UNITED CONTROLS INTERNATIONAL 10CFR21 NOTIFICATION LETTER January 13, 2020 U.S. Nuclear Regulatory Commission Document Control Desk Washington D.C. 20555-0001

Subject:

10CFR21 Notification for Ashcroft 2-1/2" & 3-1/2" 1009, 63mm 2008, 63mm &

100mm 1008 Industrial Gauges, Weksler 3-1/2" Regal Gauges The purpose of this letter is to provide the NRC with a report in general conformity of the requirements of 1 OCFR Part 21.21. On December 27, 2019 United Controls International (UCI) received a Product Advisory Letter from Ashcroft Inc, dated December 2, 2019 (attached) informing of a potential error in the tip material used on Ashcroft 2-1/2" and 3-1/2" model 1009, 63mm model 2008, 63mm and 100mm model 1008 industrial gauges along with Weksler 3-1/2" regal gauges. Refer to attached Ashcroft Inc. Product Advisory Letter dated December 2, 2019 for specific configuration and model numbers affected. Due to a supplier issue, 304SS tip material may have been supplied on the subject gauges instead of the specified 316SS material.

In certain applications, the 304SS tip material may be subjected to accelerated oxidation or degradation, resulting in potential release of process media. However, no detrimental effects are expected in non-corrosive applications (i.e. water with neutral, oils/hydraulic fluids, non-corrosive gasses such as air and nitrogen, anhydrous ammonia and ammonia refrigerant).

Required information as per 10CFR Part 21.2l(d)(4) follows:

(i)

Name and Address of the individual or individuals informing the Commission Rob.Hale (ii)

President United Controls International 205 Scientific Drive Norcross, GA 30092 Identification of the facility, the activity, or the basic component supplied/or such facility or such activity within the United States which fails to comply or contains a defect.

Ashcroft 2-1/2" and 3-1/2" model 1009, 63mm model 2008, 63mm and 100mm model 1008 industrial gauges along with Weksler 3-1/2" regal gauges. Refer to attached Ashcroft Inc. Product Advisory Letter dated December 2, 2019 for specific configuration and model numbers affected.

Page 1 of 8

UNITED CONTROLS INTERNATIONAL (iii)

Identification of the firm constructing the facility or supplying the basic component which fails to comply or contains a defect.

United Controls International 205 Scientific Drive Norcross, GA 30092 (iv)

Nature of the defect or failure to comply and the safety hazard which is created or could be created by such defect or failure to comply.

(v)

Due to a supplier issue, 304SS tip material may have been supplied on the subject gauges instead of the specified 316SS material. In certain applications the 304SS tip material may be subjected to accelerated oxidation or degradation, resulting in potential release of process media. However, no detrimental effects are expected in non-corrosive applications (i.e. water with neutral, oils/hydraulic fluids, non-corrosive gasses such as air and nitrogen, anhydrous ammonia and ammonia refrigerant).

The date on which the information of such defect or failure to comply,was obtained.

December 27, 2019 (vi)

In the case of a basic component which contains a defect or fails to comply, the number and location of these components in use at, supplied/or, being supplied for, or may be supplied/or, manufactured or being manufactured/or one or more facilities or activities subject to the regulations in this Part.

Part Number Utility Purchase Order Quantity*

2510095\\NL01L0400 Georgia Power -

SNG27679-0694 3

Hatch 101008502B160#/KP Alabama Power -

SNA27679-0585 2

Farley Page 2 of 8

UNITED CONTROLS INTERNATIONAL (vii)

The corrective action which has been, is being, or will be taken; the name of the individual or organization responsible for the action; and the length of time that has been or will be taken to complete the action.

The manufacturer of the subject items (Ashcroft Inc.) has identified that the supplier material issues affect only the identified items manufactured from October 2017 through October 2019. Utilities utilizing affected items are advised to evaluate whether the items would be subject to accelerated oxidation or degradation in the installed environment as described in the Ashcroft Inc. Product Advisory Notice dated December 2, 2019 (attached). If it is determined that accelerated oxidation or degradation applies to the items then they are to be returned for replacement with items manufactured with the correct 316SS tip material.

UCI shall implement a corrective action implementing the following:

The utilities that were supplied with affected items by UCI have been notified of this issue. Items supplied to Georgia Power - Hatch were not installed and,are being returned for replacement. UCI is awaiting evaluation by Alabama Power --

Farley as to whether replacement of items is required.

The dedication plan for supplied items includes material verification of wetted components including the bourdon tube and process connections; however, verification of the bourdon tube tip material was not included. The dedication plans for the subject items shall be revised to include verification of tip material as 3 l 6SS for all future orders along with inspection of manufacture date to ensure that items manufactured between October 2017 and October 2019 are not

  • supplied.

Training of dedication personnel regarding this issue and identification of wetted components for material verification during dedication.

(vii!)

Any Advice related to the defect or failure to comply about the facility, activity, or basic component that has been, is being, or will be given to purchasers or licensees.

The manufacturer of the subject items (Ashcroft Inc.) has identified that the supplier material issues affect only the identified items manufactured from October 2017 through October 2019. Utilities utilizing affected items are advised to evaluate whether the items would be subject to accelerated oxidation or degradation in the installed environment as described in the Ashcroft Inc. Product Advisory Notice dated December 2, 2019 (attached). If it is determined that accelerated oxidation or degradation applies to the items then they are to be returned for replacement with items manufactured with the correct 316SS tip material.

Page 3 of 8

UNITED CONTROLS INTERNATIONAL (ix)

In the case of an early site permit, the entities to whom an early site permit was transferred.

Not Applicable.

If you have any questions or wish to discuss this matter or this report, please contact:

Wesley Hickle Engineering Manager whickle@unitedcontrols.com 770-496-1406 X 165 Sincerely,

~~\\\\M Rob Hale President United Controls International Page 4 of 8

UNITED CONTROLS INTERNATIONAL

Attachment:

Ashcroft Product Advisory Letter

~ ASHCROFT~

Trust the shield~

December 2, 2019 PRODUCT ADVISORY LETTER Ashcroft* 2Yz" and 3Yz" 1009, 63 mm 2008, 63 mm & 100 mm 1008 Industrial Gauges, Weksler 3Yz" Regal Gauges:

Ashcroft* 2510095W industrial gauges in ranges of Vacuum, Compound, lSpsi through 600psi with date codes Oct 2017 to Oct 2019. (Excludes 251009AW Gauges)

Ashcroft* 351009SW industrial gauges in ranges of Vacuum, Compound, lSpsi through 600psi with date codes Oct 2017 to Oct 2019. (Excludes 351009AW Gauges)

Ashcroft* 6320085 industrial gauges in ranges of Vacuum, Compound, lSpsi through 600psi with date codes Oct 2017 to Oct 2019.

Ashcroft* 631008S industrial gauges in ranges of Vacuum, Compound, lSpsi through 600psi with date codes Oct 2017 to Oct 2019. (Excludes Center Back gauges)

Ashcroft* 1010085 industrial gauges in ranges of Vacuum, Compound, lSpsi through 600psi with date codes Oct 2017 to Oct 2019. (Excludes Center Back gauges)

Weksler* BA43, BS43, BY43 Regal industrial gauges in ranges of Vacuum, Compound, lSpsi through 600psi with date codes Oct 2017 to Oct 2019.

Weksler* BM43 & BN43 Regal Altitude Gauges with date codes Oct 2017 to Oct 2019.

Weksler* BP43 & BQ43 Regal Refrigerant and Ammonia Gauges with date codes Oct 2017 to Oct 2019.

(See figure 3 for visual identification of gauge models and figure 4 for other pressure units)

This letter is to alert you to a potential error in the tip material used on certain models of Ashcroft*

Model 1008, 2008 and 1009 pressure gauges. We have determined that due to a supplier issue, 304SS tip material may have been supplied on some 1008, 2008 and 1009 gauges instead of the specified 3165S material. The gauges in question were shipped from October of 2017 through October 2019.

(PLEASE NOTE: Bourdon tube material for all gauges was supplied with the correct 3165S material.)

In certain applications the 304SS tip material (see figure 1 below) may be subject to accelerated oxidation or degradation, resulting in the potential release of process media. However, no detrimental effects are expected in non-corrosive applications (i.e. water with neutral, oils/hydraulic fluids, non-corrosive gasses such as air and nitrogen, anhydrous ammonia and ammonia refrigerant). For additional corrosive information, please refer to our Ashcroft Corrosion Guide.

Bourdon Tube L_ TipD~

_j Wetted area

  • .. 0-~0;--W) r r

~,..

Figure 1 - C-type Bourdon Tube Assembly with Tip Details Page 5 of 8

UNITED CONTROLS INTERNATIONAL

~ ASHCROFT.

Trust the shield~

The date code can be found on a label on the back of the gauge with the month and year of manufacture or the gauge box (see figure 2).

Date Code L

UFEe GA n-100I-SWL-02L-301MV&

133HT48 llllllllllllf&IUID 30165394-18 11/tfTnl

& WARNING - C.C., _. Rep,oduclwa Hamt--.HGW.,..,,.a.;,,.

Figure 2 -Date Code Label Ashcroft is committed to the satisfaction of our customers and to the performance of our products. We apologize for any inconvenience this may cause you.

These pressure gauges can be returned for replacement. Please contact your Ashcroft supplier or go to https:ljsales.ashcroftcom/rma-form to obtain an RMA and information on how ship the gauges back to Ashcroft.

If you have technical questions regarding your specific application and potential corrosion issues, please contact:

Jeremy M. Payne Materials and Test Engineering Manager Ashcroft Inc.

250 East Main St.

Stratford, CT 06614-5145, USA Phone: +1-203-385-0564 Mobile: +1-203-767-6236 E-mail: jeremy.payne@ashcroft.com Ob nf Page 6 of 8

UNITED CONTROLS INTERNATIONAL

~ ASHCROFT.

Trust the shield~

2008 printed on the Dial 63 mm 2008 Gauges 1008 printed on the Dial DURALi FE printed on the Dial 2\\/2"' 1009SW or 3!/2"' 1009SW Gauges Lower Back lncudes both lower and lower back gauges 2 ** 1o09AWor 3'h'" 1009AW Gauges are not part of the product advisory Center Back 1008 Center back gauges are not part of the product advisory 63 mm 1008 and 100 mm 1008 Gauges not Including Center back Gauges Figure 3 - Visual Identification of 1008, 2008 and 1009 gauges 1, tf rj r ft C 1

WW w

Page 7 of 8

UNITED CONTROLS INTERNATIONAL

~ ASHCROFT.

Trust the shield~

.,,r t t nq r,.

t-psi bar kPa Mpa kg/cmz 30IMV

-1BR

-100KP

-.1MP

-tKSC

-1&.68R

-1oo&60KP

-.1&.0tiMP

-1&.6KSC 30IMV&15#

-1&1.5BR

-1oo&150KP

-.1&.15MP

-1&1.5KSC 30IMV&30#

-1&3BR

-1oo&300KP

-.1&.3MP

-1&3KSC 301MV&60t

-1&5BR

-1CIGB60a<P

-. 1'-SMP

-t&.5KSC 30IMV&100#

l&9BR

-loo&OOOKP

-.1&.9MP

-1&9KSC 15#

tBR tooKP

.lMP lKSC 20#

1.613A 1eoKP

.16MP 1.6KSC 30t 2.5BR 250KP

.25MP 2.5KSC 00#

48R 400l<P

.4MP 4KSC 6BR OOOKP

.6MP l'lKSC 1001

120, 1o8R 1000KP 1MP 1oKSC 160I 200#

168A 1600KP 1.6MP l OKSC 300I 258R 2500KP 2.5MP 25KSC 400#

500I 000#

408R 4000KP 4MP 40KSC Figure 4 - Gauge ranges including in the advisory A

ft 11 l

w" Page 8 of 8