Forwards Medtox Investigation of Specimen E2037 & Util Nuclear Quality Dept Rept of Investigation Dtd 930204,per Discovery of Blind Performance Test Discrepancy on 940111. Discrepancy Caused by Procedural Error in Dilution Process
ML20073R984
Person / Time
Site: Monticello, Prairie Island  
Issue date: 02/10/1994
From: Richard Anderson
NORTHERN STATES POWER CO.
To:
NRC OFFICE OF INFORMATION RESOURCES MANAGEMENT (IRM)
References
NUDOCS 9402160220
Download: ML20073R984 (5)


Text

.o Northem States Power Company 414 Nicoliot Mall l

Minneapolis, Minnesota 55401-1927 Telephone (612) 3345500

\\

l February 10, 1994 10 CFR part 26 Appendix A

Section 2.8.e.4 U S Nuclear Regulatory Commission Attn: Document Control Desk Washington, DC 20555 l

MONTICELLO NUCLEAR GENERATING PLANT Docket No. 50-263 License No. DPR-22 PRAIRIE ISLAND NUCLEAR GENERATING PLANT Docket No. 50-282 License DPR-42 50-306 DPR-60 Unsatisfactory Blind Performance Test l

On January 11,1994 a blind performance test discrepancy was discovered by an NSP FFD collector and subsequently reported to the NSP Medical Review Officer (MRO). According to the blind specimen supplier, El Sohly Laboratories, the blind specimen in question ( Chain of Custody # 996202) was spiked positive for opiates, with a known value of 596 ng/ml l

(Codeine) and 617 ng/ml(morphine). Initial testing of the sample by MEDTOX ccafirmed the presence of opiates, however concentration levels were detected and recorded at approximately double the known values. NSP's MRO uses a deviation rate of 15% for acceptance.

In accordance with 10 CFR part 26, Appendix A, Section 2.8.e.4, NSP completed an investigation into the unsatisfactory blind test result and identified correciive actions to prevent reoccurrence. (See attached Nuclear Quality Department (NQD) report of investigation dated l

February 2,1994)

Results of NSP's investigation indicate that a procedural error in the dilution process was the l

cause for the testing discrepancy. MEDTOX failed to dilute the sample by a factor of (2) as was annotated on the GC/MS confirmation cover sheet. A MEDTOX analyst believing that the sample had been diluted by two, calculated the test results by doubling them, thus accounting for the 100% discrepancy.

l l

f O

9402160220 940210 PDR ADOCK 0500o282 i,-

L 11i P

PDR tj

USNRC February 10,1994 Page 2 Northern States Power Company As further required by 10 CFR 26 Appendix A, Section 2.8.e.4, attached is a record, signed and dated by the individuals responsible for day-to-day management and operation of MEDTOX, identifying investigation findings and corrective actions to prevent recurrence.

This letter contains no NRC commitments.

Please contact Mel Opstad (612-337-2038) if you have any questions related to this matter.

M TCh sor Roger O. Anderson Director, Licensing & Management Issues C:

Regional Administrator, Region lli NRR Monticello Prcject Manager, NRC Monticello Sr Resident inspector, NRC NRR Prairie Island Project Manager, NRC Prairie Island Sr Resident inspector, NRC Jay E. Silberg Attachments:

MEDTOX Investigation of Specimen # E2037, Signed & Dated 01-31-94 NSP Nuclear Quality Department Investigative Report, Signed & Dated 02-03-94 NSP Distrbution :

1 Doug Antony George Miserindino Randy Cleveland Craig Johnson Brian Anderson Frank Evitch Dr. James Woodburn l

l l

02/09/94 12:08 FAI 812 838 5351 MEDTDI IABS* *

@002/002 MEDTD(S LA40lATotIE 4

612-636-7466 SaintPaul, Minnesota $5112 e

402 West County RoadD e

l

\\

MEDTOX INVESTIGATION OF SPECIMEN _

J ME203704) his document relates to the investigation of a blind specimen submitted by Northem States Power to MEDTOX Laboratades on 1/10/94. The specimen was identified with the social security number 530-30-8445 and chain of custody number 996202. This geTen The sample was positive for opiates was assigned the MEDTOX acr*=im number E203704.

by immunoassay and confirmed by GC/MS with codeine - 1244 ng/mi and morphine - 1322 ng/ml. The results were reported on 1/12/94. NSP requested that the specimen be sent to Smith Kline Laboratories in Chicago for an opiate zetest. NSP requested MEDTOX to retest the specimen on 1/27/94. He retest was still positive for codeine and morphine. He codeine - 656 ng/ml and morphine - 638 ng/ml were appnndmately one-half of the miginal values reported.

The calibrators and controls were evaluated for both GC/MS assays and found to be i

acceptable. All three controls on both assays were well within their reference ranges. In both assays the specimens were diluted by 2 due to the high immunoassay value (greater than 125). De dilution is pufmmcd in order to accommmhwe the high codeine and maphine specimens encountered so that less specimens need to be repeated and turn around time j

decreased i

The first specimen set up on 1/11/94 by the analyst was not diluted in half btfore the assay was run but was multiplied by two thus sw" dog for the two-fold difference in results.

The following cmrective action has been instituted. The GC/MS confirmation cover I

sheet has been modlBed to more effectively distinguish which specimens are to be diluted.

This pmblem has been discussed with the analyst and the proper pmeedure for insuring that the diluted samphs are handled correctly reinforced. The Forensic Supervisor, observed the analyst perform a GC/MS confirmation set-up with dilution protocol to insure that proper procedures are being followed. Documentation will be put into the individuals pe.rsvunel file.

j In addition, the proper dilution handling procedure has been discussed with the GC/MS staff and the importance of the proper handling reinforced.

/ M Y

l'AP Q

MEDT0X Laboratodes.Inc.

MEDT0XBio-Analytical MEDT0XBio Analytical Los Angeles, Calfornia Chicago, Illinois e

SaintPaul. Minnesota e

612-636-7466 312-880-6900 818-718-0115

Internal Correspondence case February 2,1994 FroA LE lirehm, Sr QA Engineer uaion RSQ 08 To RD Cleveland, Personnel Secur Team Leader location RSQ 06 suNect Review of Illind Sample, chain of custody #996202 On January 27, 1994, you requested from the Supervisor Quality Assurance Audits that an investigation be conducted and documented with regards to the error of a blind sample sent by NSP to MEDTOX for analysis on January 10,1994. The purpose of this memo is to document the results of the review of the MEDTOX investigation by NQD of that blind sample.

The sample was provided to NSP by El Sohly Laboratories, Inc., Batch 0-0319, cor trol number l

530-30-8445, and submitted under the chain of custody number 996202. The known value for the specimen was codeine = 596 ng/ml and morphine = 617 ng/ml, The sample was confirmed l

positive for opiates at MEDTOX on January 11,1994 by GC/MS with codeine = 1244 ng/ml l

and morphine = 1322 ng/ml. This is an approximate error of 100%. The MRO at NSP uses an error ofless than 15% for acceptance.

Upon notification of the discrepancy, the MRO requested on January 18,1994 that the sample be sen 'o SmithKline Beecham Laboratories in Chicago. These results were codeine = 624 ng/ml and morphine = 720 ng/ml, an approximate error of 5% for codeine and 15% for morphine. The MRO then requested that MEDTOX retest the sample. These results were codeine = 656 ng/ml and morphine = 638 ng/ml, an approximate error of 10% and 3%,

respectively.

Laboratory Codeine (ng/ml)

Morphine (ng/ml)

El Sohly 596 617 l

MEDTOX (initial) 1244 1322 SmithKline 624 720 MEDTOX (retest) 656 638 l

On January 31,1994,1 met with Dr. Gary Hemphill and Ms. Jennifer Collins of MEDTOX to discuss the cause of the above discrepancy. MEDTOX found upon investigation that this was a dilution error. Typically, when the initial immunoassay indicates a level greater than 125 ng/ml, the GC/MS Confirmation Cover Sheet is annotated to dilute the sample by two. This reduces turn around time due to reducing the need to rerun samples because of instrument limitations.

In this case, the sample was not diluted by two as was annotated on the GC/MS Confirmation Cover Sheet, but when calculating the result, analyst assumed that the sample had been diluted as indicated and multiplied the results by two, thus accounting for the 100% discrepancy. It appears that the sample was not diluted due to an error by the analyst, who is a relatively new employee (3-4 months). The GC/MS Confirmation Cover Sheet used to direct the dilutions is

i confusing. The need for dilution, altnough identified,is not readily distinguishable. This type of error has not occurred prior to this at MEDTOX. It should be noted that NSP had no other l

specimens in the original batch of ten specimens (011194A). It should also be noted that the possibility for a false positive to have occurred due to this type of error is very sn ' The l

initial inununoassay screens the specimen with no dilutions such that those below or cutoff 1

level are not confirmed using GClMS.

In order to prevent recurrence, MEDTOX revised the GC/MS Confirmation Cover Sheet to-better identify those specimens that are to be diluted. In addition, the Forensic Supervisor observed the analyst responsible for this error perform a GC/MS setup with dilution to verify proper performance. No performance problems were found. During the setup process, the l

analyst physically moves the sample in the rack after the dilution. The analyst was counseled l

regarding the procedure for insuring that the diluted samples are properly handled and a letter l

describing this incident was placed in his file. The proper dilution handling procedure was l

discussed with the staff and the importance reinforced.

l I conclude that no false positives should have resulted due to this error and that MEDTOX has implemented corrective action to prevent recurrence.

This error does not reduce the effectiveness of NSP's Fitness for Duty Program.

Any questions, I can be reached at 337-2204.

Sincerely, m

LE Brehm Sr QA Engineer cc:

Dr. J Woodburn Dr. DG Hemphill(MEDTOX)

TE Amundson l