Health Physics Procedure 1750,Revision 1: Operation & Calibr of G.S.M. 5 Survey Instrument
ML19224B385
Person / Time
Site: Crane 
Issue date: 06/14/1976
From:
Metropolitan Edison Co
To: Mullinix W
NRC/IE
References
1750, TM-0128, TM-128, NUDOCS 7906150030
Download: ML19224B385 (8)


Text

~

TMI DOCUMENTS

],j~O,[b DOCUMENT NO:

/

0F DOCUMENT l'ROVIDED BY COIT MADE ON _

METROPOLITAN EDISON COMPANY.

N Wilda R. Mullinix, NRC 790615003o D

N 202 023

.CONTROLED cun. IV AS~El CO?Y Revi: ion 1

~

~

~

DO NO RE V OV _.

1;so THREE Mit,E ISLAND NUCLE AR STATION STATION HEALTH PHYSICS PROCEDURE 1750 OPERATION AND CALIBRATION OF G.S.M. 5 SURVEY INSTRUMENT Table of Effective Pages Pay Date Revision M

Date Revision Pm Date Revision 51.0 1.0 06/14/76 1

26.0 2.0 06/14/76 1

27.0 52.0 3.0 06/14/76 1

28.0 53.0 4.0 06/14/76 1

29.0 54.0 5.0 06/14/76 1

30.0 55.0 6.0 06/14/76 1

31.0 56.0 7.0 02/08/74 0

32.0 57.0 8.0 33.0 58.0 9.0 34.0 59.0 10.0 35.0 60.0

.1.0 36.0 61.0 12.0 37.0 62.0 13.0 38.0 6?.0 14.0 39.0 64.0 15.0 40.0 65.0 16.0 41.0 66.0 17.0 42.0 67.0 1 R.0 43.0 68.0 19.0 14.0 69.0

, 20.0 45.0 70.0 21.0 46.0 71.0 22.0 47.0 72.0 23.0 48.0 73.0 49.0 74.0 24.0 50.0 75.0 25.0 Unit 1 Staff Recommends Approval Unit 2 Staff Recommends Approval 3

Approval Date3 Approvals tALA Date i

QIcgnizant Dept. Head ognizant D'ept. Head Unit 1 PORC Re7 mmencls Appi val Unit 2 PORC Recorhmend Approval l

Date Date Chiiri\\, in of PORC Chairmar' of' ORC '

\\

/

\\

/

included included PORC comnlents of PORC comments of da 'e)

(date)

~

~ Date

~~

By Dgte By i

_7 k

.D Da e d -/ d-76 ~

~

prov I

ape ('in3ntent/ /

~

  • 56' A "'"

/

S atio

~

Unit Superintendent

~

202 024

1750 7'

Revision 1 06/14/76 TiiREE MILE ISLAND NUCuEAR STATION STATION tiEALTH PliYSICS PROCEDURE #1750 Operation and Calibration of GSM-5 Survey Instrument 1.0 PURPOSE The purpose of this. procedure is to explain in detail the operation and calibration of the GSM-5 survey instrument.

2.0 DISCUSSION The model GSM-5 survey meter is a multipurpose instrument with application and versatility for measuring iadioactivity.

There are three probes that are available, an open end GM probe, and a side window GM probe and an alpha scintillation Jetector.

A regulated 900 volt power supply provides potential for all probes.

There are ranges corresponding to 500, 5,000 and 50,000 CPM or alternately, 0.2, 2 and 20 mR/hr.

3.0 REFERENCES

3.1 WM B. Johnson and Associates Technical Manual.

3.2 Health Physics Procedure 1749.

4.0 EQUIPMENT 4.1 GSM-5 Survey Meter 4.2 AFD 2-A Alpha Scintillation Probe 4.3 GP-90 Side Window GM-Probe 4.4 GP-200 End. Window'GM-Probe 4.5 MSC-1 Manual Sample Changer 4.6 DIG-1 Digital Timer and Scaler 4.7 VPE 2 Varipulse (Pulse Generator) or equivalent

~

4.8 CM-1 Count Multiplier 1.0

-1

~

~

=-

=

202'~025~~

1750 bb#MS

/

5.0 OPERATING INSTRUCTIONS 5.1 Operating Instructions for GSM-5 Survey Meter with side window probe and ASA-2 speaker.

5.1.1 Turn range selector switch on GSM-5 to "X1" position.

5.1.2 Depress battery test button on GSM-5.

Indicating needle should move into green coloreo section of meter scale.

5.1.3 Move speaker switch to "0N" position.

If no blips are heard from speaker as mete. needle is responding, battery replacement in speaker assembly may be necessary.

5.1.4 When using side window probe, twist knurled knob slightly and push probe out of Beta shield.

5.1.5 Equipment is now ready for use.

5.1.6 CAUTIONS 5.1. 6.1 Switches on survey meter and accessories should be turned off when equipment is not in use.

5.1.6.2 Probes and/or accessory equipment should be neither connected nor disconnected while any power supply switch is in ON" position.

5.1. 7 CALIBRATION 5.1.7.1 The GSM-5 survey meter with side window probe will be calibrated at least once every 3 months.

Set Victoreen InstI Calibration 64-764, 100 mci Cs-137, in 5.1.7.2 position for calibration.

5.1.7.3 Place detector at proper location to detect 15 MR/,HR.

Adjust calibration pot if necessary as described in HPP 1636.

5.1.7.4 Place detector at proper location to detect 1.5 MR/HR._ Adjust

"~

calibration pot if necessary as described in HPP 1636.,

2.

202 026

~

1750

.l Revision 1 06/14/76 5.1.7.5 Place detector at proper location to detect 0.15 MR/HR above background.

5.1.7.6 Recheck instrument at each calibration point.

5.1.7.7 Record the calibration in the Instrument Calibration Log Book located in the H. P. Lab.

5.2 Operating procedure for GSM-5 survcy meter with end window probe, DIG-2 Count Multiplier and Digital Timer Scaler, MSC-1 manual sample changer, and ASP-2A alpha scentillation probe.

5.2.1 Turn range selector switch on GSM-5 to "XI" position.

.5.2.2 Depress Battery Test button on GSM-5.

Indicating needle should move into green colored section of meter scale.

If indicatino needle reads below green portion of meter scale, battery reple-7nt may be necessary.

5.2.3 Insert selected probe through top tole in MSC-1 manual sample changer.

Use adapter ring with end window GM probe.

If ASP-2A probe is used, switch toggle to "Orl".

5.2.4 Insert shelf in second position (middle) on MSC-1 without a sample.

5.2.5 Move toggle switch on DIG-2 to the "Ori" position.

5.2.6 Select background or sample count interval by rotating black knob to desired increment.

(.5,1, 2, 3, 4 minutes ).

5.2.7 Reset reg'ister by depressing black button on register.

tiote that all decades rcturn to zero.

5.2.G Commence background or sample count by switching toggle from "stop timer reset" position to " count" position.' " Count terminai.ed" lamp will glow at completion of count.

Return

~~

" count" toggle switch to "stop timer reset" position. '

~~

'-~

.__ _ _ _ _ ~_.J :_.'__ _Y 3.0 202 027-

~

ito0 J.

Revision 1 06/14/76 5.2.9 Place sample on second shelf on MSC-1.

5.2.10 Adjust range selector switch on GSM-5 so that needle reads on scale.

5.2.11 Select count rate multiplier by rotating black knob to desired -

increment.

5.2.12 Repeat steps 5.2.5 through 5.2.10 for the sample.

At completion of count interval, determine net cpm by the following procedure:

(Total Counts) x (Count Multiplier) -- (Background) = Net C/M (Lou 61 T ime in MTT1. )

5. 2.1.1 Calibration - Efficiency 5.2.13.1 The calibration of GSM-5 survey meter with end window probe, ASP-2A Alpha Scentillation probe, DIG-2 Count Multiplier and Digital Timer Scaler will be calibrated at least once every refueling interval.

5.2.13.2 Follow steps 5.2.1 through 5.2.5.

5.2.13.3 Count interval will be set at 1 minute.

5.2.13.4 Follow steps 5.2.7 through 5.2.11 with the exception that step 137 230 5.2.9 will be the Cs or Th sources depending upon the probe used.

5.2.13.5 At completion of count interval determine the percent efficiency by using the following formula:

fa/

x 100 = efficiency in %

To a our 5.3 DIG-2 Count Multiplier and Digital Timer-Scaler Combination.

5.3.1 Calibration of timer selector o

y._.

4.0

=

202'028~

~

1750 Revision 1 06/14/76 5.3.1.1 Set timer selector switch to.5 minutes.

5.3.1.2 Turn "on" "off" switch to "on" position.

5.3.1.3 Using a stop datch simultaneously turn timer reset switch to count position and activate stop watch.

Stop the stop watch when " time end" light comes on.

5.3.1.4 Repeat steps 5.3.1.1 through 5.3.1.3 for timer selector settings of 1 minute, 2 minutes, 3 minutes, and 4 minutes.

5.3.1.5 Record the calibration in the Instrument Calibration Log Book located in the H. P. Lab.

5.4 Calibration of Scaler Register (to be completed by Instrument Department).

5.4.1 Hook BNC Connector from Vari-Pulser " Pulse" connection tc input connector of DIG-2.

5.4.2 Set "Pos" "tieg" switch on the vari-pulser to the "Pos" position.

5.4.3 Set output selector to 1.0 volts.

5.4.4 Set minutes selector switch on the DIG-2 to the 1 minute position.

5.4.5 Set count selector on the DIG-2 to the X1 position.

5.4.6 Set the frequency CPM selector switch on the Vari-Pulser to a setting of 100 cpm.

5.4.7 Depress register reset button on the DIG-2.

5.4.8 Turn the "on" "off" switch on the CIG-2 to the on position.

5.4.9 Turn the timer reset switch on the DIG-2 to the count position.

5.4.10 Repeat step 5.4.5 through 5.4.9 using the following settings.

1.

DIG-2 Count Selector Switch - X100

~

2.

Vari-Pulser (frequency) - 10,000 cpm

-- -~~

~

~

~~

5.0

~

~

202 029

~

i' 1750 Revision 1 06/14/76 5.4.11 Record the calibration in the Instrument Calibration Log Book located in the H.P. Lab.

e e

9

.-.-e-em--ae

. 6.0

n -- -

~

~. '~~

~2g2=J Q.

k

Q,g : ' t ' '.

Revision 0 p

)

(

(

b

(

OFF XI CAL.

4 o

XIO oc o,

(

BAT.

9 o

o oo TEST XIOO O

prose C1 PHONE x' o O@O o

i \\ \\ \\ \\\\ \\\\\\\\\\3 '

j w

L J

c ;,. -

(

G S M -5 SURVEY lh STRUMEiN T -

a

... n,., ~,

n.

...: n,

Figure l'50-1 202.031 e..

.pJ

  • f.g, 5

6-..

4M

.e e e e -

e mea e apeupume se -

em.es -

e O

%e

.e um e.s.e e

.p

,,, ems,y

.e i e e. e ese me p,

e

.me-4w

-? _.

_..---
-~--: __. _

7.0

_.w.

~

G g,

A,..w

._ _.