ERA UNITED STATES tA* #88&
- p 9'o,,
NUCLEAR REGULATORY COMMISSION R EGION il g
101 M ARIETTA STREET,N.W.
g p g
-.g 8
ATLANT A, oE ORGI A 30303 "o
$*****/
APR 191979 VIRGINIA CONHONwEALTH UN!vERSITY MEALTH SCIENCES CENTER RICHMUND VA 23219 Gentlemen:
The enclosed IE Circular 79-06 is forwarded to you for information. If there are any questions related to your understanding of the suggested actions, please contact this office.
t w
(
lg James P. O'Reilly Regional Director
Enclosure:
IE Circular No. 79-06 m
THIS DOCUMEtlT C0tlTAlliS POOR QUAUTY PAGES g
L
?S0507GM7
UNITED STATES NUCLEAR REGULATORY COM'!ISSION OFFICE OF INSPECTION AND ENFORCEMENT WASHINGTON, D.C.
20555 April 19, 1979 IE Circular No. 79-06 FAILURE TO USE SYRINGE AND BOTTLE SHIELDS IN hTCLEAR MEDICINE Description of Circumstances:
The Nuclear Regulatory Commission (NRC) is taking this opportunity to reinforce the importance of the use of radiation shields "or syringes and bottles for nuclear medicine.
In many cases they are not always being used by persons preparing and administering doses of radiophar-This is a potentially unsafe practice as these shields maceuticals.
reduce the radiation dose to the fingers and hands of persons handling the syringes and bottles and can result in unnecessary and undesirably Calculations and high total dose to the fingers over a period of time.
studies show that failure to use the protective shields can result in a radiation dose rate to fingers of hospital staff personnel on the order of 100 mrads to one rad per minute, or a projected lifetime dose to the fingers in some cases of about 4,000 to 100,000 rada.
Notice to Medical Licensees:
Medical licensees are required to have and use these protective shields.
is imposed by either a specific condition of the NRC This requirement As license or by procedures which are incorporated into the license.
part of tneir routine audit program, NRC licensees authorized to use byproduct material pursuant to 10 CFR 35 should take the following actions:
Verify that protective shields are available and that nurses and other medical personnel are using them, except in special a.
circumstances, such as pediatric cases, when the use of syringe shields would compromise the patient's well-being.
Inform the people involved in this type of work of the potentially b.
high and serious radiation doses that can be received by the hands and fingers sf the protective shields are not used.
to the The information contained in this Circular should be brought attention of all personnel who prepare and administer licensed radio-pharmaceuticals to patients.
No written response to this Circular is required. If you require additional information regarding this subject, please contact the Director of the appropriate NRC Regional Office.
O y3o3o2oito 3
Enclosure IE Circular No. 79-06 April 19,1979 Page 1 of 3 LISTING OF IE CIRCULARS ISSUED IN LAST TWELVE MONTHS Circular Subject Date issued To of Issue No.
78-01 Loss of Well Logging 4/5/78 All Holders of Source Well Logging Source Licenses 78-02 Proper Lubricating 011 4/20/78 All Holders of Reactor OLs or for Terry Turbines cps 78-03 Packaging Greater Than 5/12/78 All Holders of Type A Quantities of Reactor OLs, cps, Low Specific Activity Fuel Cycle, Radioactive Material Priority I Material for Transport and Waste Disposal Licenses 78-04 Installation Error That S/15/78 All Holders of Could Prevent Closing of Reactor OLs or Fire Doors cps 78-05 Inadvertent Safety Injection 5/23/78 All Holders of Reactor OLs or During Cooldown cps 78-06 Potential Common Mode 5/23/78 All Holders of Reactor OLs or Flooding of ECCS Equipment Rooms at BWR Facilities cps 78-07 Damaged Components of a 5/31/78 All Holders of Bergen-Paterson Series Reactor OLs or cps 25000 Hydraulic Test Stand 78-08 Environmental Qualification 5/31/78 All Holders of Reactor OLs or of Safety Related Equipment at Nuclear Power Plants cps 78-09 Arcing of General Electric 6/5/78 All Holders of cps Company size 2 Contactors 3
Enclosure IE Circular No. 79-06 Page 2 of 3 April 19, 1979 LISTING OF IE CIRCULARS ISSUED IN LAST 1%'ELVE MONTHS Date of Issued To Circular Subject Issue No.
78-10 Control of Sealed 6/14/79 All Medical Licensees in Sources Used in Radiation Therapy Categories G and G1 78-11 Recirculation M-G 6/15/78 All Holders of BWP. OLs or cps Set Overspeed Stops 78-12 HPCI Turbine Control 6/30/78 All Holders of Valve Lift Rod Bending BWR OLs or cps for plants with HPCI Terry Turbine 78-13 Inoperability of Multiple 7/10/78 All Holders of Reactor OLs and cps Service Water Pumps except for plants located in: AL, AK, CA, FL, GA, LA, MS, SC 78-14 HPCI Turbine Reversing 7/12/78 All Holders of BWR Cha=ber Hold Down Bolting OLs or cps for plants with a HPCI Terry Turbine excepting Duane Arnold and Monticell 78-15 Checkvalves Fail to 7/20/78 All Holders of Close In Vertical Reactor OLs or cps Position 78-16 Limitorque Valve 7/26/78 All Holders of Reactor OLs or Actuators cps L l.
78-17 InadequateGu$rdtraining/
10/13/78 All Holders of Qualification;and Falsified and applicants for Reactor OLs Training Records qt i
e Enclosure IE Circular No. 79-06 Page 3 of 3 April 19, 1979 LISTING OF IE CIRCULARS ISSLTD IN LAST IVELVE MONTHS Date of Issued To Circular Subject Issue No.
11/6/78 All Holders of 78-18 UL Fire Test Reactor OLs or cps 78-19 Manual Override (Bypass) 12/28/78 All Holders of cps of Safety Actuation Signals 79-01 Administration of 1/12/79 All Medical Licensees except Unauthorized Byproduct Teletherapy Medical Material to Hu=ans Licensees and each Radiopharnaceutical Suppliers 79-02 Failure of 120 Volt Vital 2/16/79 All Holders of Reactor OLs AC Power Supplies and cps 79-03 Inadequate Guard Training-2/23/79 All Holders of Qualification and Falsified and r.pplicants for Special Nuclear Training Records Material Licenses in Safeguards Group I 79-04 Loose Locking Nut on 3/16/79 All Holders of Reactor OLs or cps Limitorque Valve Operators 79-05 Moisture Leakage in Stranded 3/20/79 All Holders of Reactor OLs or Wire Conductors cps