ML20071D017

From kanterella
Jump to navigation Jump to search
Responds to NRC Re Automated Analytical Instrumentation Sys Discussed in IE Insp Repts 50-254/82-13 & 50-265/82-15.Peak Search Algorithm Used in Measuring Radioactivity Will Include Resolution of Multiplet Peaks
ML20071D017
Person / Time
Site: Quad Cities  Constellation icon.png
Issue date: 02/14/1983
From: Rybak B
COMMONWEALTH EDISON CO.
To: James Keppler
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
Shared Package
ML20071D014 List:
References
5592N, NUDOCS 8303090107
Download: ML20071D017 (2)


Text

Commonw:alth Edison z

ey'

[i one First N 1 on21 Pia 7a Chicsgo, lil. noir L'

-/' Addr:ss R; ply to Post Office Box 767

("

Chicago lilinois 60690

~

6 February 14, 1983 Mr. James G. Keppler, Regional Administrator Directorate of Inspection and Enforcement - Region III U.S. Nuclear Regulatory Commission 799 Roosevelt Road Glen Ellyn, IL 60137

Subject:

Quad Cities Station Units 1 and 2 Inspection Report Nos. 50-254/82-13 and 50-265/82 Open Items NRC Docket Nos. 50-254 and 50-265 References (a):

C.

J. Pareriello letter to Cordell Reed dated October 1, 1982.

(b):

T.

J.

Rausch letter to J. G. Keppler dated October 1, 1982.

(c):

C. J. Paperiello letter to T.

J.

Rausch dated December 6, 1982.

Dear Mr. Keppler:

The referenced letters discussed resolution of an open item concerning multiplet analysis capability of our Automated Analytical Instrumentation System (AAIS) to which we committed to review our software system.

A review of the AAIS Sof tware has been completed and improvements will consist of the following:

1.

The peak search algorithm used in measuring radioactivity will include resolution of multiplet peaks.

2.

The calucations for error analysis will be included.

3.

The ability to subtract background radiation wi 1 be included.

4.

Where applicable, a determination of the lower limit of detectability will be included.

5.

The algorithms for classifyireg peaks will be modified.

6.

The algorithms for determining detector efficiency will be modified.

G63 9303090107 830304 PDR ADOCK 05000254 fgg 16 O

PDR

4

'/

2 J. G. Keppler February 14, 1983 A schedule for completion of the software changes will be pr'vided by June 15, 1983.

Please direct any questions you may have on this matter to this office.

i Very truly yours, B. Ry x

Nuclear Licensing Administrator 1m cc:

Region III Inspector - Quad Cities 1

5592N l

t