Notice of Violation from Insp on 810903.Noncompliance Noted: Last Linearity Check Performed on 810105 & Accuracy Checks Performed Using Only Co-57
ML20010G704
Person / Time
Issue date: 09/18/1981
From: Sreniawski D
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
To:
Shared Package
ML20010G697 List:
References
NUDOCS 8109220430
Download: ML20010G704 (1)


Text

- _ _ - _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

~

Appendix A NOTICE OF VIOLATION l

l Marvin Dean Dziabis, M.D.

License No. 13-12204-01 As a result 9f the inspection conducted on September 3, 1981, and in accordance with the Interim Enforcement Policy, 45 FR 66754 (October 7, 1980), the following violation was identified:

License Condition No. 17 requires that licensed material'be possessed and used in accordance with statements, representations, and procedures contained in application dated January 15, 1979.

The above referenced application states, the procedures deacribed in Appendix D of the draft " Guide for Preparation of Applications for Medical Programs" will be followed. Appendix D requires that your dose calibrator be checked and recorded quarterly for linear 1Ly and the annual accuracy be performed using several radionuclides such as cesium-137,'carium-133, and cobalt-57.

Contrary to the above requirement, it wao determined through statements by license 2 representatives and the NRC inspectors review of records that this condition is not being met.

Specifically, as of the day of this inspection, the last linearity check was performed on January 5, 1981.

In addition, accuracy checks are being performed using only cobalt-57.

This is a Severity Level V violation (Supplement VII).

Parsuant to the provisions of 10 CFR 2.201, you are required to submit to this office within thirty days of the date of this Notice a written statement or explanation in reply, including for each item of noncompliance:

(1) corrective action taken and the results achieved; (2) corrective action to be taken to avoid f-2rther noncompliance; and (3) the date when full compliance will be achieved. Under the authority of Section 182 of the Atomic Energy Act of 1954, as amended, this response shall be submitted under oath or affirmation. Consideration may be given to extending your response time for good cause shown.

l Dated 9 -l b SI D. J. Si en13wski, Chief Materia s Radiation Protection Section 2 P109220430 810918' NMS LIC30 13-12204-01