.Efstathiss G. Metropoulos, M.D.
License'No. 21-12972-01 4
As a result of the inspection conducted on December 16, 1981, and in ac-I cordance with the Interim Enforcement' Policy, 45 FR 66754 (October 7, 1980),
l the following violations were identified:
License. Condition 14 requires the licensee to possess and use licensed material'in accordanu with statements, representations,'and procedures 1
contained in application dated September 1, 1978.
I 1.
In the referenced application, the licensee indicates that the pro-cedures described in Appendix D Section 2 will be used for calibration of the dose calibrator. Appendix D Section 2 (of NUREG-0338, Revision 1, dated November 1, 1977) requires that the dose calibrator linearity be checked at installation and quarterly thereafter.
Contrary to this requirement, the licensee failed to check the dose calibrator linearity from August 30, 1979 to July 31, 1981 and no check after July 31, 1981, periods in excess of a calendar quarter.
This is a Severity Level V violation (Supplement VII).
i 2.
. Appendix D Section 2 requires that the licensee measure and record the l
activity of at least one reference source on a daily basis or before each use of the dose calibrator.
i Contrary to this requirement, the licensee failed to record the measured i
activity of their reference source on November 13, 16 and 23, 1981 and i
j December 11, 1981, and other dates during which licensed material was administered to patients.
This is a Severity Level VI violation (Supplement VII).
l 3.
10 CFR 20.401(a) requires each licensee to maintain records showing the radiation exposures of all individuals for whom personnel monitoring is' required and such records be kept on Form NRC-5 or contain all the information required by that form.
Contrary to this requirement, the licensee's radiation exposure records
- do not include the date of birth and social security number for all l
fndividuals for whom personnel monitoring is required.
This is a Severity Level VI violation (Supplement IV).
t.
J l
4 92020103S1 820122 NMS LIC30 21-12972-01 PDR
N Appendix I Pursuant to the provisions of-10 CFR 2.201, you are required to submit to
.this office within thirty days of the date of this Notice a written state-ment or explanation in reply, including for each item of noncompliance:
(1) corrective action taken and the results achieved; (2) corrective action to be taken to avoid further noncompliance; and (3) the date when full com-pliance will be achieved. Under the authority 'of Section 182 of the Atomic Energy Act of 1954, as amended, this response shall be submitted under oath or affirmation. Consideration may be given to extending your response time for good cause shown.
)
r/gijg3 Datel
/
D. J.
Sreniawski, Chief Ma(er als Radiation Protection She ion 2