FCTC: RHO T8 71-9201 OCT 0 3 I!NE Nuclear Packaging, Inc.
ATTN: Dr. Charles J. Temus 1010 South 336th Street j
Federal Hay, UA 98003 Gentlenen:
l This refers to your application dated August 30, 1985, requesting approval of the l'.odel No. ELEKTA TIGER package.
The application has not been accepted for detailed review since information has not been provided which would pernit a neaningful review. Primary inadequacies are noted in the enclosure to this letter. The eight j
copies of the safety analysis report for which withholding was requested i
are being returned under separate cover.
I If you have any questions regarding this natter, we would be pleased to meet with you and your staff.
If you decide to pursue the application, a new application should be subnitted which takes into account the enclosed comnents and the requirements of Part 71.
Sincerely, cristi Signed by gygJ,gs I. MACD088 l
Charles E. MacDonald, Chief Transportation Certification Branch l
Division of Fuel Cycle and
!!aterial Safety, INSS
Enclosure:
As stated Distribution: w/ encl RH0degaarden (2)
DTHuang Docket File l
NRC PDR i
IE HQ e5}OOJ$$$ $N 1
N
/
FCTC R/F d/Walwans 1
QJg{
".c o.
FCTC Im I,Hua g T
CEMa nald !N/J//g.n.r i.u,,,J egaarden:a' l
~ 10/eb/85~
10/32./85-10/7/85-- Ljg/g/gf j
..{..
nne rone na oo-soi nncu ouo OFFICIAL RECORD COPY
i l
Nuclear Packaging, Inc i
Model No. ELEKTA TICER Packaae l
Docket No. 71-9201 OCT O 3198$
Encl to ltr dtd:
t DRAWINGS t
1.
Engineering drawings and naterial specifications were not provided 2
jt for the LGS radiation unit and ISO fittings.
j 2.
Drawing No. JR-20-1000 j
a.
Wall thickness shown in Details C and F are not in agrecaent.
I
)
b.
Length of tube as shown in Detail C is not specified.
l STRUCTURAL L
i i
1.
Justification was not provided for relaxed stress allowables used j
for the design of the ELEKTA TIGER.
)
2.
A fracture control plan was not provided for all components of the i
packaging including LGS radiation unit.
)
3.
Brittle fracture analysis for the LGS radiation unit did not consider j
localized bearing stresses by the support dunnage or identify the support dunnage.
l 4.
A denonstration was not provided which shows that the sealed sources will remain positioned within the LGS radiation unit or the unit remains effective following the accident conditions.
4 5.
Top or bottom edge drop of the ELEKTA TIGER was not considered.
The side edge drop may not be the worst case in deternining the l
adequacy of the lid holts.
6.
Puncture analysis which compares packages of different materials and weights is not substantiated. Also, the puncture analysis did not take into account sequential damage caused by previous drop tests.
00ALITY ASSIIRANCE PROGRAM I
An NRC approved Quality Assurance Progran was not identified for the design and fabrication of the LGS radiation unit.