ML20080E111

From kanterella
Jump to navigation Jump to search
Rev 1 to Procedure 08-S-04-109, Chemistry Instruction: Operation of Orion 701 Ion Analyzer,Safety Related
ML20080E111
Person / Time
Site: Grand Gulf  
Issue date: 09/24/1983
From: Wedgeworth B
MISSISSIPPI POWER & LIGHT CO.
To:
Shared Package
ML20080E069 List:
References
08-S-04-109, 8-S-4-109, NUDOCS 8308300727
Download: ML20080E111 (8)


Text

f i

~

g PLANT OPERATIONS MANUAL

~

P.

Volume 08 08-s-04-109 Section 04 Revision 1 Date:

9-24-82 r

CHEMISTRY INSTRUCTION f

OPERATIO_N OF ORION 701 ION ANALYZER l

SAFETY RELATED

/

Prepared:

J/

[

bj ft/1, A

[1 Reviewed:

% /

NYM

~

tec rtij:fl1Revit!,

q 44' g independ'ent Beview

/

. ~ ' ' II h

, ij f,&w A

Approved:

'J

' i' / O _$ %

Plaa t r.

hedItt4 Q4,ij fj g,

'a -

Concurre.nce:

/

T2

/

2' Chemis t ry/Radiat t6n4 Tl pe, tftenQ /

List of Effective Pages:

Page I

1 1-7 l

List of TCN's Incorporated:

Revision TCN No.

O None 1

None 9

-r 83083C0727 830825

[

PDR ADOch 05000 i

E

GRAND GULF NUCLEAR STATION CHEMISTRY INSTRUCTION

Title:

Operation of Orion 701 Ion No.:

05-S-04-109 Revision:

1 Page:

1 Analvzer

(-

i 1.0 INSTRUMENT DESCRIPTION Orion Digital pH/Mv Meter 701A and Manual Electrode Switch Model 605

2.0 REFERENCES

2.1 Orion Research Instruction Manual IM701A, 460000949 2.1 Orion Research Instruction Manual IM605, 460001037 2.3 Chemistry Section Procedure 08-S-03-1, Qualificacion of Chemistry Program 2.4 Chemistry Section Procedure 08-S-03-7, Control of Chemistry Support j

Equipment 3.0 DEFINITIONS l

3.1 Specific lon Electrode - A device that dev' elopes an electrical potential i

proportional to che logarithm of the activity of an ion in solution.

4.0 PREREQUISITES 4.1 Specific ion electrode 4.2 Reference electrode 4.3 An assortment of reference standards 4.4 Reference electrode filling solution 5.0 PRECAUTIONS 5.1 Always make measurements or calibrations with the test solution in a non-metal conr;.iner.

5.2 Place the instrucent function switch in standby prior to removing the electredes from solution.

e N_-

GRAND CULF NUCLEAR STAT' ION CHEMISTRY INSNCTION

Title:

Operaticn of Orion 701 Ion No.:

08-5-04-109 Revision:

1 Pega:

2 Annlyznr 5.3 To eliminate warm-up time and increase ' omponent life, always leave the

-)

c meter in the STDBY mode rather than disconnect from line power.

Warm-up

..thee is 30 minutes.

5.4 Check and ensure reference electrode filling solution is at least one inch above the solution being measured.

5.5 Allow all samples and standards to reach the same temperature before attempting a measurement.

5.6 Stir both standards and samples with a magnstic stirrer while a specific ion measurement is being made.

Some magnetic stirrers generate enough heat to increase the te perature of the solution.

To r/o ld th is, p'.2 :e t piece of thermal insulating material between the stirrer and beaker.

5.7 Rinse electrodes with decineralized water between measurements.

Do not blot dry.

5.4 Never use an individual reference and a common reference electrode in the same solution.

6.0 INSTRUCTIONS 6.1 Functional Check - 701A Ionalyzer s

6.1.1 Connect the shorting strap-between the input jack and reference

')

jack, and turn the function switch to STDBY.

6.1.2 Conn =ct the power cord of the 701A to a suitable power source.

~

Let the meter varm up for thirty minutes.

6.1.3 With the. function switch on STDBY, verify that 2 decimal point only is displayed in the middle of the display.

6.1.4 Turn the function switch to REL MV and verify a display of polarity sign and four digits.

D 3

4%

'* m,

2

f RfND GULF NUCLEAR STATION CHEMISTRY INSTRUCTION l

Title:

Operaticn of Oricn 701 Ion No.:

08-S-04-109 Rsvision:

1 Pega:

3 Analyzer p

6.1.5 Using the calibhtion control, verify that,the display is i

adjustable from about -200mg to +2.00mv.

Adjust the calibration control for a disjilay of'000.0.

6.1.6 Turn the function switch to pH/.001.

6.1.7 Turn the ISO ADJ screw' to obtain the isopotential point (7.000).

Verify that turning the temperature compensator' knob does not change the reading by more' than +0.002 pH.

6.1.8 Turn the function switch to,MV.

Verify a display of 000.0+ 0.1.

If the. display is outside these limits, repeat steps 6.1.6 and

~

6.1.7.

If still out, terminate test and notify Chemistry Supervisor.

(

6.1.9 Turn the function switch to pH/.001.

Set the slope indicator dial to 100% and turn the temperature compensator knob until the white arrow points to 29'C.

At this temperature a change of one pH unit corresponds to a 60 my change in potential.

6.1.10 Turn the calibration control until a reading is obtained that is 0.50 pH units less than the isopotential point.

6.1.11 Set function switch to REL MV and verify a display of 30.0_+ 0.3 mv.

If the ceter does not check out in any of the above steps, repeat the step.

If the second attempt fails, contact the Lead Laboratory Chemist.

6.2 605 Electrode Switch 6.2.1 Operation The Channel Selector Switch on the instrument is used to a.

select from any chsnnel 0 through 5.

khen channel 0 is selected, standardization is accomplished utilizing the pH/mV ins t rument standardization control and not one of the five chcnnel :entrols on the electrode s itch.

Channel O cust be standardized before switching the electrode switch on.

l

lTitlo: Operctica of Orion 701 Ion No.:

08-S-04-109 lRsvision:

1 Pcga:

4 l

Anqlyzne l

b.

Standardize each channel u' sed by placing the selector switch

-)

to the corresponding channel and rotating the calibrace control for that respective channel to the desired reading.

c.

To verify the original setting at any time, connect a shorting lead between channel 0 reference and sensing electrode input jacks and note the readings.

If the readings do not coincide with the original readings, the controls have been moved and the channel will require restandardization.

6.2.2 Battery Check a.

To perform an instrucent battery check, verify that the electrode switch is connected to the respective, meter,.i.e.,

701A.

b.

Set the Meter Selector Swith to the MV position.

c.

Switch the electrode switch off.

d.

Connect a shorting lead between the battery ' check jack on the rear panel and the sensing electrode input jacket of channel-0.

Verify thac the electrode swit.h channel _ selector points to j

e.

channel 0.

\\

f.

Read the battery voltage on the meter. A reading of 250 millivolts or greater indicates proper batcery voltage.

If less, contact the Chemistry Supervisor.

6.3 Operation 6.3.1 The 701A Ion Analyser can be used for both pH measurements or specific ion analysis.

a.

pH Measurements e

15 4

/.

I

\\',

)

U

- ~-~

GRAND GULF NUCLEAR STATION CHEMISTRY INSTRUCTION

Title:

Operation of Orion 701 Ion No.:

08-S-04-109 Revision:

1 Page:

5 Analyzer r~

n I

\\.

~

(1) SelectekobufferswhosepHv,aluesbrackettheexpected sample pH, with one buffer between pH 6 and pH 8.

Check the label on the buffer packet or bottle for the exact pH at the solution temperature.

o (2)

Settheisopotenti'a}2 point to the pH of the buffer between pH 6 and 8.

(To change the isopotential point from the factory setting of 7.000, connect the shorting strap between the input and reference jacks.

Turn the functica switch to,REL MV and set the reading to 000.0 with calibration control. Turn the function switch to pH/.01. Turn the ISO ADJ screw on the back of the meter until the desired isopotential point is displayed.)

(3) Remove the shorting strap.

Attach the combination pH l

electrode or pH electrode and reference electrode securely to the meter.

Turn the function switch to i

pH/.01.

(4)

Place electrodes in the buffer whose pH is closest to isopotential point.

Turn the calibration control until the isopotential point is displayed.

j

(

(5) Remo e the electrodes from the isopotential buffer and 1

rinse.

(6) Place the electrodes in the second buffer solution.

(7)

If necessary, turn the temperature compensator until the pH value of the second buf fer is displayed.

Turn the-slope indicator dial antil the temperature compensator arrow points to the buf fer temperature.

Read the percent of the theoretical Nernst slope at the "% slope" index line.

If the % Nernstian slope is less than 90%, repeat the standardization with fresh buffers.

If the % slope is again less than 9CL, see KEASUREMENT DIFFICULTIES, page 16 of Reference 2.1.

e 4

-- bra saasasacsoeacasa f-~

Titto:

Opercticn of Orica 701 Icn Na,:

08-S-04-109 R0 vision:

1 Pag 2:

6 '"'"

Ancivzcr

)

)

(8) Remove electrodes, rinse, and place in the unknown s ample.

(9) Allow about one minute for the reading to stabilize, and

- record the pH measurement directly from the display.

(ib)

Place the instrument function switch in standby, rinse the electrodes and store in demineralized water.

b.

Specific Ion Measurements NOTE Individual instructions are written for each specific icn measurement.

The following is to be used as a general instruction for use of the 701/605 c omb ina t ion.

Whe re conflicts exist, specific lon measurement instructions take precedence:

(1) Multiple specific ion measurements can be made by connecting the Orion Model 605 manual electrode switch to the ion analyser.

(2) Prepare three standards that bracket the unknown sample.

concentration.

Add the ion:c strength adjustor, or pH adjustor, recommended by.he electrode instruction

-}

manual.

For greater accuracy, more standards may be prepared and measured.

(3) Place electrodes in the mid range standard and select proper channel on the electrode switch.

(4) Set the function switch to the mV (absolute) position.

Wait for the prescribed time, read, and record.

(5) Set the function switch to the REL mV position. Wait for,

a stable reading.

Adjust the individual calibration control until the meter reads 000.0.

Since the of fset potential in the REL MV mode is -about +200my, it is t

S ',

4

'6

.\\

)

GhAND GUII NUCLEAR STATION CHEMISTRY INSTRUCTION

Title:

Operation of Orion 701 Ion No.:

08-S-04-109 Revision:

1 Page:

7 Analyzer e.

g M

impossible to zero the meter when the electrode potential is outsi,de these' limits; in these cases, simply turn the function switch to t.he MV position and record the electrode potenciaIs to 0.1mv,.

(6) Rinse electrodes and place in the most dilute standard.

Wait for the prescribed time, read and record.

(7) Rinse electrodes and place in the most concentrated standard.

Wait for the prescribed time, read, and record.

i (8) Using semilogarithmic graph paper, prepare a calibration curve by plotting the electrode potentials measured (millivolt readings) on the linear axis and the activity or concentration value of the standards on the logarithmic axis.

(9) Rinse electrodes and place in the unknown sample. Wait for the prescribed time, read and record.

(10) Using the calibration curve,' determine the concentration of the unknown.

(

7.0 DOCUMENTATION / CORRECTIVE iCTION 7.1 Enter operator's initials and date check performed on standardization card attached to the instrument.

7.2 Document results on proper reporting forms as per Chemistry Procedure 08-S-03-10.

w2 e

e

__________m____.-__._____----_

_