  • Revision 1 12/01/77 THREE MILE ISLAND NUCLEAR STATIOrjV ASTER COPY STATION HEALTH PHYSICS PROCEDURE 1763 DO NO' RENOVE H.P. IrlSTRUMEtlT CALIBRATIOil/ STATUS

.00ATROD ED COPY Table of Effective Pages Pace Date Revision Pace Date R evision Pace Date Revision 1.0 12/01/77 1

26.0 51.0 2.0 12/01/77 1

27.0 52.0 3.0 12/01/77 1

28.0 53.0 4.0 29.0 54.0 5.0 30.0 55.0 6.0 31.0 56.0 7.0 32.0 57.0 8.0 33.0 58.0 9.0 34.0 59.0 10.0 35.0 60.0 11.0 36.0 61.0 12.0 37.0 62.0 13.0 38.0 63.0 v.

14.0 39.0 64.0 15.0 40.0 65.0 16.0 41.0 66.0 17.0 42.0 67.0 18.0 43.0 68.0 19.0 44.0 69.0 20.0 45.0 70.0 21.0 46.0 71.0 22.0 47.0 72.0 23.0 48.0 73.0 24.0 49.0 74.0 25.0 50.0 75.0 Unit 1 Staf Recommends ppr val Unit 2 Staff ecommends A roval Approval A

it

>o Date (f-2 ? -77 Approval '

. ry -

<< A Date it -1F-M Cognizant Oept. Head Cognizant Oept. Head Unit 1 PORC Re mmend Ar oval Unit 2 PORC Recommen Approval Date i

Date Chairmfn a PO Chairma o PO R PORC commen of I

/

included PORC commente of\\- '

uitruded e

By I

Date By Date l

Approvb II

/m)

Date N/dO/>7 Approval i

Date '

Approva

//h

} 9 9Dap#N ///7 '?

Mgr.. O perational M/ 7 W

~

Quality Assurance Unit Superintendent

1763 Revision 1 12/01/77 THREE MILE ISLAND HUCLEAR STATION STATION HEALTH PHYSICS PROCEDURE 1763 H.P. Instrument Calibration / Status 1.0 PURPOSE The purpose of this procedure is to explain in detail the method of record keeping utilized for the calfbration, and status of the Health Fhysics Portable Survey Instrument'..

2.0 DISCUSSION A record must be maintained for all Health Physics Portable Survey Instruments to insure that proper equi,cment control is maintained and that calibratior s have been performed at the specified time intervals.

3.0 REFERENCES

3.1 H. P. P. 1749 4.0 E0_UIPMENT 4.1 Calibration / Status and Battery Check Leg Book 5.0 OPERATING INSTRUCTIONS 5.1 Form 1763-1 is the form to be utilized to maintain a record for the Calibration / Status of all the Health Physics Portable Survey Instruments.

Form 1763-1 contains the following information.

5.1.1 Instrument type and instrument serial number.

Each portable survey instrun:ent will have a Form 1763-1.

5.1.2 Column #1, Indicates the instrument status; in-service, out-of-service, not in use.

5.1.3 Column #2, Indicates the calibration date.

This is the date the instrument was last calibrated.

~

p me 1.0

1763

,'(

Reyi.sion 1 12/01/77 5.1.4 Column #3, Indicates by signature who calibrated the instrument.

5.1.5 Column #4, Indicates the calibration due date.

Refer to H.P.P.1749 for proper frequency.

5.1.6 Column #5, Is for remarks.

5.2 The status of an instrument will be determined by the following:

5.2.1 IN-SERVIC[,-The instrument is operable and has an up to date calibration sticker.

5.2.2 OUT OF SERIVCE-The instrument is not available for use because:

5.2.2.1 Out of Calibration-Recove calibration sticker and attach an "Out of Commission" tag denoting " calibration required".

5.2.2.2 Recairs Recuired-State on an "Out of Conunission Tag" the problem with the instrument and if possible a suggested repair.

Attach the tag to the instrument.

5.2.3 NOTE IN USE-The instrument is not available for use.

The instrument snould be stcwed in the "Nct in Use" locker with an "Out of Commission" tag on it denoting, "not in use".

5.3 If an instrument is "Out of Service" due to repairs required or if it is "Not in Use", denote this in Column #1 of Form 1763-1.

~

_ ;z _

_7 -

x-

~

~

2.0

~

Me+

i==ume-

--M

-e

6 4

HEALTH PHYSICS PORTABLE INSTRUMENT ev ion 1 12/01/77 CAllBRATION/ST ATUS LOG SHEET INSTRUMENT TYPE _

(NSTRUMENT SERIAL NO.

Column #1 Column 42 Column 33 Column d4 Column d5 I

INSTRUMENT STATUS CA tl8R A TIC N GLIBR^ ' "

PERFORMED U

In Service CAU 8R ATION R EM AR KS tsqdure)

CATE ATE Out of Service Not in Use qm

_ g = -.

. = _.

~~~ f h k f]__.:--

3.0

_ Form 1763-1