Safety Insp on 930811 of License 50-29059-01.Insp Findings: Linearity Test Performed to 323 Mci & Linearity Test Performed to 316 Uci
ML20149D626
Person / Time
Site: 03033011
Issue date: 08/24/1993
From:
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION V), VALLEY HOSP., PALMER, AK
To:
Shared Package
ML20149D612 List:
References
NUDOCS 9309210105
Download: ML20149D626 (1)


See also: IR 05000290/1959001

Text

! rid: FORM Q1 PART 1

U.S. NUCLEAR REGUuTCITY COMMISSION

y.,,,

SAFETY INSPECTION

,o u . e.roi

1

Page 1 of

1.UCEV EE

2. REGIONAL OFFICE

_ valley Hospital

agg,gg y

Imaging Department

U S NUCLEAR REGULATORY COMMISSION

515 East Oahlia

1450 MAR!A MNE SUITE 210

Palmer, Alaska 99645

WALNUT CREEK CA 94596-5368

3. DOCKET NUMBERIS)

4. UCENsE NUMBER (S)

5. DATE OF INSPECTION

030-33011

50-29059-01

August 11, 1993

LICENSEE:

The inspection was an exarnination of the activities conducted under your license as they relate to radiation safety and to compliance with the Nuclear

Regulatory Commission (NRC) rules and regulations and the conditions of yoc iicense. The inspection consisted of selective examinations of

procedures and representative records, interviews wi:h personriel, and observations by the inspector. The findings as a result of this inspection are as

follows:

1.

Within the scope of this inspection, no violations w6re observed.

~

2.

The inspector also verified the steps you have taken to correct the violations identified during the last inspection. We have no further

questions on those actions at this time.

-

3.

During this inspection certain of your activit;es, as described below or attached, were in violation of NRC requirements. This form is a

NOTICE OF VKXATION, which is required to be posted in accordance with 10 CFR 19.11.

A.

was not properly posted to

indicate the presence of a

10 CFR 20 203(b),(c),[d),(e) or 34.42.

B.

of seated sources were not

periormed at the proper frequencies.10 CFR

or Ucense Condition Number

C. Records of

were not property maintained.

10 CFR

or Ucense Condition Number

D. Documents were not property posted or otherw;se made available.10 CFR 19.11.

E. Reports or notification of

were not made in accordance with

10 CFR

or Ucense Condition Number

X

10 CFR 35.50(b)(3) requires that linearity tests be performed over the range of

F.

its use between the highest dosage that will be administered and ten microcuries.

Contrary to the above, on March 22, 1993, the linearity test was performed to

323 microcuries and on June 30, 1993, the linearity test was performed to 316

microcuries.

I hereby state that, within 30 days, the actions described by me to the inspector will be taken to correct the violations identified in the items checked

above. This statement of conective actions is made in accordance with the requirements of 10 CFR 2.201. No further response will be submitted

unless required by the NRC.