Notice of Violation from Insp on 820225-26.Noncompliance Noted:No Thyroid Checks Performed on Personnel Handling Quantities of Radioiodine & No Lab Surveys Performed in Applicable Labs at Required Frequencies
ML20041F736
Person / Time
Issue date: 03/11/1982
From: Sreniawski D
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
To:
Shared Package
ML20041F733 List:
References
NUDOCS 8203170365
Download: ML20041F736 (2)


Text

.

e Appendix NOTICE OF VIOLATION Eli Lilly and Company License No. 13-01133-02 As a result of the inspection conducted on February 25 and 26, 1982, and in accordance with the Interim Enforcement Policy, 45 FR 66754 (October 7, 1980), the following violations were identified:

1.

License Condition No. 20 requires that licensed material be possessed and used in accordance with statements, representations, and proce-dures contained in.various dated applications and letters.

Eli Lilly radiation safety manual, page 9, submitted with application dated October 28, 1977, states that personnel performing radioiodine labeling who handle more than 10 millicuries per week will have weekly or biweekly thyroid checks.

Radiation safety manual, page 9, submitted with application dated August 3, 1978, states that individuals handling 4

500 microcuries of radioiodine for labeling or inorganic radioiodine salts must have a thyroid check every two weeks.

Contrary to these requirements, personnel handling quantities of radioiodine exceeding one or both of the above action levels, have not had thyroid checks every two weeks.

Specifically, an individual in laboratory N-22 did not have thyroid checks performed in March thru June 1979, and June thru September 1980.

This is a Severity Level V violation (Supplement VII).

2.

License Condition No. 20 requires that licensed material be possessed and used in accordance with statements, representations, and proce-dures contained in various dated applications and letters.

Radiation safety manual, pages 33-36, submitted with application dated August 3, 1978, requires. laboratory surveys to be performed at frequencies dependant on the radioisotope, its quantity and use.

i Contrary to this requirement, laboratory surveys were not performed j

l in all applicable labs at the required frequencies. Specifically, monthly surveys were not performed in R. J. Herberg's lab (s) at your Greenfield facility between June 20, 1980 and January 7,1981, nor i

in J. R. Hutchins' lab at the McCarty Street facility between October f

31, 1980 and March 9, 1981.

This is a Severity Level V violation (Supplement VII).

i l

l 0203170365 820311 b#fS LIC30 13-01133-02 PDR

c Appendix Pursuant to the provisions of 10 CFR 2.201, you'are required to. submit to this office within thirty days of the date of this Notice a written statement or explanation in reply, including for each item of noncompliance:

(1) corrective action taken and the results achieved; (2) corrective action to be taken to avoid further noncompliance; and (3) the date when full compliance will be achieved. Under the authority of Section 182 of the Atomic Energy Act of 1954, as amended, this response shall be submitted under oath or affirmation. Consideration may be given to extending your response time for good cause shown.

3-M*E V

.::! =

ma Dated D. J. Sreniawski, Chief

?*

Materials Radiation Protection Section 2 1