Forwards Spec Pages II 4-16 & II 4-17 for Incorporation Into 850423 Application for Renewal of License SNM-696.Pages Include Addl Info Re Testing,Calibr & Maint of Criticality Alarms
ML20128E906
Person / Time
Site: 07000734
Issue date: 05/20/1985
From: Mowry W
GENERAL ATOMICS (FORMERLY GA TECHNOLOGIES, INC./GENER
To: Crow W
NRC OFFICE OF NUCLEAR MATERIAL SAFETY & SAFEGUARDS (NMSS)
References
25297, 696-7056, NUDOCS 8507080032
Download: ML20128E906 (8)


Text

-

.r.w j~ -

In Reply Refer To: 696-7056 GA Technologies Inc.

s2N 20 May 1985 Go UFoRNIA 92138 o

O,'

(619) 455-3000 v

j' RECENED L f 39$

U\\iSd Q&} >

d.S. NUCLEAR PEGUly0gr' '

Dr. William T. Crow certam U.S. Nuclear Regulatory Commission 6

ne:

g b!! suctma Office of Nuclear Material Safety & Safeguards Washington, D.C. 20555 G

g

Subject:

Docket 70-734: SNM-696; Supplement to Request for Amendment of Certain License Specifications and Related License Conditions.

(6 copies)

Dear Dr. Crow:

GA Technologies Inc. (GA) has recently submitted a request for amendment of certain licinse specifications and related license conditions. 'Ihis request was submitted April 23, 1985.

Subsequently in discussions with your staff, it was agreed that additional changes to the section dealing with testing, calibration and maintenance of criticality alarms was in order. Attached herewith are four pages containing additional clarifi-cations of the testing and calibration requirements as well as some ex-panded statements regarding the conditions under which licensee would be exempt from certain requirements of 10 CFR 70.24.

Please substitute these pages for the Pages II 4-16 and II 4-17 of the recently renewed Specifications Volume.

Page II 4-16 of this submittal supersedes II 4-16 of our April 23, 1985 correspondence. An instruction sheet is enclosed.

We believe that the expanded text of this submission answers all the ques-tions raised by your staff in the matter of criticality alarm testing and calibration.

We look forward to your early approval of this and our April 23 request since we have scheduled a shutdown of the criticality alarm system within the SVA facility so we may begin modificatic,ns of the system on May 30.

1 C

(o f

4 f0CKETED 2

USNRC j

MAY2 41985 > 5h app

, p_

n.m m

i Mall HCTiON j M,d" s

a

^

DocxEr cttu m,, x,As pff N,

j eso70eoosa e50520

~ _ -

PDR ADOCK 07000734 C

PDR J

e William T. Crow ' 696-7056 Thank you for your staff's' cooperation in expediting our request. Should there be any additional questions in the matter, please don't hesitate to contact me at (619) 455-2823 Very truly yours, William R. Howry Licensing Administrator WRM:he

Enclosures:

1.

' Specifications Volume:

Pages II 4-16. II 4-17. II 4-17a, II 4-17b 2.

Instruction sheet oc: NRC, Region V (1 copy) i J

=

0 Instructions for updating SNM-696 Specifications Volume 20 May 1985 Remove Add Page II 4-16, Rev. 2, dated 11/83 Page II 4-16, Rev. 5, dated 4/23/85 Page II 4-17,. Rev. 3, dated 4/84 Page II 4-17 Rev. 5, dated 4/23/85 Page II 4-17a, Rev. 5, dated 4/23/85 Page II 4-17b, Rev. 5, dated 4/23/85 k

'a.

~.

4.2.1 3 Radioactive Material Detection and Assay Portable instrumentation shall be available and utilized for detection and assay. This instrumentation includes:

X ray and gamma - Geiger-Muller (GM) detectors, scintillation detectors, scintillation detectors coupled with a portable scaler for assay of low levels of activity, scintillation detector coupled with a single-channel analyzer, and ion chambers.

Beta - GM detectors and ion chambers.

Alpha - Air proportional and scintillation detectors.

Neutron - mrem or Rem reading neutron monitor.

Fixed instrumentation shall be available and utilized as required for radiological safety purposes. Such instruments shall include counting equip-ment, spectrometers for analysis of samples, calibration sources (s) to cali-brate instruments, and meteorological equipment for measurement of certain meteorological conditions.

4.2.1.4 Criticality Monitoring and Alarm System The licensee shall maintain in each area where SNM is handled, used, or stored a monitoring system using gasma or neutron-sensitive radiation detec-tors which will energize clearly audible alarm signals if accidental criti-cality occurs. This system shall meet the performance requirements of 10 CFR 70.24(a).

Each system shall be tested monthly using internal check sources or portable sources.

Individual channels of the system will be recalibrated when whichever occurs first:

1.

detector and amplifier response to a check source is > +50% from expected value.

License No.

SNM-696 License Amendment No.

Page

,, y c no4-4-,

e n,

.. r - e m e

,,,,,-,a n-

2.

detector and amplifier response to internal " keep alive" source or known external radiation field is not within +100% of expected value.

3 prior to reentry into service following any required maintenance.

The alarm trip levels may be set at levels between 5 mR/hr and 20 mR/hr for non-coincidence systems and a correspondingly appropriate predetermined level if the system is a coincidence type. They will be readjusted if the alarm point fails to activate within 25 see on at least three of four trials.

l Any area which meets the criteria of 1, 2 or 3 below and which has prior CRSC approval is exempt from the monitoring system described above.

1.

General Spaces i

a.

The area shall be a defined nuclearly isolated space, area, laboratory, building or facility established to control SNM activities, b.

The area shall have established an MBA which shall be limited to not more than 700 grams contained U-235, 500 grams U-233, 400 grams plutonium or 400 grams of a combination thereof.

i c.

The area's activity shall be under an approved work authoriza-tion.

d.

Administrative and procedural and/or physical constraints exist to preclude the introduction of more than the authorized amount of SNM to the controlled area.

l' l

[

\\

License No.

SNM-696 License Amendment No.

Page Revision No.

5 Date 4/23/85 Released By II 4 17

2.

Hot Cells a.

Heavily shielded cells designed to contain highly irradiated fuel, reactor components, or byproduct material sources, b.

The operating and service galleries of the cells shall be alarmed.

3.'

Other a.

Any area used for temporary storage of SNM in authorized ship-ping containers on transport vehicle pending its shipment or delivery to on-site storage area.

In addition, the licensee is exempt from the 10 CFR 70.24 monitoring system requirements when performing system repair or modifications provided that:

1.

special attention is given to minimizing the period of inoperability.

2.

no material handling will be allowed in the area while the alarms are inoperative unless such handling is to mitigate a significant Health Safety problem or provide required physical protection and receives prior approval of operating management, the managers of Nuclear Safety & Health Physics, and the Chairman of the CRSC.

3 and facility operating staff will be informed of the special operating circumstances limiting SNM handling.

- License No.

SNM-696 License Amendment No.

Page Revision No.

5 Date 4/23/85 Released By II 4-17a

The licensee also is exempt from the 10 CFR 70.24 alarm audible require-ment'when planned operations will result in radiation levels above the 5-20 mR/hr monitoring system trip levels provided that:

1.

the radiation level is continuously measured and is under observa-tion during the interval of bypass of the alarm's audio; 2.

any unrelated SNM handling in the area shall be suspended during the period of bypassed alarm audio; 3

the bypassing of the audio alarm is accomplished by or under di_rection of Health Physics and s

4.

The system will be tested for complete operability at the time it is returned to normal service.

y.

License No.

- S NM-696 License Amendment No.

Page Revision No.

5 Date 4/23/85 Released By II 4.17b

,,../

DOCKET NO._

["

CONTROL 10._ MDMf DATE OF DOC.__ Ob[h DATE RCVD._ [I[Mk PCUF PD1 FCAF LPDR WM_.

ISE REF._

WMUR__

SAFEGUARDS FCTC OTilER DESCRIPTION:

/