Health Physics Procedure 1703,Revision 1:background Determination for Beckman Wide Beta II
ML19224B179
Person / Time
Site: Crane Constellation icon.png
Issue date: 09/08/1976
From:
Metropolitan Edison Co
To: Mullinix W
NRC/IE
References
1703, TM-0110, TM-110, NUDOCS 7906140169
Download: ML19224B179 (3)


Text

_.

w t

s TMI DOCUMENTS

~

DOCUMENT NO:

COPY MADE ON

)

f OF DOCUMENT PROVIDED BY METROPOLITAN EDISCN COMPANY.

Wilda R. Mullinix, NRC I

7906140169 6

I 190 209

~

MiASTER CO?Y 1703 DO NOT REMOVE

";%7"7s

)

THREE MILE ISLAND NUCLEAR STATION i

STATION HEALTH PHYSICS PROCEDURE #1703

.J l1 BACKGROUND DETERMINATION FOR BECKMAN WIDE BETA II Table of Effective Pages Paae Date R evision Pace Date Revision Pace Date Revision 1.0 09/08/76 1

26.0 51.0 2.0 09/08/76 1

27.0 52.0 3.0 28.0 53.0 4.0 29.0 54.0 5.0 30.0 55.0 6.0 31.0 56.0 7.0 32.0 57.0 8.0 33.0 58.0 9.0 34.0 59.0 10.0 35.0 60.0 11.0 36.0 61.0 12.0 37.0 62.0 13.0 38.0 63.0 14.0 39.0 64.0 15.0 40.0 65.0 16.0 41.0 66.0 17.0 42.0 67.0 18.0 43.0 68.0 19.0 44.0 69.0 20.0 45.0 10.0 21.0 46.0 71.0 22.0 47.0 72.0 23.0 48.0 73.0 24.0 49.0 74.0 2E.0 50.0 75.0

- Una 1 Staff Recommends Approval Unit 2 Staff Recommends Approval Approval MObb t

Date b

Approval @

%MN Date h h

Cdgnizant Deat. Head C gnizant Dept. Head

' pproval Unit 1 PORC R ommen s Ap roval Unit 2 PORC ecomm nd A

l Date

\\

Date C h6irm, n o PORC

/

Cha rm-i of P R I

,f i

included PORC comm nts f included PORC commen's of 1

(dae)

(d ite)

By Date By l

i Date S[ation Suleridtendent/ /

/

~

  • '-5 5-^ ""

Unit Superintendent

1703

...)

Revision 1 09/08/76 THRtt HILC ISTAND NUCLEAR STATION

~

STATION HEALTH PHYSICS FROCEDURES #1703 Background Determination for Bechman Wide B?ta II 1.0 PURPOSE The purpose of this procedure is to list the steps taken to determine the background count rate for the Beckman Wide Beta II.

2.0 DISCUSSION The Beckman Wide Beta II is an automatic counting device for determining alpha, and/or beta radiation.

Pure,100% oxygen free, methane gas is used with a window detector, and cosmic or noise pulses are automatically rejected.

3.0 REFERENCES

The Beckman Wide Beta II Instruction Manual #1437.

4.0 EOUIPMENT 4.1 Beckman Wide Beta II Counting System.

4.2 A Teletype Printer.

4.3 100% 0xygen free Methane Counting Gas.

4.4 A Background Planchet and Dish.

5.0 OPERATING INSTRUCTIONS 5.1 Open valve behind instrument controlling gas supply. Check that gas flow reads approximately 0.2 CFH.

If no gas flow is apparent, l

a gas cylinder refill may be required.

(Located outside of building).

5.2 After a 30 minute wait, open the lower right panel and turn the Plateau Point Knob to the OPERATE position.

[

w-e 1.0

?0 ]ll

s

..r

>1703 Fevision I 09/08/76 5.3 Turn analyser Mode Selector kneb in lower right panel to Beta then Alpha.

5.4 Set the preset min kncb cn Front Panel to count time.

(10 min. or as desired).

5.5 Insure Count Moda switch in NET.

insure Operate-Leas sulten (in upper rigid p=al) in LOAD _.

5.6 After changer has stepped, press add cage advance butten (upper right panel) to bring changer to #1 cage positicn.

5.7 Place background dish and planchet in changer magazine on top of #1 cage.

(This is sample position 10).

5.8 Set first sample selector and final sample collector to 10.

5.9 Set Operate / Load switch to OPERATE. After counting starts,

turn Operate / Load switch to LOAD.

5.10 Set Cutput-Selector switch to TELETYPE.

(lower right panel) 5.10.1 Teletype Output Format.

(EXAMP LE )

Beta Alpha Time Time Of 2c Of 2c Samole# Dav Counts Time STD. DEV.

Samole# Day Counts Time STD. DEV.

000.10 3:42 00010 0010.00 00.0 000.10 3:42 000001 0010.00 C0.0 6.0 Background Counts / Min.

= Total Counts Counting Time (MIN. )

Eg. 10 Counts l.0 Counts 10 Min.

Min.

T

'190 2;2

.m

.