t August 5, 1997 1

i J

John R. Dukes Dukes Associates 236 Weydon Road Worthington, Ohio 43085 l

l

Dear Mr. Dukes:

This letter is in response to your application dated January 24,1997, requesting

)

registration of the Model TG-2 gauging device. We are in the process of evaluating your request. However, in order to continue our evaluation, the following information is necessary:

1.

Your application indicates a change in the air gap to 77 cm (30 inches) for low energy gamma sources. This exceeds the max air gap defined in registration certificate NR-109-D-123-B. Please provide external radiation levels and dose assessments for devices with air gaps up to 77 cm.

2.

The rotary shutter in the Description Section of NR-109-D-123-B is stated to be used only in devices containing Sr-90 beta sources with maximum activity of 400 mci. However, this statement was not addressed in the application letter.

l Therefore, please indicate the circumstances in which the rotary r,hutter be used i

instead of the linear drive shutter.

Please note that if ABB wishes to transfer registration certificate NR-109-D-123-B to inactive status, they will need to submit an amendment request to the NRC.

Please provide the requested information as soon as possible. If you have any questions, please contact me at (301) 415-5799 or Mr. John Lubinski at (301) 415-7868.

L j

Sincerely,

/s/

Eric B. Compton, Engineering Co-op Sealed Source Safety Section

//

Medical, Academic, and Commercial

/c Use Safety Branch Division of Industrial and Medical Nuclear Safety Office of Nuclear Material Safety And Safeguards q%9) cc: Donald Riesbeck, IRMS, Inc.

9708080186 970805 PDR RC SSD PDR n m e n,,,e u w }

pmW g

n.

I l

i e

i i

Distribution:

4 SSSS r/f SSD-97-05 NE01 SSD File # NR-109-D-123-B DOCUMENT NAME: H:\\ERIC\\COMPLTR\\lRMS.DEF T,,.c.i

. c.,y.: inie 4.com.ni, ineic.i. w th. box: c = copy wiinout.ttachment/ enclosure "E" = Copy with attachment / enclosure "N' = No copy OFFICE IMAB 16 IMAB lQ l

l NAME ECompton Fc JLubinski'2J DATE 08/r/97 08/f/97 OFFICIAL RECORD COPY 1

1 l

i