ML033240134

From kanterella
Jump to navigation Jump to search
Part 21 Notification (Initial Report) Re Potential Exists for Incorrect Solder Used on the Internal High Voltage & Signal Solder Connections in Model 877-1 High Range Containment Monitor Detectors
ML033240134
Person / Time
Site: Hatch, Summer, Vermont Yankee  File:NorthStar Vermont Yankee icon.png
(DPR-028, DPR-057, NPF-005, NPF-012)
Issue date: 11/18/2003
From: Castleberry C
Cardinal Health Radiation Management Services
To:
Document Control Desk, Office of Nuclear Reactor Regulation
References
Download: ML033240134 (7)


Text

Date: November 18, 2003 To:

Document Control Desk U.S. Nuclear Regulatory Commission Washington, DC 20555 From: Christine Castleberry Regulatory and Quality Assurance Manager Cardinal Health Radiation Management Services 6045 Cochran Road Solon, OH 44139 Re:

1 OCFR Part 21 Notification (Initial Report)

In accordance with 10CFR21, please find the attached notification.

Should you have any questions or require further information, please contact me at (516) 870-0100 or via email at christine.castleberrv(acardinal.com or Zisimos Giatis at (440) 498-2550 or via email at zisimos.piatisDcardinal.com.

Christine Castleberri

-'

I 6i 1 of 7

li IOCFR Part 21 Notification (Contlued) 1.0 Name and address of individual informing the Commission:

Christine Castleberry Regulatory and Quality Assurance Manager Cardinal Health Radiation Management Services 6045 Cochran Rd.

Solon, OH. 44139 (Formerly Victoreen and Inovision) 2.0 Identification of the basic component supplied which fails to comply or contains a defect:

The potential exists for incorrect solder to have been used on the internal High Voltage and Signal solder connections in Model 877-1 High Range Containment monitor detectors manufactured between 6/1/00 and 11/1/03. While the incorrect solder will maintain its integrity at the maximum operating temperature of the unit, the temperature margin designed into the unit will not be realized.

Detectors manufactured prior to 6/1/00 are not affected.

Field Termination kits, PIN 877-1-15 and 877-1-15-1 are not affected by this notification.

3.0 Identification of the firm supplying the basic component which fails to comply or contains a defect:

Cardinal Health Radiation Management Services 6045 Cochran Rd.

Solon, OH. 44139 (Formerly Victoreen and Inovision) 4.0 Nature of the defect or failure to comply and the Safety Hazard created:

Background:

The Model 877-1 is a High Range Containment radiation monitor detector. The detector is an all welded, stainless steel, hermetically sealed, Ion Chamber that operates at atmospheric pressure.

General

Description:

In 1981, the Model 877-1 High Range Containment detector was qualified for operation during and after a Loss of Coolant Accident (LOCA). The qualification program is documented in Victoreen Qualification Type Test Data Report 950.301. The LOCA temperature profile used in the 2 of 7

IOCFR Part 21 Notification (Contiued) qualification test was based on the Combined PWRIBWR LOCA profile defined in IEEE-323, Appendix A. To include margin in the LOCA test, the maximum temperature stated in Appendix A of IEEE-323 was increased by 5%, from 340 OF to 357 OF. The internal detector high voltage and signal connections, for the unit tested, were soldered with a lead free, 96.5% Sn and 3.5% Ag silver solder. This type of solder has an established melting temperature of 430 F, providing a 73 OF margin in the solder joint design.

Nature of Failure to Comply: In the units subject to this notification, the internal detector high voltage and signal connections may have been soldered with a 60% Sn and 40% Pb solder. This type of solder has an established melting temperature range of 361 OF to 374 F. While this is still above the 357 OF maximum test temperature, the same temperature margin designed into the unit will not be realized with the 60% Sn and 40% Pb solder.

From a worst case analysis, if the LOCA temperature increased 4 to 17 IF above the 357 OF qualification maximum temperature and the 361 - 374 OF melting range, it would be possible for both the signal and high voltage solder connections to melt, potentially resulting in an open and/or short circuit. The melted solder would flow, by capillary action, along the buss wire and the solder post or ground lug connection. Due to the capillary action of the solder, it is doubtful that the electrical connections would break, resulting in an open circuit.

In the event of an open circuit, however, the Electronic Check Source (ECS) test, that automatically runs every 17.1 minutes, would actuate and indicate a channel failure.

During a LOCA, the steam pressure released would result in vibrations in the detector, the magnitude of which would vary, based on actual location of the detector relative to the leak. This vibration, coupled with a temperature above the melting range, could result in solder "spatters" inside of the chamber. These 4 spatters" could result in a short circuit in the connector cavity in the base plate of the detector. If the solder shorted the signal to ground, the radiation value would decrease. Within 17.1 minutes, the Electronic Check Source (ECS) test would actuate and indicate a channel failure. If the solder shorted the High Voltage to ground, the High Voltage will begin to drop from its normal 500Vdc value. When the High Voltage drops to approximately 400 Vdc, the FAIUSAFE circuit will turn off, and the channel fail relay will change state. In either event, a channel failure would be indicated.

In the original design, melting of the internal solder connections would not occur until the ion chamber internal temperature increased to 430 0F.

SAFETY IMPACT: The overall impact of the use of the incorrect solder is that the temperature margin designed into the unit, and demonstrated in the qualification test program, will not be achieved if the lower temperature solder was used.

Note the temperature at which the 60/40 solder enters the paste range, 361 OF, is still above the maximum test temperature, 357 OF.

For a PWR, the maximum temperature specified in Appendix A of IEEE-323, Appendix A, is 300 OF. This is well below the 361 OF paste range of the 60140 3 0f7

IOCFR Part 21 Notification (Contlued) solder. Based on the above, the unit should continue to operate when subjected to the PWR LOCA event For a BWR, the maximum temperature specified in Appendix A of IEEE-323, Appendix A, is 340 OF. This is still below the 361 F paste range of the 60/40 solder, and, based on the above, the unit should continue to operate when subjected to the BWR LOCA event Because the temperature margin designed into the unit will be reduced if the 60/40 solder was used, it was decided to provide this notification to all affected Users. A Risk Analysis summary is provided in Attachment 2.

5.0 The date on which the information of such defect or failure to comply was obtained:

The problem was discovered while reviewing a routine Engineering Change Notice for the Part Number 877-1-80 Detector sub-Assembly. During the review, it was learned the Vendor part number for the original lead free silver solder used in the qualification was no longer available, and that a 60/40 Tin/Lead solder was used for the eight (8) internal signal, high voltage and ground connections in a unit presently being manufactured.

A preliminary meeting between Quality Assurance, Systems Engineering, and Manufacturing was held November 11, 2003 to review the solder issue. After further analysis, a follow-up meeting was held November 17, 2003. As a result of this meeting it was determined that a 10CFR21 notification was required.

6.0 Affected facilities:

The following NRC licensed users are affected by this failure to comply:

OwnerlPlant:

Entergy Nuclear/ Vermont Yankee (1)

S Carolina Electric & Gas / VC Summer (3)

Georgia Power Co. / Hatch (2)

Other affected customers are located in Korea (2) and in Belgium (1).

A complete list of affected users, and detector serial numbers, is provided as an attachment.

4of7

10CFR Part 21 Notification (Contiued) 7.0 Corrective Action:

1. As a result of the initial meeting, an immediate Hold was placed on Model 877-1 detectors being manufactured, and shipment of any new 877-1 detectors.
2. ECN 3977 has been initiated to define a replacement for the MSA-0015 silver solder. The replacement will be required to provide the same 96.5% Sn, 3.5%

Ag composition.

3. A review of our complete product line was performed. The review confirmed that only the products listed here-in were affected by the solder substitution.
4. Notification letters will be sent to the affected utilities, advising them of the potential for reduced temperature margin in the event of a LOCA event. The letters will be mailed by November 28, 2003.

8.0 Advice related to the defect or failure to comply about the basic component that has been, is being, or will be given to purchasers:

As stated in the analysis above, we believe there is no need to replace or repair the detectors subject to this failure to comply notification.

For further information please contact:

Zisimos Giatis Quality Control Manager Cardinal Health Radiation Management Services Ph: 440-498-2550 Email: zisimos.aiatis(ecardinal.com 5 of 7

OCFR Part 21 Notification (Contiued)

Attachment I - List of 877-1 Users with Potentially Incorrect Solder Connections Customer:

Plant:

Purchase Sales Ship Serial Order:

Order Date:

No.:

N o.:

Entergy Nuclear Vermont VY014544 185667 10/31/02 102463 Vermont Yankee Yankee Georgia Power Hatch 6052356 175573 10/31/02 102465 Georgia Power Hatch 6052356 175573 10/31/02 102466 South Carolina V. C. Summer SR715810 181000 10/31/02 102464 Electric & Gas South Carolina V. C. Summer SR715810 181000 7/31/02 102462 Electric & Gas South Carolina V. C. Summer SR709121 155438 2/28/02 102461 Electric & Gas Non-US Installations:

Centrale Tihange 3 4500196326 172821 1/6/03 102467 Nucleaire De Tihange Korea Electric KORI 3/4 or COONYG2CO 162871 11/30/01 102459 Power Co.

Yeonggwang 58480 1 /2 Korea Electric KORI 3/4 or C99NKRB02 157151 11/30/01 102460 Power Co Yeonggwang 26230 1/2 6 of 7

OCFUR Part 21 Notification (Contiued)

ATTACHMENT 2 - RISK ANALYSIS Risk Analysis Table SYSTEM RISK ANALYSIS TABLE Cardinal Health Radiation Management Services Model 877-1; Reduced Temperature (357 OF) Margin on Internal Solder Connections; Design based on use of 96.5%, SN and 3.6% Ag solder, with a melting temperature of 430 OF Page 1 of I Hazard Cause Effect Severity Probability Detectability Risk Index Reduced safety 60/40 Sn/Pb None; 60/40 a

A 1

I margin as LOCA Solder used Solder melting temp increases to instead of range, 361-374 357"F max 96.5/3.5 Sn/Ag F, is above qualified LOCA temperature:

temperature Reduced safety 60/40 Sn/Pb Melted solder b

B I

I margin as LOCA Solder used Shorts or Opens temp increases instead of H V Connection; Above 357 OF 96.5/3.5 Sn/Ag Channel Fails in Qualification 17.1 mins temperature:

Reduced safety 60/40 Sn/Pb Melted solder b

B I

I margin as LOCA Solder used Shorts or Opens temp increases instead of Signal Above 357 "F 96.5/3.5 Sn/Ag Connection; Qualification Channel Fails in temperature:

17.1 mins Engineering Operating Procedure Number EOP-04-12 Re=ision:

A Risk Index I = Acceptable I Acceptable as Implemented Method of Management:

For temperatures below 357 "F, No Action required.

For temperatures above 357 OF, the Qualification Limit of the detector, the detector may fail, and alternate means of measurement may be required.

7 of 7