ML19320B657

From kanterella
Jump to navigation Jump to search
Forwards Comments & Recommendations on State of Nc Radiation Control Program
ML19320B657
Person / Time
Issue date: 06/27/1980
From: Kerr G
NRC OFFICE OF STATE PROGRAMS (OSP)
To: David Brown
NORTH CAROLINA, STATE OF
References
NUDOCS 8007140569
Download: ML19320B657 (3)


Text

s&* %.

l

7i g

UNITED S TATEs 7

p, NUCLEAR REGULa FORY, COMMISSION 3

E' WASHINGTON, D. C. 20555 e

x.....)

JUN 271980 Ref: SA/ LAB Mr. Dayne H. Brown, Chief Radiation Protection Section Division of Facility Services Box 12200 Raleigh, North Carolina 27605

Dear Mr. Brown:

This is to confirm the comments made to you and Mr. Cecil Brown by Lloyd Bolling at the conclusion of our recent radiation control program review.

Based on the results of our review, the staff believes that the North Carolina program for control of agreement materials is adequate to protect the public health and safety apd is compatible with the NRC's program.

Specific comments and recomendationf are enclosed.

I would appreciate your review of our recomendations and receiving your comments on them...

The State's efforts in responding swiftly and adequately to incidents involving radioactive materials are commendable.

I appreciate the courtesy and cooperatica extended to Lloyd Bolling during the review.

Sincerely, U s eth v)

G. Wayne Kerr, As.4stant Director for State Agreements Program Office of State Programs

Enclosure:

As stated cc: Sarah T. Morrow, M.D.

NRC Public Document Room, w/ encl.

State Public Document Room, w/ encl.

80 0 7140 SM j

COMMENTS AND RECOMMENDATIONS ON THE NORTH CAROLINA lifDTfTION CONTROL PROGRAM I.

Licensing A.

Comment A review of selected license files showed that at least two medical licenses lacked adequate discussion of their procedures to check the performance of their dose calibrators.

Recommendation We recommend that applicants for medical licenses be required to submit step-by-step procedures for checking the performance of dose calibrators.

Applicants may submit the name of an approved consultant who will perform quarterly and yearly accuracy checks. The named consultant must also have step-by-step procedures on file with your agency. A sample of acceptable procedures is enclosed.

B.

Comment 4

A review of selected license files indicated that at least three licenses lacked sufficient discussion of procedures for the receipt of radioactive material.

Recommendation We recommend that each applicant be required to adequately address j

procedures for receipt and safe opening of packages containing radioactive material. A sample of acceptable procedures is enclosed.

We also i

i recommend the use of checklists (sample enclosed) by license reviewers to insure'that all pertinent safety issues are addressed.

II. Compliance A.

Comment A review of selected compliance files showed that while inspection reports._

routinely mentioned previous inspections and the number of noncompliance items, there was no mention as to what the noncompliance were.

We did note that there was a statement on the present status of previous non-compliance items.

Recommendation We recommend that each inspection report mention very briefly, what-i-

noncompliance items were cited during the previous inspection.

  • _ 2' B.

Comment A review of selected compliance files showed the need to amend the wording of enforcement letters.

Recommendation We recommend that licensees be routinely required to state what steps would be made to prevent a recurrence of noncompliance items. These steps could be technical (e.g. revision of a procedure) or administrative (e.g. audits of technical activities by management).

4

.