ML20004C067

From kanterella
Jump to navigation Jump to search
Notice of Violation from Insp on 810512.Noncompliance Noted: 206 Uci Sealed Cs-137 Ref Source Not Leak Tested Since Dec 1978 & Radiation Levels Found Exceeding Allowable Limit in Unrestricted Hallway Adjoining Room 2D50
ML20004C067
Person / Time
Issue date: 05/21/1981
From: Wiedeman D
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
To:
Shared Package
ML20004C062 List:
References
NUDOCS 8106010477
Download: ML20004C067 (2)


Text

O-Appendix A NOTICE OF VIOLATION Hennepin County Medical Center License No. 22-11070-01 As a result.of the inspection conducted on May 12, 1981, and in accordance with the Interim Enforcement Policy, 45 FR 66754 (October 7, 1980), the following violations were identified:

1.

10 CFR 20.105(b) requires that radiation levels in unrestricted areas be limited so that if an individual were continuously present in the area, he could not receive a dose in excess of two millirens in any hour or 100 milliress in any seven consecutive days.

Contrary to this requirement, on the day of the inspection, within an unrestricted hallway adjoining room 2D52, radiation levels existed of such a magnitude that if an individual had been continuously present in the area, he could have received a dose in excess of 100 milliress in any seven consecutive days.

This is a Severity Level V violation (Supplement IV).

2.

10 CFR 20.201(b), " Surveys," requires you to make surveys such as may be necessary for you to comply with all sections of Part 20.

Contrary to this requirement, prior to September 1980, you failed to make such surveys as were necessary to determine that individuals who handled 100 sci quantities of tritium were not exposed to airborne concentrations exceeding the limits specified in 10 CFR 20.103.

Specifically, no urine bioassays were made during the preparation of tritiated sodium borohydride.

This is a Severity Level V violation (Supplement IV).

3.

10 CFR 34.14(e)(1)(i) requires that sealed calibration or reference sources possessed pursuant to.10 CFR 35.14(d) be tested for leakage and/or contamination at intervals not to exceed six months.

Contrary to this requirement, from December 1978, to the day of the inspection, you have failed to leak test your 206 microcurie sealed cesium-137 reference source.

This is a Severity Level V violation (Supplement VII).

81060/o_ %

Appendix A Pursuant to the provisions of 10 CFR 2.201, you are required to submit to this office within twenty-five days of the date of this Notice a written statement or explanation in reply, including for each item of noncompliance:

(1) corrective action taken and the results achieved; (2) corrective action to be taken to avoid further noncompliance; and (3) the date when full com-pliance will be schieved. Under the authority of Saction 182 of the Atomic Energy Act of 1954, as amended, this response shall be submitted under oath or affirmation.

S-A /-F a*L = =-

Dated D. G. Wiedeman, Acting Chief Materials Radiation Protection Section 1 e

m

,w g

,y.

.y

_,,--g

- -