Notice of Violation from Insp on 820520.Noncompliance Noted:Failure to Calibr Dose Calibrator Annually & Failure to Conduct Weekly Wipe Tests
ML20053E167
Person / Time
Issue date: 06/01/1982
From: Wiedeman D
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
To:
Shared Package
ML20053E160 List:
References
NUDOCS 8206070664
Download: ML20053E167 (2)


Text

-

~

I Appendix NOTICE OF VIOLATION St.-Joseph Mercy Hospital License No. 14-14700-01 i

As a result of the inspection conducted on May 20, 1982, and in accordance

.with the NRC Enforcement Policy, 47 FR 9987 (March 9, 1982), the following violations were identified:

1 i

1.

License Condition No. 17 requires that licensed material be possessed and used in accordance with statements, representations, and procedures contained in application dated May 4, 1977.

Item 11 of the application

-states that the dose calibrator will be checked for accuracy at installation and annually.

i Contrary to this requirement, it was learned through statements _of licensee representatives and the NRC inspector's review of records that this requirement is not being met Specifically, the dose calibrator was last calibrated for accuracy January 24, 1981.

This is a Severity Level V violation (Supplement VI).

2.

License Condition No. 17 requires that licensed material be possessed and used in accordance with statements, representations, and procedures contained in' application dated May 4,-1977 and letter dated August 5,

)

1977. The August 5, 1977 letter states that specific laboratory areas-will be wipe tested weekly and the results recorded, t

Contrary to this requirement, it was learned through' statements of licensee representatives and the NRC inspector's review of records that this requirement is not being met.

Specifically,' weekly surveys of the specific laboratory areas were not performed.

This is a Severity Level V violation,(Supplement VI).

3.

10 CFR 35.14(e)(1)(1) requires that sealed sources used as calibration or reference' sources be tested for leakage and/or contamination at intervals not to exceed six months.

l Contrary to this requirement, between June 21, 1979 and May 20, 1982, a 210 microcurie' cesium-137 scaled source v s not leak tested.

This is a Severity Level'V violation (Supplement VI).

i i

1 l

8206070664 820601 e

NMS LIC30 i

14-14700-01 PDS r

r --

r, r-

--.,--v----.,...,-,-,-+-..----_-----

--,,..w

-,.-3~----,_,-,--.

- -,,,, + ~, -

y

-v

e Appendix 2

Pursuant to the provisions of 10 CFR 2.201, you are required to submit to this of fice within thirty days of the date of this Notice a written state-ment or explanation in reply, including for each item of noncompliance:

(1) corrective action taken and the results achieved; (2) corrective action to be taken to. avoid further noncompliance; and (3) the date when full com-pliance-will be achieved. Consideration may be given to extending your response time for good cause shown.

1 l

4-N2-M-

Dated D. G. Wiedeman, Acting Chief Materials Radiation Protection Section 1 l

i

_ _ _. _ _. _ - _ _ _ - - - - - - - - - - - - - - - - -