Appendix NOTICE OF VIOLATION Bothwell Memorial Hospital License No. 24-16275-01 As a result of the inspection conducted on April 1,1982, and in accordance with the NRC Enforcement Policy, 47 FR 9987 (March 9, 1982), the following violations were identified:
1.
License Condition No. 17 requires that licensed material be possessed and used in accordance with the statements, representations, and procedures contained in certain referenced applications and letters.
The referenced application dated December 19, 1979, states in Item 10 that the survey meters will be calibrated at annual intervals.
Contrary to the above requirement, the Texas Instruments Model 9121 Log Series survey meter has not been calibrated since December 31, 1980.
This is a Severity Level V violation (Supplement VII).
2.
License Condition No. 17 requires that licensed material be possessed and used in accordance with the statements, representations, and procedures contained in certain referenced applications and letters.
The referenced application dated December 19, 1979, states in Item 10 that the dose calibrator will be checked for activity linearity at quarterly intervals and following repair.
Contrary to the above requirement, linearity checks on your dose calibrator were not performed during the second quarter of 1980 and the third quarter of 1981.
This is a Severity Level V violation (Supplement VII).
Pursuant to the provisions of 10 CFR 2.201, you are required to submit to this office within thirty days of the date of this Notice a written state-ment or explanation in reply, including for each item of noncompliance:
(1) corrective action taken and the results achieved; (2) corrective ac; ion to be taken to avoid further noncompliance; and (3) the date when full com-pliance will be achieved. Consideration may be given to extending your response time for good cause shown.