Appendix NOTICE OF VIOLATION University of Wisconsin - Superior License No. 48-20005-01 As a result of the inspection conducted on May 11, 1982, and in accordance with the NRC Enforcement Policy, 47 FR 9987 (March 9, 1982), the following violation was identified:
License Condition No. 13 requires that your chromatograph detector contain-ing nickel-63 be tested for leakage and/or contamination at intervals not to exceed six months.
Contrary to this requirement, you failed to test your chromatograph detector containing nickel-63 for contamination or leakage from January 1980 to March 31, 1982, an interval exceeding six months.
This is a Severity Level V violation (Supplement VI).
Pursuant to the provisions of 10 CFR 2.201, you are required to submit to this office within thirty days of the date of this Notice a written state-ment or explanation in reply, including for each item of noncompliance:
(1) corrective action taken and the results achieved; (2) corrective action to be taken to avoid further noncompliance; and (3) the date when full compliance will be achieved.
Consideration may be given to extending your response time for good cause shown.