Discusses Rept of Unsatisfactory Performance Testing. Findings & Supporting Documents by Smithkline Beecham Clinical Labs & Elsohly Labs Encl.Encl Withheld
ML20045E731
Person / Time
Site: Farley  
Issue date: 06/28/1993
From: Woodard J
SOUTHERN NUCLEAR OPERATING CO.
To:
NRC OFFICE OF INFORMATION RESOURCES MANAGEMENT (IRM)
Shared Package
ML19303F677 List:
References
NUDOCS 9307020366
Download: ML20045E731 (2)


Text

Southern Nciear Operating Company Post CMfice Box 1295 Birmingham, Alabama 352011295 Telephone 205 868-5000 k

L Southern Nudear Operating Company the s othemtectnc sprem June 28, 1993 Docket Nos.

50-348 50-364 U.S. Nuclear Regulatory Commission Attn.: Document Control Desk Washington, DC 20555 Joseph M. Farley Nuclear Plant Report of Unsatisfactory Performance Testina Gentlemen:

In accordance with 10 CFR 26 Appendix A, paragraph 2.8, Southern Nuclear Operating Company (Southern Nuclear) requires that blind performance test specimens be submitted to the HHS-certified laboratory used to-confirm the presumptive positive on-site screens. On June 2, 1993 a known blind positive sample was forwarded from the Southern Nuclear screening facility at the Joseph M. Farley Nuclear Plant to SmithKline Beecham Clinical Laboratories (SKBL) located in Atlanta, Georgia for confirmation testing.

On June 4,1993 the sample was reported as negative by the Medical Review Officer. An investigation of the discrepancy was begun immediately upon receipt of the Medical Review Officer report.

The subsequent investigation determined a suppression in discrimination above the cut-off for the THC-100 assay used by SKBL which was isolated to channel one on the SKBL Olympus screening instrument. Corrective actions by SKBL include switching channels on the Olympus machine and requesting that the manufacturer repair the malfunctioning channel.

Enclosed are the findings and supporting documents by SKBL and ElSohly Laboratories.

This letter is considered to satisfy the reporting requirements of 10 CFR Part 26 appendix A, paragraph 2.8.

Respectfully submitted, 020045 d,mL4 J. D dard (Exe ve Vice President JDW/JMG Enclosures o

/