Notice of Violation from Insp on 810928.Noncompliance Noted: in Absence of Area Monitor,Personnel Entering Teletherapy Room Following Irradiation Have Not Been Determining Condition of Beam W/Required Survey Instrumentation
ML20031H413
Person / Time
Issue date: 10/15/1981
From: Wiedeman D
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
To:
Shared Package
ML20031H406 List:
References
NUDOCS 8110270461
Download: ML20031H413 (2)


Text

-

k.

-Appendix A NOTICE OF VIOLATION Luther Hosptial License No. 48-02122-04 As a result of the inspection conducted on September 28,-1981,.and in accordance with the Interim Enforcement Policy, 45 FR 66754 (October 7, 1980), the following violations were identified:

1.

The order modifying teletherapy licenses dated May 7, 1980,. states that whenever your area monitor is not operational, any person entering the teletherapy room following an irradiation shall enter with an operable and calibrated radiation survey meter to determine the beam condition.

Contrary to the above requirement, it was determined through state-ments by licensee representatives that this requirement is not being met.

Specifically, a licensee representative stated the Primalert l'" Model 05-433 was sent back to the manufacturer for repairs'on 0

September 8, 1981.

In the absence of the area monitor, personnel entering the teletherapy room following irradiation have not been determining the condition of the beam with the required survey instrumentation.

This is a Severity Level V violation (Supplement VII).

2.

10 CFR 35.25(a) and (c) requires the licensee-to maintain for.

inspection by the Commission records of calibration-of the instru--

ments used to make full calibrations and records of the licensee's evaluation of the qualified experts training and experience.

Contrary to the r.bove requirement, it was determined through state-ments by licensee representatives that this condition is not being met.

Specifically, on the day of this inspection, the above records were not maintained for inspection.

This is a Severity Level VI violation (Supplement VII).

Pursuant to the provisions of 10 CFR 2.201, you are required to submit j

to this office within thirty days of the date of this Notice a written 4

I l

t 8110270461 811'015 NMS !.IC30 l

48-02122-04

~PDR.

Appendix A statement or explanation in reply, including for each item of noncom-pliance:

'l) corrective action taken and the results achieved; (2) cor-rective

. tion to be taken to avoid further noncompliance; and (3) the date wher: full compliance will be achieved. Under the authority of Section 132 of the Atomic Energy Act of 1954, as amended, this response shall be submitted under oath or affirmation. Consideration may be given to extending your response time for good cause shown.

Dated )D/) 5/ S/

b.

UAly&v!

D. G. Wiedeman, Acting" Chief' Materials Radiation Protection Section 1