Notice of Violation from Insp on 820310.Noncompliance Noted: Failure to Perform Wipe Tests Between 810807-0904,1116-30, 820219-0305,nor During Several Other Periods in Excess of Weekly Requirement
ML20042C468
Person / Time
Issue date: 03/24/1982
From: Sreniawski D
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
To:
Shared Package
ML20042C467 List:
References
NUDOCS 8203310426
Download: ML20042C468 (2)


Text

___

Appendix NOTICE OF VIOLATION St. Thomas Hospital Medical Center License No. 34-03329-02 As a result of the inspection conducted on March 10, 1982, and in accord-ance with the Interim Enforcement Policy, 45 FR 66754 (October 7, 1980),

the following violations were identified:

1.

License Condition No. 18 states that the licensee shall possess and use licensed material in accordance with tatements, representations and procedures listed in applications dated September 21, 1977, November 28, 1978, December 13, 1979, and September 10, 1981.

The referenced applications state that weekly wipe tests (for removable contamination) will be performed in your nuclear medicine department.

Contrary to the above, no wipe tests were performed between August 7, and September 4, 1981; November 16 and 30, 1981; February 19 and March 5, 1982; nor during several other periods in excess of the weekly requirement.

This is a Severity Level V violation (Supplement VII).

2.

License Condition No. 18 states that the licensee shall possess and use licensed material in accordance with statements, representations and procedures listed in applications dated September 21, 1977, November 28, 1978, December 13, 1979, and September 10, 1981.

The referenced applications state t. hat quality control tests of your dose calibrator will include daily constency checks.

Contrary to the above, no constency checks were performed between July 10 and 15, 1981; July 15 and 20, 1981; May 21 and 27, 1981; nor on February 4, 16, 17, and 19, 1982; nor on several other occassions when byproduct materials were administered.

This is a Severity Level V violation (Supplement VII).

Pursuant to the provisions of 10 CFR 2.201, you are required to submit to this office within thirty days of the date of this Notice a written state-ment or explanation in reply, including for each item of noncompliance:

8203310426 B20324 NMS LIC3K) 34-03329-02 PDR

'e Appendix 2

(1) corrective action taken and the results achieved; (2) corrective action to be taken to avoid further noncompliance; and (3) the date when full com-pliance will be achieved. Under the authority of Section 182 of the Atomic Energy Act of 1954, as amended, this response shall be submitted under oath

- or affirmation. Consideration may be given to extending your response time for good cause shown.

5DAf8h f

D. J SveniaQ' ski, Chief Dated [

j Mate is.s Radiation Protection Sect on 2

+