Appendix A NOTICE OF VIOLATION Vermillion County Hospital License No. 13-18604-01 As a result of the inspection conducted on September 15, 1981, and in ac-cordance with the Interim Enforcement Policy, 45 FR 66754 (October 7, 1980),

the following violations were identified:

1.

License Condition No. 15 requires that licensed material be possessed and used in accordance with statements, representations, and procedures contained in application dated October 6, 1978.

Item No. 9 ci the above referenced application states that your survey meter will have a measuring range of 0 mR/hr to 2 R/hr.

i.

Contrary to the above requirement, it was determined through an obser-vation of the NRC inspector that this requirement has not been met.

j Specifically, your CDV-700 Model No. 6A survey meter is only capable i

of measuring 0 mR/hr to 50 mR/hr.

J l

This is a Severity Level V violation (Supplement VII).

?

2.

License Condition No. 15 requires that licensed material be possessed and used in accordance with statements, representations, and procedures contained in application dated October 6, 1978.

Item No. 10 of the above referenced application states that quarterly linearity' tests will be performed by assaying the syringe containing technetium-99m at six hours, 24 hours2.777778e-4 days <br />0.00667 hours <br />3.968254e-5 weeks <br />9.132e-6 months <br />, 30 hours3.472222e-4 days <br />0.00833 hours <br />4.960317e-5 weeks <br />1.1415e-5 months <br />, and 48 hours5.555556e-4 days <br />0.0133 hours <br />7.936508e-5 weeks <br />1.8264e-5 months <br />.

Contrary to the above requirement, it was determined through state-t ments by licensee representatives that this condition is not being met.

Specifically, a licensee representative stated lineerity tests performed since August 12, 1980 did not include a six hour, 30 hour3.472222e-4 days <br />0.00833 hours <br />4.960317e-5 weeks <br />1.1415e-5 months <br />, or 48 hour5.555556e-4 days <br />0.0133 hours <br />7.936508e-5 weeks <br />1.8264e-5 months <br /> assay. Furthermore, records of these tests were not available for review.

This is a Severity Level V violation (Supplement VII).

5 S110060322 8109287 NHS LIC30 13-18604-01 PDR

_~

_ = -

o D

Appendix A Pursuant to the provisions of 10 CFR 2.201, you are required to submit to this office within thirty days of the date of this Notice a written state-ment or explanation in reply, including for each item of noncompliance:

(1),

corrective action taken and the results achieved; (2) corrective action to be taken to avoid further noncompliance; and (3) the date when full compliance 5

will be achieved. Under the authority of Section 182 of the Atomic Energy Act of 1954, as amended, this response shall be submitted under oath or affirmation.

Consideration may be given to extending your response time for good cause shown.