' APPENDIX A NOTICE OF VIOLATION Teachers Hospital License No. 52-15086-01 As a result of the inspection conducted on January 12, 1982, and in accordance with the Interim Enforcement Policy, 45 FR 66754 (October 7,1980), the following violation was identified.
License Condition 17 requires that licensed material be used in accord-ance with statements and procedures contained in your application dated March 30, 1977 and letter received August 31, 1977.
Your letter received on August 31, 1977 requires that the linearity of your dose calibrator be checked at annual intervals.
Contrary to the above, the linearity of your dose calibrator had not been checked in three years.
This is a Severity Level V Violation (Supplement VII).
Pursuant to the provisions of 10 CFR 2.201, you are hereby required to submit to this office within thirty days of the date of this Notice, a written statement or explanation in reply, including: (1) admission or denial of the alleged viola-tion; (2) the reasons for the violation if admitted; (3) the corrective steps which have been taken and the results achieved; (4) corrective steps which will be taken to avoid further violations; and (5) the date when full compliance will be achieved. Consideration may be given to extending your response time for good cause shown. Under the authority of Section 182 of the Atomic Energy Act of 1954, as amended, this response shall be submitted under oath or affirmation.
l l
FEB 0 31982 Date:
8204140361 820407 NMS LIC30 52-15086-01 PDR
' -