ML19224B380

From kanterella
Jump to navigation Jump to search
Health Physics Procedure 1748,Revision 0:operation & Calibr of Eberline PIC-6A
ML19224B380
Person / Time
Site: Crane 
Issue date: 02/14/1977
From:
Metropolitan Edison Co
To: Mullinix W
NRC/IE
References
1748, TM-0126, TM-126, NUDOCS 7906150021
Download: ML19224B380 (4)


Text

e g

Y TMI DOCUMENTS DOCUMENT t;O:.

0F DOCUMENT l'ROVl:'<.D BY C01'Y MADE ON METROPOLITAN EI)1 SON COMPANY.

f Wilda R. MulJinix, NRC 79061500at se e

201 318

Revisiors 0 s.

1748 02/14/77 Thfh; M!L E ISLAND NUCLE AR STATION STAi!ON HE*i.TH PHYSICS PROCEDURE 1748 OW cq 0'i %D

-,c IE PAi!ON OF THE EBERLINE PIC-CA E f f e-tva Fages fi evision Paoe Da:e Revisior.

Pm Date Revision 1.0 02/14/77 0

2a.C 51.0 2.0

@!?:/;,

0 2 7. t>

...-$ 53.0

^

.. ;52.0 3.0 0 2 / l ',' 7 23.0

.p f,, c p,6 j--

54.0

.qt 4.0 79.0 L.

5

'.j{,[h-{,['.h, p

' 55.0 5.0

't' l

% v -

56.0 6.0 3'm

{

57.0 7.0

,..-s, 8.0 t J:

nn i ' G~I jy i g i._, b 8',0 9,0

.a u L/ J

."U 5

~~

59.0 1 J.0 30.0 60.0

1. 0 36.0 61.0 12.0 37.0 62.0 13.0

'30 0 63.0 14.0

.n G4.0 l b.0 30 C5.0 16.b

'. o 66.0 17.0

.'T 67.0 18.0 40 68 0 19.0 44.0 69.0 20.0 45.0 70.0 21.0 46.0 71.0 22.0 47 n 72.0 23.0 48.0 73.0 24.0 49.0 74.0 25.0 50.0 75.0 Unit 1 Staff Recommends Ap rova' Unit 2 Staff Pecommends Approval Date R[9[7 /

Date A/9b7 Approval 2

-A p m oval 4

Cognizant D ept. Head Cc;ninni Dept. H ead Unit 1 PORC Tbf,ammen s A h. oval Unit 2 PORC omm nds gpprcvat f

{

fi

. ___ _ o _. i _ _ D ate fi Date Chairan o. Pu fd

}

Chair la f PC RC g I 1

l PORC commem) of

,_L included PORC comme its b

included (d te)

I date{

\\

j t

Date By i

4.

Da t e ____ ___

By 1,

)

Approval Date

._ Approval

/

Dat 0 h M p. p. m v M

~~

- ~ ~ ~ ~

Mgr., O perational Quality Assurance Unit erintendent 201 319

1748 Revision 0 02/14/77 THREE MILE ISLAND NUCLEAR STATION STATION HEALTH PHYSICS PROCEDURE 1748 OPERATION AllD CALIBRATION OF THE EBERLINE PIC-6A 1.0 PURPOSE _

The purpose of this procedure is to explain in detail the operation and calibration of the Eberline PIC-6A.

2.0 DISCUSSION The PIC-6A is a small, lightweight portable instrument which measures the dose rate from gamma radiation.

The detecting element is a gas filled ionization chamber operating in the proportional (gas multiplication) reg on.

Six decades of dose rate, from (1) MR/hr. to 1000 R/hr.,

i are measured in two ranges of three decades each.

A single rotary switch turns the instrument 0FF, provides a BATTERY check, and selects the r.ange.

A beta window in the bottom of the instrument provides for the detection of energetic beta particles.

2.0 REF EP.ENCES 3.1 Eberline PIC-6A Manual 3.2 HPP 1636 Handling #682-50 Ci CS-137 Instrument Calibrator 4.0 EQUIPMENT 1

4.1 PIC-6A 4.2 50 Ci Cs-137 Source 4.3 100 mci Cs-137 Saurce

^ ~

5.0 OPERATING INSTRUCTIONS 5.1 To use the instrument for dose rates 5.1.1 Turn the ', ingle rotory switch to the BATTERY check position to insure the Datteries are in the battery check range.

__._____._.1...._

l.0 201 320

~ =~

~

-=

Revision 0 02/14/77 t

5.1.2 Turr. the single rotary switch to the MR/hr. range, and check that the instrument responds to radiation by use of a check source, located in the f;1e cabinet in the H.P. Lab, marked source drawer.

~

5.1.3 If the battery check is good and the instrument responds to radiation, the instrument can be used for surveys providing it has an up to date calibration on it.

5.1.4 Upon completion of use of the instrument turn the instrument "0FF" and return it to the H.P. Lab.

5.2 Calibration of the PIC6A.

5.2.1 Each range is individually calibrated by a ZERO and a SPAN control.

5.2.2 Unlach the latches on each end of the instrument and remove the top from the bottom.

5.2.3 For calibration of the MR/hr. scale the R-10 pot is for the ZERO control, and the R-12 pot is for the SPAN control.

5.2.4 Place the instrument in a (1) one MR/hr. radiation field and adjust R-10 until a reading of (1) MR/hr. is indicated on the meter.

5.2.5 Place the instrument in a (1) R/hr. radiation field, and adjust R12 until a reading of (1) R/hr. is indicated on the meter.

5.2.6 Repeat steps 5.2.4 and 5.2.5 until both.he (1) MR/hr. and the (1) one R/hr. is indicated on the meter.

5.2.7

' For calibration of the R/hr. scale the R-9 pot is for the ZERO control and the R-ll pot is for the SPAN control.

3

}}---

=-

201

';21

~~

2.0

_;7-

1748 Revision C s - -

02/14/77 Place the instrument in a (1) one R/hr. radiation field and S.2.d adjust R-9 until a reading of (1) one R/hr. is indicated on

- the meter.

t 5.2.9 Place the instrument in a radiation field, in the 50 Ci CS-137 Source, that is the highest obtainable and adjust R-ll until this reading is indicated on the meter.

See HPP 1635 for exposure rates associated with the 50 Ci Cs-137 source.

5.2.10 Repeat steps 5.2.8 and 5.2.9 until both readings are indicated correctly on the meter.

5.2.11 Upon completion of the calibration olace the top section into the bottom section and latch both 'st ;hes, one located on each end of the insturment.

5.2.12 Place a calibration sticker on the instrument, log in the calibration log book in the H.P. lab.

These instruments will be calibrated on a quarterly basis according to H.P.P.1749.