ML19309H130

From kanterella
Jump to navigation Jump to search
Status Rept on Deficiency Re Rosemount Model 510 DU Trip/ Calibr Units & Model 1152 Pressure Transmitters.Transmitter Has Common Mode Failure of Capacitor When Transmitter Operating Temp Greater than 175F.Review Underway
ML19309H130
Person / Time
Site: Grand Gulf  
Issue date: 04/25/1980
From:
MISSISSIPPI POWER & LIGHT CO.
To:
Shared Package
ML19309H129 List:
References
NUDOCS 8005080484
Download: ML19309H130 (1)


Text

800 5080yIf

?

O ATTACHMENT TO AECM-80/86 i

STATUS REPORT ON PRD-80/11 4

I.

Description of Deficiency Defects in two components supplied by Rosemount are identified i

under PRD-80/11. These components are the Model 510 DU trip /cali-bration unit and the Model 1152 pressure transmitter.

The trip / calibration unit has a common mode failure of an essential switch that determines the trip output logic. The malfunction open-circuits the logic, thereby providing an inoperable signal.

The transmitter has a common mode failure of a capacitor when the transmitter operating temperature is greater than 17SOF or when the damping potentiometer is rotated clockwise. The failure is manifested as a drop in transmitter output current of up to 10 percent of the correct current.

II.

Investigation Status Our preliminary investigation indicated that the defect noted in the trip / calibration units was previously reported by MP&L as PRD-79/18.

Our final report on PRD-79/16 was submitted to the NRC in the letter dated January 30, 1980. That deficiency was determined to be reportable under 10 CFR 50.55(e). Corrective actions associated with PRD-79/18 are documented under Bechtel Management Corrective Action Request (MCAR)

GGNS No. 56.

The defect noted in the Model 1152 pressure transmitters was not previously reported by MP&L. The investigation into this defect, the component's use at GGNS, and the associated safety implications, is in progress.

_d e

e 1

l

_