Text
Ref:
SA/ LAB k
e afa, UNITED STATES g
NUCLEAR REGULATORY COMMISSION
/
-[ 'h - )[# -
[
WASHINGTON. D. C. 20555 y.
\\,l'...~*fl b. '?,
f Qiji 2 0 gggg i
?
lir. Dayne H. Brown, Chief D'
~c ' >N,.
O 10,
9g
'Ag Radiation Protection Section s
'M j,
,s Division of Facility Services
/h'7 %s I
Box 12200 Raleigh,liorth Carolina 27605
Dear fir. Brown:
This is to confirm the comments made to you and fir. Cecil Brown by Mr. Lloyd Bolling at the conclusion of our recent radiation control program review.
Several deficiencies of a serious nature were noted in our letter to Dr. florrow. A copy of this letter is enclosed for your information.
Enclosed also are specific comments and recommendations based on our recent regulatory program review.
I would appreciate your review of our recommendations and woul. like to receive a response from you.
I appreciate the courtesy and cooperation extended to ilr. Bolling during the review.
Sincerely,
.n s
34 -
onald A. liussbaumer Assistant Director for State Agreements Program Office of State Programs
Enclosure:
As stated cc:
J. P. O'Reilly, Administrator, NRC Region II Sarah T. Ilorrow,l1.D NRC Public Document Room, w/ encl.
State Public Document Room, w/ encl.
8202100002 820120 PDR STPRG ESGNC FDR
~
Commints and Recomm:ndations on ths North Carolina Radiation Control Program I.
Licensing Licensing Procedures is a Category II Indicator. The following minor deficiency was noted.
A.
Comment A review of selected license files indicates that in at least J
two cases medical applications lacked adequate procedures and equipment to perform quality checks on dose calibrators.
It was noted also that a similar comment was made in our letter to you following our 1980 program review.
Recommendation We recommend that each medical applicant submit step-by-step procedures for use and performance testing of dose calibrators.
l i
Each applicant should also be required to have a set of check / calibration sources of appropriate strengths and energies.
II. Compliance Independent Measurements is a Category II Indicator. The following minor deficiency was noted.
A.
Comment A review of selected inspection reports indicates that in at least five cases there was a lack of adequate documentation of f
independent measurements.
Recommendation We recommend that independent measurements (i.e., wipe samples) be performed, analyzed and documented for each compliance inspection, when applicable. We feel that these confirmatory measurements should be an integral part of ti': inspection report.
L I