Responds to NRC Re Violations Noted in IE Insp of Licenses 21-12826-01 & SNM-1777 on 810224.Corrective Actions:New Procedures Instituted to Assure Semiannual Performance of Sealed Calibr Leak Tests
ML20004A039
Person / Time
Site: 07002707
Issue date: 04/06/1981
From: Kreiger D
INGHAM MEDICAL CENTER, LANSING, MI
To: Sreniawski D
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION III)
Shared Package
ML20004A034 List:
References
NUDOCS 8105110605
Download: ML20004A039 (5)


Text

,

401 WEST GREENLAWN, LANSING, MICH1GAN 48909 (517) 374-2261 M

INGHAM MEDICAL' CENTER JOHN C. SODERHoLM, MHA Executive Vico. President April 6, 1981

'D. J. Sreniawski Acting Chief, Materials. Radiation Protection Section 2 United States Nuclear Regulatory Commission Region III 799 Roosevelt Road Glen Ellyn, Illinois 60137

Dear Mr. Sreniawski:

This correspondence is provided in response to your letter of March 16, 1981, concerning the Notice of Violation resulting from the safety inspection survey conducted at Ingham Medical Center on February 24, 1981.

Two Severity Level V (Supplement VII) violations were identified.

In accordance with the requirements of 10 CFR 2.201, and for your convet tence, the specific violations are addressed below:

1.

10 CFR 35.14 (e)(1)(1) requires that sealed calibration or reference sources possessed pursuant to 10 CFR 35.14(d) be tested'for leakage and/or contamination at intervals not to exceed six months.

Response - Contrary to this requirement, the last documented Cs-137 calibration source leak test was performed on November 24, 1976.

A wipe test was performed on March 27, 1981 with the following result:

114 cpm background, 106 cpm wipe with a difference of 8 cpm.

To assure compliar.ce with 10 CFR 35.14, Ingham Medical Center procedure 15.14.34 (cop, appended) has been approved and implemented.

2.

Condition No. 16 of your license states the dose calibrator shall be calibrated in accordance with the procedures described in Appendix D, Section 2, NUREG-0338, Revision 1, dated November 1, 1977.

Appendix D, Section 2 requires a quarterly linearity calibration of the dose calibrater.

Responne - Contrary to this requirement, the last documented calibration was performed on May 5, 1980.

1a an offort to rectify this violation, t if ncarity test was perf ormed on March 30, 1981 (documentation appended).

Additionally, to insure continued compliance with Condition No. 16, Inghan Medical Center Procedure 15.14.35 (copy appended) has been approved and implemented.

GR 1 ".) 1981 810s n oboS

=~*

INGHAM MEDICAL CENTER The raw' data and calculations for both the Cs-137 leak test and linearity test is being reviewed currently by the Physicist, with a final written report expected prior to May 1, 1981.

By my signature, I hereby certify that the above info'rmation is true and to the best of my knowledge and belief.

correct Sincerely, INGHAM MEDICAL CE::TER

/

/.c.

.,L M

David A. Krei,c[

Assistant Vic resident DAK/kb enc 1:

cc:

Mr. Henry Orth Signed before me this 7th day of April, 1981

  • (\\.' n b.

~aw~--

M. Kathleen Bonneau Nota ry Public, Ingham County Commission Expires Ety 18, 1983 i

u:

p.

~

O l15.14.35 t

. RADIOLOGY -

  • NUCLEhR HEDICINE ;

PROCEDURE

. Ingham Me' dical Centur.

.sua,eer; DOSE CALIsRArOR CALIaRArION a

1

'1 PAGE -

. OF 5LL' NUCLEAR PERSONNEL SCOPE:

I v

ISSUED SY:

H. Orth, R. T.* (R) i

>}

EFFECTIVE:

REvJWED:

REVISED:

APPAOVED Bv:

m \\\\n 9l

\\

~

-2SMAYB1 w JJ D. Kreige -,h

~

1 1

/

3 i

t

Purpose:

To be' in compliance with condition #16 of the IMC license with the NRC which states the dose calibrator'shall be calibrated {

in accordance with the' procedures described in Appendix D, t

Section 2 NNUREG-0338, Revision 1, dated Novembcr 1,1977.

i I.

To assure accurato operation of the dose calibrator, the following t

method will be implemented.

3 A.

Test for the following:

1.

Instrument constancy (da'ily) 2.

Instrument accuracy (at installation and annually thereafter). l 3.

Instrument linearity (at installation and quarterly thereafter)i 4

Geometrical variation (at installation)

Y t

B.

The actual testing of the doso calibrator will be performed i

ht; the radiation physicist and th. supervisor as outlined in Appendix D, method for calibration of dose calibrator section

}

2, procedure 10.8-27 in the U.S. Nuclear Regulatory Commission Regulatory Guide Offico of Stsndard Development.

3 i

II.

Documentation of the dose calibrator Linearity test will be kept on file with the Manager of the Department of Radiology

't f

/

f f

DISTRIBUTION:

l - President 1 - Vice-President 1 - Archivos 1 - R.3diology 1 - Nuc1 car Mcdicina J

001-1

)

J l

^

i 15.14.34 i

j

  1. ' c Iitgham' Medical Center 2

PROCEDURE RAMON - NUCLEAR MENCJ2E.

SueuC1*

Cs-137 CALIBRATION SGUT'JE LEAK TEST '- CONTAM!?lATION Pron 1

cy 1

+

SCOPE:

ALL NUCLEL' PERSONNEL 4

t ISSUED BY:

M r

H. Or.,t,h, R. 'C. (R) y EFFECTIVE:

REVIEWED:

REVISED.

APP

{

.e

  • " ~. ~

^

\\v J

.,, a

_1 L,,

25 MAR 81 w

. Kr g PURPOSE: To be in compliance with NRC regulation' 10 CFR 35.14 which requires that scaled' calibra tion or reference sources -

possessed pursuant to 10CFR 35.1Q (d) be tested for loakage and/or contamination at intervals not to exceed l

six months.

I.

The 1 mci Cs-137 calibration source leak test will be performed quarterly.

e i

The test for leakage and/or contamination will be performed A.

by our radiation physicist.

B.'

Documentation of the leakage and/or contamination test will be kept on file-with the Manager of the Department of Radiology

\\

l i

f h

r DISTRIBUTION:

I 1 - Prcsident 1 - Vice-Prcsident 1 - Archives 1 - Radiology 1 - Nuc1 car Mcdicina l

f 001-1

'[J

\\

.,m.w x c 1

p.r.;

.,2.

s

_... f _ _.t_ _ _........

i

... L-t.........-

t

.l....

1

-l=

c

.. ~....... _,

i.

.._.t..

t!._ _

.....L_._.,.

_l I

T-- -- -

DOSECALIBRf0RLIN$ARITYTEST-~

l l

j..

t l'

i I

n, IFacil_ity:

U ^ I" b

'5/

.;._ p.j j u 2

._:.y

.._:.4...

l r

i 1.

..l l

.l

. a.....jDose: Calibrator: h0R(M~/ i

.l.

l

..J. _. 1.

3... _.f

.H

}

1 t_.

_.._...a....._t

.Date:

0,M.

Radionuclide: g?*1% !

t

... (.

4, g hl WSsh.

I Performed By:

_. t

.......a

- j j

~ l i

1,!

i.

.i.

s.

1 I.

2 t

e

.I i

I e

.'l.

t p

i.

i.'

t

.i 1

t I

3....

Activit'y-

_ f.. 7..__t

" -- - 1date -

' time

- i i

i i

hrs.

Activity (mci)

(mci).

i

'N), ?/..

8CO g

_. ().

(300..

7 s.

u.a 4

.=

p"

a. - s. -.'

.. 3.

. p ]... ?A ct._.., ii l

r.

E I

i.

3,31el ' P (w

. 0..,...

.-.g.-........q...,

J e

e l

g

1. _...I '

..h-

..... i h. O l f tjnr b.0. _

$ DD I

7

-.\\ i (O

r.

h I

l

.i

..j._...,!_.

i s.

.y t

.j

k.,i..

i l

j

')

l t.

._...i.

t n

.t

\\

'l j

(f) *-,

e t

M,:

  • y-t.

a.

4,

\\

)

r N

I

\\

I I.

1 l

l

'b I

r

\\ !

o..

l' l

+

i 1

.i s 6.,

....i..

c,

\\ )l f

I i

s l

i-i i

y i _. t J

'j ; j

,;...... ]..... -... p,

.t.

.2,.

- l...

4.......,_. s l

..i....

1

.c,.,

i l

.. t... -. n,.. _ _ _

u i

I h

}

.-l E

.. s.

g r.-

s.

.-. I s

)

s.

g, e

4 6

j I

\\

/.M.,,

.3 l

i

l l

I I'

.[

.6.._..

4

+

9 1

{

t e

e n

I 6

6 t

l l

{

I h !.

g l

6 r

i

/ t i

(D k!

N

[

6 Y

I 4

e i

f f

3 j

i h

l.

.l l

.i i

f

. h.

I