Two 18 Ultrasonic Flowmeters Serial Numbers Wmp #2, & Wmp #3, Westinghouse Electric Corporation Purchase Order Number 87GC-WJ-47042
ML031080534
Person / Time
Site: Indian Point Entergy icon.png
Issue date: 04/10/2003
From:
Worcester Polytechnic Institute
To:
Office of Nuclear Reactor Regulation
References
Download: ML031080534 (33)


Text

CALIBRATION INDIAN POINT TWO 18" ULTRASONIC FLOWMETERS SERIAL NUMBERS WMP .2, and WPP #3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 MAY, JUNE 1979 - ARL NO. 106-79/C91 121 HUH.

.> .. 1.,

.4. .. ___ =. .. = .. A>. 1 -_ 4 . v.

SUMMARY

Two 18" Ultrasonic Flowmeters, Serial Numbers WMP #2 and WMP #3, were calibrated at the ALDEN RESEARCH LABORATORY of Worcester Polytechnic Institute for WESTINGHOUSE ELECTRIC CORPORATION under their Purchase Order Number 87GC-WJ-47042.

Both 18" ultrasonic flowmeters were calibrated using straight pipe configurations. The details of the piping immediately upstream and downstream of the ultrasonic meters are shown on the piping diagram.

The results are presented in the form of a tabulated data sheet as well as a plot of the average temperature and pressure corrected mean meter factor (T,P) versus the average flow rate (q) in cubic feet per second or pounds per hour.

The flow meters were calibrated at seven flow rates ranging from 5 to-35 cubic feet per .second, or 1 to 7.7 million pounds per hour.

Tests 1 and 4 in May, 1979 identified a scaling error of approximately 0.45% which was corrected for Tests 5 and 6 conducted in June, 1979.

The error was determined to be due to incorrect placement of the paths in the LEFM meter section.

SUMMARY

WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 TEST NUMBER 1 f-INDIAN POINT DATE: 24 MAY 1979 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #2 Average Average Corrected Flow Rate Mean Meter cfs. Factor (T,P)

First Series of Seven Flows 34.812 0.9941 29.791 0.9956 24.554 0.9958 19.970 0.9955 14.795 0.9960 9.956 0.9958 4.839 0.9957 Second Series of Seven Flows

,4.825 0.9934 9.975 0.9967 14.909 0.9968 19.997 0.9964 24.815 0.9962 -

29.719 0.9964 34.726 0.9964 Data includes 0.45% error due to path misplacement AME, -S I_~

SUMMARY

WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 TEST NUMBER 4 INDIAN POINT DATE: 23 MAY 1979 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #3 (WITHOUT DAMPING) Average Average Corrected Flow Rate Mean Meter cfs. Factor (T,P)

First Series of Six Flows 34.463 0.9900 29.480 0.9901 24.557 19.816 0.9919 0.9917 l

14.686 0.9939 10.0485 0.9942 Second Series of Six Flows 10.055 0.9947 14.992 0.9926 19.809 0.9913 24.837 0.9917 29.538 0.9922 34.845 0.9936 Third Series of One Flow 14.710 0.9935 Data includes 0.45% error due to path misplacement

- -- -- -b - 4' A - - - ---- W -4I

-A-- -- -- --

page 1 of 3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 INDIAN POINT DATE: 16 JUNE 1979 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #3 ARL DATA WESTINGHOUSE DATA -

Line Pressure Wreight Time Flow Average Mean Corrected Test Run Temp Downstream W t q LEFM Meter Mean Meter No. No. 0F psi: lbs.- sec. cfs. Flow Factor Factor (T,P) 3 Sig %

First Series of Seven Flows 5- 1 104 42 98174 45.325 35.020 - 34.9257 1.0029 1.0027 0.617 5 2 104 42 96951 44.828 34.967 34.8728 1.0030 1.0027 0.565 5 3 104 42 96971 44.869 34.942 34.8410 1.0032 1.0029 0.556 5 4 104 42 97297 45.045 34.923 34.8811 1.0014 1.0012 0.550 5 5 104 42 97164 45.019 34.895. 34.8462 1.0016 1.0014 0.580 5 6 104 42 97265 45.075 34.888 34.8184 1.0022 1.0020 0.492 5 7 104 48 96687- 51.957 30.087 30.0239 1.0024 1.0021 0.578 5 8 104 48 96799 52.074 30.054 29.9850 1.0025 1.0023 0.656 5, 9 104 48 96758 52.095 30.029 29.9930 1.0014 1.0012 0.942 5 10 104 48 96839 52. 124 30.038 29.9392 1.0035 1.0033 0.484 5 11 104 48 96534 51. 983 30.024 29.9402 1.0031 1.0028 0.571 12 105 58 96122 61.889 25.116 24.9637 1.0064 1.0061 0.719 5

13 105 58 96077 61.860 25. 116 24.9737 1.0060 1.0057 0.558 5

14 105 58 96097 61.913 25. 100 24.9801 1.0051 1.0048 0.467 5

105 58 95995 61.881 25.086 24.9513 1.0057 1.0054 0.627 5 15 105 58 95731 76.492 20. 239 20. 1463 1.0049 1.0046 0.779 5 16 105 58 95772 76. 600 20. 219 20.1384 1.0043 1.0040 0.617 5 17 I Im

a64,-

---. -r4 - - -4 -- -- -. 4 .z l- - --l 2- f page 2 of 3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 INDIAN POINT DATE: 16 JUNE 1979 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #3 ARL DATA WESTINGHOUSE DATA ------------

Line Pressure Weight Time Flow Average Mean Corrected Test Run Temp Downstream W; t q - LEFM Meter Mean Meter No. No. OF psi lbs. sec. cfs. - Flow Factor Factor (T,P) 3 Sig %

5 18 105 58 95741 76.632 20. 204 20.1235 1.0044 1.0041 0.939 5 19 105 60 95310 102. 170 15.085 14.9593 1.0087 1.0084 0.741 5 20 105 *60 95294 102.426 15.045 14.9389 1.0074 1.0071 0.800 5 21 105 60 94959 152.586 10.064 9.9535 1.0114 1.0111 0.993 5 22 105 60 94887 152.595 10.056 9.9505 1.0109 1.0106 1.031 5 23 105 60 94846 310.097 4.946 4.9038 1.0089 1.0086 1.780 Second Series of Seven Flows 5 24 105 60 94846 313. 548 4.892 4.9305 0.9925 0.9922 1.860 5 25 106 60 95173 154.592 9.958 9.9421 1.0019 1.0016 1.018 5 26 106 60 95060 154. 283 9.966 9.9441 1.0025 1.0022 0.833 5 27 106 60 95289 102.256 15.073 15.0534 1.0016 1.0013 0.642 5 28 106 60 95442 102.461 15.067 15.0504 1.0014 1.0011 0.899 5 29 106 58 95457 76.220 20. 257 20.2813 0.9992 0.9988 0.774 5 30 106 58 95853 76.498 20.267 20.2913 0.9991 0.9988 0.539 5 31 106 58 95498 76. 275 20.251 20.2916 0.9983 0.9980 0.816 5 32 106 58 96027 62.093 25. 014 25.0365 0.9994 0.9991 0.453 5 33 106 58 96057 62.082 25.027 25.0320 1.0001 0. 9998 0.567

SUMMARY

WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 TEST NUMBER 5 INDIAN POINT DATE: 16 JUNE 1979 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #3 L Average Average Corrected

.L Flow Rate (cfs.)

Mean Meter Factor (T,P)

First Series of Seven Flows 34.939 1.0022 30.046 1.0023

25. 105 1.0055 20.221 1.0042 -

15.065 1.0078 10.060 1.0109 4.946 1.0086 Second Series of Seven Flows 4.892 0.9922 9.962 1.0019 15.070 1.0012

- 20.258 0.9985 25.014 1.0000 29.901 0.9973 34.788 0.9975 i

mma

SUMMARY

WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 TEST NUMBER 6 INDIAN POINT DATE: 16 JUNE 1979 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #3 Average Flow-Rate Mean Meter lbs/hr x 10 6 Factor First Series of Seven Flows 7.763 0.9970

'6.628 0.9984 5.575 1.0019 4.476 1.0000 3.277 1.0029 2.207 1.0013 1.089 0.9820 Second Series of Seven Flows 1.098 0.9933 2.205 1.0000 3.326 1.0021 4.426 0.9998 5.496 1.0028 6.602 1.0003 7.726 1.0002

APPARATUS 100,000-POUND WEIGHING TANK FACILITY 36" TEST LINES The flow element was installed in a test line on the main floor of Laboratory Building No. 1 as shown on the Laboratory sketch.

Water for the tests was pumped from a 180,000-gallon capacity sump by means of one (or two) 300-horsepower centrifugal pump(s) arranged to operate in parallel.

The piping arrangement immediately upstream and downstream of the element to be tested is shown on the piping diagram. Careful attention was paid in aligning the flow element in the pipe. Vents were provided at various intervals in the pipe to purge the system of air.

The flow rates were obtained by means of a weigh tank mounted on Toledo scales with a capacity of 100,000 pounds, BLH Load Cell Back-up, a Hewlett-Packard "5300A" measuring system crystal reference timer using a 5301A 10 MHZ Frequency Counter Module, A Toledo Systems Crystal Reference Timer as a back-up timer, and a mercury thermometer.

rr- . r- K.I. .-. r . .

" , .I h / .- I I-1 TEST LINE 1 1 Timor Switch 0

I I

i I

I I TEST LINE 3 High Head Line _I I

I Sump Il L. _ _ _ _ _ _ _ _ _ ____ _ _ _ _ _ _ _ _ __- - - I Breakdown Valves Scale Sump 10,000 Lb.

Weighing Tank Manifold CALIBRATION FACILITIES 0 5 10 15 30 L I I I i I LABORATORY BUILDING NUMBER 1 SCALE - FEET

-i -- Fr - r-- I - F7 r-- F-- F-- r-- Fr - r-.

- - .--- r,--- . - - - ---- -- l- -

PLAN VIEW B.UILDING 1, LINE 2 36 x 24

-FLOW STRAIGHTENER 18 x 24

/ 24 x 20 7-20 x 18 DRESSER COUPLINGS 36" 27' 2' 1 PIPING DIAGRAM TWO 18" ULTRASONIC FLOWMETERS WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 JUNE 1979

PROCEDURE After the LEFM had been installed in the test line, water was pumped into the system and all air was purged from the lines. The control valve at the downstream end of the test line was set to give the desired flow rate with the discharge being diverted to the sump.

When steady state conditions were reached, the discharge valve on the weighing tank was closed and the scale indicator and the electronic timer were both zeroed. The water was then diverted to the weighing tank and the electronic timer and the Westinghouse readout on the Indian Point LEFM were actuated at the instant of diversion.

At the end of the run the water was diverted to the sump and the electronic timer and the LEFM readout were stopped, terminating the test run. The accumulated weight in the tank, the time, the water temperature, and the readout data were recorded.

The flow was then adjusted to the next desired flow rate and the procedure repeated.

Testing on the 18" pipe section and LEFM was done at seven flow rates of 35, 30, 25, 20, 15, 10, and 5 cubic feet per second. In order to accumulate enough data from the LEFM, at least six weigh tank readings were done at 35 cfs, five at 30 cfs, four at 25 cfs, three at 20 cfs, three at 15 cfs, two at 10 cfs, and one at 5 cfs. This combined testing procedure was called a series of seven flows in this report.

The following table summarizes the number of tests that were conducted on the 18" pipe sections and Indian Point LEFM.

Test Flow Piping Number Pipe S/N No. Unit Configuration Of Tests WMP #2 1 cfs Straight Two series of seven flows WMP r3 4 cfs Straight (without Two series of six flows damping) One series of one flow WMP #3 5 cfs Straight Two series of seven flows WdMP #3 6 lb§/hr Straight Two series of seven flows

Page 1 of 2 CALCULATIONS Actual Flow Rate q W

Lqa= ty1

.q = actual flow rate in cubic feet per second E W = weight of water accumulated during run in pounds t = duration of run in seconds Ylia = specific weight of water in air at run temperature in pounds per cubic foot Note: See Appendix Temperature Correction - (TC)

TC = 1 + 3 (u) (t - T) u = coefficient of thermal expansion = 6.5 -6 x 10 T = temperature of meter section = 701F (during measurement of pipe dimensions) t = temperature of fluid during run -F L Pressure Correction - (PC)

LC 1 +(1-f-) PR 1 + (0.5-v)

IL th.E v = Poission's ratio = 0.3 E = modulus of elasticity = 3.0 x 107 th = pipe wall thickness = 0.5" R = pipe radius P = fluid pressure during run

Page 2 of 2 CALCULATIONS Total Correction Factor = (TC) (PC)

Mean Meter Factor = MF MF = Weigh Tank Average LEFM Flow Weigh Tank = ARL actual flow rate in cubic feet per second Average LEFM Flow = Qavg = N1 Q4 sec N = number of LEFM samples during run Temperature and Pressure Corrected Mean Meter Factor - MF (T,P)

MF (T,P) = MF x TC x PC Standard Deviation - a%

N Q.

~avg 1 N-1 Qavg 1i o = standard deviation Qj= individual sampled LEFM flow

RESULTS The tabulated data and result sheets indicate the experimental testing information for the individual test points and the plot of temperature and pressure corrected mean meter factor (T,P) versus average flow rate show the trend of the coefficient over the flow range tested.

Since the flows were measured to the nearest ten pounds and the time to the nearest 0.001 second, the average flow rate for each run should be within +/-0.25% of the true value.

The Laboratory's standards for the measurement of temperature, time, weight and length are traceable to the National Bureau of Standards.

K.

L Ll L

L L.

!1E

F- -

I r -

[-

r -

r- -,--

.r r---

!n .

F---

- ~ r--- I~I. -

f----

n - I- F--- , ~ r4 1 page 1 of 3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 INDIAN POINT DATE: 24 MAY 1979 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #2 Data includes 0.45% error due to path misplacement ARL DATA --- WESTINGHOUSE DATA --------

Line Pressure .Vlight Time Flow Average Mean Corrected Test Run Temp Downstream W t q LEFM Meter Mean Meter No. No. OF psi lbs. sec. cfs.- Flow Factor Factor (T;P) 3 Sig %

First Series of Seven Flows 1 58 96 40 97646 45.242 34.838 35.0448- 0.9942 0.9941 0.463 1 59 96 40 97526 45.234 34.802 35.0332 0.9935 0.9934 0.429 1 60 96 40 97475 45. 176 34.;828 35.0382 0.9941 0.9940 0.361 1 61 96 40 97414 45.189 34.796 35.0096 0.9939k> 0.9939 0.511 1 62 96 40 97414 45.183 34.801 35.0040 0.9943 0.9942 0.373 1 63 96 40 97434 45. 189 34.804 34.9754-i 0.9952 0.9951 0.563 1 64 96 48 96977 52.402 29.872 30.0342 0.9947 0.9946 0.641 1 65 96 48 97150 52.630 29.796 29.9007 0.9966 0.9965 0.649 1 66 96 48 97110 52.638 29.779 29.9106 0.9957 0.9956 0.569 1 67 96 48 97039 52. 637 29.758 29.8925 0.9956 0.9955 0.545 1 68 96 48 97029 52.646 29.750 29.8725 0.9960 0.9959 0.471 1 69 96 54 96521 63. 346 24. 595 24.6938 0.9961 0.9960 0.348 1 70 96 54 96490 63. 400 24. 566 24.6523 0.9966 0.9965 0.635 1 71 96 54 96460 63.429 24. 547 24.6505 0.9959 0.9958 0.573 1i 72 96 54 96420 63.510 24. 506 24.6316 0.9950 0.9949 0.566 1 73 96 57 96156 77.704 19.975 20.0734 0.9952 0.9951 0.753 1 74 97 57 96106 77.695 19.970 20.0623 0.9956 0.9954 0.798

r- -

I I -,I ra i--- r--- o - , ~r--

F- F

. .;..6 V

rlo.

- r-- .r-> ,. f, . ( - - I -

-i - b-t Isa . .1 --.. ... A* ,..

, I.I.-I -I Ybi&

, -11 I- - -- -

.1-Jz1,9L-2, -...

-I,-

I.

I-, i xmw-4ft, page 2 of 3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 INDIAN POINT 18" ULTRASONIC FLOWMETER DATE: 24 MAY 1979 SERIAL NUMBER WMP #2 ARL DATA ------------- WESTINGHOUSE DATA ------------

Line Pressure Weight Time Flow Test Run Temp Average Mean Corrected Downstream W t No. No. OF q LEFM Meter Mean Meter psi lbs. sec. cfs. Flow Factor Factor (T,P) 3 Sig %

1 75 97 57 95994 77.625 19.965 20.0472 0.9960 0.9959 0.552 1 76 97 60 95801 103. 865 14.891 14.9538 0.9960 0.9958 0.819 1 77 97 60 95750 104.694 14.765 14.8139 0.9969 0.9967 0.645 1 78 97 60 95527 104.708 14. 729 14.7941 0.9957 0.9956 0.792 1 79 97 60 75651 122. 625 9.960 9.9990 0.9962 0.9961 1 80 97 60 0.773 95548 155.017 9.951 9.9960 0.9956 1 81 97 60 0.9955 0.725 88793 296.221 4.839 4.8599 0.9958 0.9957 1.285 Second Series of Seven Flows 1 82 98 60 - 46901 156. 971 4.825 4.8571 0.9935 0.9934 1 83 98 60 1.273 85843 138.987 9.973 10.0050 0.9970 1 84 98 0.9968 0.877 60 85712 138.746 9.976 10.0100 1 85 0.9968 0.9966 0.683 98 60 95781" 103. 726 14. 911 1 14.9679 0.9964 0.9962 0.719 86 98 60 95801 103.778 14.907 14.9459 0.9975 0.9974 0.685 1 87 98 57 95963 77.513 19.991 20.0693 0.9963 0.9961 0.673 1 88 98 57 95892 77.430 19.998 20.0743 0.9964 0.9962 0.587 1 89 98 57 96054 77.550 20.001 20.0652 0.9970 0.9968 0.825 1 90 98 54 96399 62.655 24.845 24.9248 0.9969 0.9968 0.488 1 91 98 54 96379 62,708 24.818 24.9152 0.9963 0.9961 0.408 1 92 98 54 96521 62.859 24.795 24.9021 0.9958 0.9957 .0.469 1 93 98 54 96440 62.790 24.802 24.9016 0.9961 0.9960 0.662 Data includes 0.45% error due to path misplacement

, A -~ _

-r- -, -, '-* -f- --- j- -- 11_1_r---_,J-,F--,,---F k- r -~A ,~LJ 1,4- ,- ,,d64 , 11,r ".

- I 11- .-I r , I page 3 of 3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 INDIAN POINT DATE: 24 MAY 1979 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #2 ARL DATA WESTINGHOUSE DATA ------

Line Pressure Weight Time Flow Average Mean Corrected Test Run Temp Downstream W t q LEFM Meter Mean Meter No. No. 9F psi lbs. sec. cfs. Flow Factor Factor (T,P) 3 Sig %

1 94 98 48 96775 52. 531 29.748 29.8495 0.9968 0.9966 0.451 1 95 98 48 97008 52.735 29.705 29.8093 0.9966 0.9965 0.712 1 96 98. 48 96755 52.583 29.713 29.8353 0.9960 0.9959 0.517 1 97 98 48 96988 52.699 29.719 29.8144 0.9969 0.9968 0.671 1 98 98 48 96846 52.635 29.711 29.8273 0.9962 0.9961 0.666 1 99 99 40 97474 45.320 34.738 34.8635 0.9965 0.9964 0.629 1 100 99 40 97231 45. 223 34.725 34.8715 0.9959 0.9958 0.497 1 101 99 40 97282 45. 248 34.724 34.8425 0.9968 0.9966 0.523 1 102 99 40 97272 45.244 34.724 34.8425 0.9967 0.9966 0.506 1 103 99 40 97130 45.186 34.718 34.8609 0.9961 0.9959 0.354 1 104 99 40 97322 45. 266 34. 725 34.8330 0.9971 0.9969 0.456 WITNESSED AND ASSISTED BY CLIENT REPRESENTATIVES ARL DATA CERTIFIED BY:

WITNESSED BY CONSOLIDATED EDISON REPRESENTATIVES Data includes 0.45% error due to path misplacemefnt

_ _ _I--

,-2 2X- -- U -- .17716O . 11,1,0

- ..17 - J - xz- X i7I'77-[ & - -- ii m ri UWMIt M .....

NS#S-it1m4SttEtSli I :tM P:

......... W - Ik I..g i~ W g i gmlmIData includes 0.45% error due to path misplacement M o-

1. nrn_ 3Mi-TMWIMENIM1NSMINN .... ~~~ l li il 011-01111E lsi .-- il lil Il-r1-1--

. . ... o~...

rlra...... rr mIS M .n111nlfM IIII11I-1flWM0

4 1111 IIII 1 Ill I Ili Ill I... IiII..1111 IIII

.~m U U1 Ii.§ i I................ 1 11 111....I.....................~l fit-fl-11t 44Th1i4-fl- I fI I HA ,'14 itNiI1tIIH 1141M 1t 1+/-

i11,

¢ P4 W: 0

' - -I ------II

¢ Z4

¢C 0.99-IIlul lml l e+ee I- 4 J-U itl-1ttrtttt~t

.... m1

-t-l td 1 t tdtd+ 1- td1-t1 idtdI 11-MA-M-i

- -i +i-i

-M 1. I-1-4 1tioiifmi 10 20 30 AVERACE FLOW RATE - q (efs)

CERTIFIED BY: aVAS ' -a TEST NUMBER 1 (3 - First Series of Seven Flows CD- Second Series of Seven Flows INDIAN POINT 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #2 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 MAY, JUNE 1979

-- .- -- I -L--F -~ - r 7; -r, F J page 1 of 3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 INDIAN POINT DATE: 23 MAY 1979 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #3 (WITHOUT DAMPING)


ARL DATA ------------------------ ---------- WESTINGHOUSE DATA -------------

Line Pressure Weight Time Flow Average Mean Corrected Test Run Temp Downstream W t q LEFM Meter Mean Meter No. No. OF psi lbs. sec. cfs. Flow Factor Factor (T,P) 3 Sig %

First Series of Six Flows, 4 1 93 40 97434 45.582 34.484 34.8570 0.9893 0.9893 0.406 4 2 93 40 97525 45.631 34.479 34.8410 0.9896 0.9896 0.533 4 3 93 40 97384 45.586 34.484 34.8253 0.9902 0.9902 0.390 4 4 93 40 97616 45.707 34.454 34.7880 0.9904 0.9904 0.387 4 5 93 - 40 97555 45.706 34.433 34.7878 0.9898 0.9898 0.463 4 6 93 40 97535 45.681 34.445 34.7613 0.9910 0.9909 0.624 4 7 93 48 97010 53.008 29.524 29.8132 0.9903 0.9903 0.656 4 8 93 48 96839 52.985 29.485 29.8009 0.9895 0.9894 0.462 4 9 93 48 96899 53.062 29.460 29.7666 0.9898 0.9897 0.448 4 10 93 48 96869 53.044 29.461 29.7526 0.9902 0.9902 0.683 4 11 93 48 97000 53.100 29.470 29.7406 0.9909 0.9909 0.486 4 12 93 54 96464 63.204 24.622 24.8331 0.9915 0.9915- 0.631 4 13 93 54 96414 63.340 24.556 24.7590 0.9918 0.9918 0.631 4 14 93 54 96212 63.268 24.533 24.7383 0.9917 0.9917 0.539 4 15 93 54 96333 63.394 24.515 24.7027 0.9925 0.9924 0.745 4 16 93 57 96101 78.194 19.827 20.0151 0.9907 0.9906 0.614 4 17 93 57 95969 78.173 19.805 19.9627 0.9921 0.9921 0.685 4 18 93 57 95898 78.066 19.817 19.9668 0.9925 0.9925 0.578 4 19 93 60 95586 104.746 14.722 14.8168 0.9936 0.9936 0.852 4 20 93 60 95495 105.159 14.650 14.7355 0.9942 0.9942 0.830 Data includes 0.45% error due to path misplacement I

- -; - rw b -F-F F----- A-

'I --- (( - -W ------ --- 4 page 2 of 3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 INDIAN POINT 18" ULTRASONIC FLOWMETER DATE: 23 MAY 1979 SERIAL NUMBER WMP #3 (WITHOUT DAMPING)

__--- ------------- -ARL DATA WESTINGHOUSE DATA ----------

Line Pressure Weight Flow Flow Average Mean Test Run Temp Corrected Downstream W t q LEFM Meter Mean Meter No. No. OF psi lbs. sec. cfs. Flow Factor Factor (T,P) 3 Sig %

4 21 93 60 75459 121. 115 10.051 10.1046 0.9948 0.9947 0.941

.4 22 93 60 75520 121.293 10.044 10. 1087 0.9936 0.9936 0.804 Second Series of Six Flows Data includes 0.45% error due to path misplacement 4 23 93 60. :75515 121. 129 10.057 10. 1014 0.9956 0.9956 0.866 4 24 93 60 75934 121.851 10.053 10.1167 0.9938 0.9937 0.713 4 25 93 60 95487 102.760 14.991 15.1073 0.9924 0.9923 0.884 4 26 93 60 95812 103.097 14.992 15. 1053 0.9925 0.9925 0.954 4 27 93 60 95578 102.843 14.993 15.0972 0.9932 0.9931 0.612 4 28 93 57 95923 78.077 19.820 19.9859 0.9918 0.9917 0.746 4 29 93 57 95984 78.192 19.803 19.9788 0.9912 0.9912 0.763 4 30 93 57 96207 78.375 19.803 19.9849 0.9910 0.9909 0.772 4 31 93 54 96328 62.548 24.822 25.0500 0.9910 0.9909 0.630 4 32 93 54 96460 62.647 24.840 25.0403 0.9921 0.9920 0.736 4 33 93 54 96298 62. 547 24.838 25.0408 0.9920 0.9919 0.651 4 34 93 54 96197 62.455 24.848 25.0459 0.9922 0.9921 0.660 4 35 93 48 96846 52.846 29.564 29.8174 0.9915 0.9915 0.584 4 36 93 48 96897 52.926 29.535 29.7522 0.9927 0.9927 0.448 4 37 93 48 97059 53.044 29. 519 29.7511 0.9923 0.9922 0.562 1 -k-k

near ir, ,*r.-I I-- I - l --- r-,I - -- ____-- --.-

-I ESIHU - CORPORATI -f~

-- -i pa, 3ge 3 of page 3 of 3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 INDIAN POINT 18" ULTRASONIC FLOWMETER DATE: 23 MAY 1979 SERIAL NUMBER WMP #3 (WITHOUT DAMPING)

ARL DATA WESTINGHOUSE DATA Line Pressure Weight Time Flow Average Mean Test Run Corrected Temp Downstream W t LEFM No. No. q Meter Mean Meter 0F psi lbs. sec. cfs. Flow Factor Factor (T,P) 3 Sig %

4 38 93 48 96755 52.848 29.536 29.7652 0.9924 0.9923 0.543 4 39 93 48 96917 52.938 29.535 29.7582 0.9926 0.9925 0.318 4 40 93 40 97424 45.114 34.838 35.0589 0.9937 0.9937 0.505 4 41 93 40 97323 45.046 34.855 35.0972 .0.9932 0.9931 0.621 4 42 93 40 97495 45.141 34.843 35.0568 0.9939 0.9939 0.427 4 43 93 40 97475 45. 122 34.850 35.0886 0.9933 0.9932 0.496 4 44 93 40 97495 45.150 34.836 35.0569 0.9937 0.9937 0.483 4 45 93 40 97434 45. 103 34.850 35.0674 0.9938 0.9938 0.496 Third Series of One Flow 4 46 93 60 95680 104.489 14.772 14.8806 0.9928 0.9927 0.699 4 .47 93 60 95517 104.600 14.732 14.8269 0.9937 0.9936 0.946 4 48 93 60 95568 105.411 14.626 14.7128 0.9942 0.q941 0.738 WITNESSED AND ASSISTED BY CLIENT REPRESENTATIVES ARL DATA CERTIFI ED BY:

WITNESSED BY CONSOLIDATED EDISON REPRESENTATIVES Data includes 0.45% error due to path misplacement

r4t- - -; , --- r1 > F F F 1-- F-

-- r-V r~- [--- r--- r- r- r-- [-- - r--

-~ -- - ata includes 0.457. error 1 --- - due to path misplacement U r4 M~O

-; o ggt1 - iI-lli1t1; ~

10 20 30 AVERAGE FLOW RATE - q (cfs)

TEST NUMBER 4 (without damping)

- First Series of Six Flows CERTIFIED BY:

[ - Second Series of Six Flows

- Third Series of One Flow-INDIAN POINT 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 MAY, JUNE 1979

-.-- -1 - - - - - --

em![ -,1--- -_ r --~ i - r- I I -~ [. F, f--

--', FV [- [7- ----- f-- r---- w--; h 4 page 3 of 3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 INDIAN POINT DATE: 16 JUNE 1979 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #3

__-_----------------- ARL DATA WESTINGHOUSE DATA Line Pressure Weight Time Flowv Average Mean Corrected Test Run Temp Downstream W t q LEFM Meter Mean Meter No. No. OF psi lbs. sec. cfs. Flow Factor Factor (T,P) 3 Sig %

5 34 106 58 96102 62. 157 25.008 24.9955 1.0008 1.0005 0.338 5 35 106 58 95986 62.085 25.007 24.9945 1.0008 1.0005 0.698 5 36 106 48 96565 52. 191 29.927 29.9930 0. 9981 0.9978 0.366 5 37 107 48 96555 52.241 29.902 29.9799 0.9978 0.9974 0.695 5 38 107 48 96382 52. 140 29.906 30.0020 0. 9971 0. 9968 0.554 5 39 107 48 96351 52.176 29.876 29.9389 0. 9982 0. 9979 0.472 5 40 107 48 96708 52.338 29.894 29.9900 0.9971 0. 9968 0.776 5 41 107 42 97104 45. 111 34.825 34.9228 0.9975 0.9972 0.402 5 42 107 42 96809 45.028 34.783 34.8807 0. 9975 0.9972 0.550 5 43 107 42 96789 44.998 34.799 34.8583 0.9986 0.9983 0.413 5 44 107 42 97063 45. 166 34.768, 34.8412 0. 9982 0. 9979 0.466 5 45 107 42 96849 45.044 34.785 34.8617 0.9981 0.9978 0.438 5 46 107 42 97083 45. 176 34.767 34.8786 0. 9971 0. 9968 0.643 WITNESSED AND ASSISTED BY-CLIENT REPRESENTATIVES ARL DATA CERTIFIED BY:

WITNESSED BY CONSOLIDATED EDISON REPRESENTATIVES

' .2E-

i e ra[

Wrat r- ' re [--a r--- r- F -1 r - w-- Ir" rum r ---r r--"-, >-- f -- 7-r FMT 1.01-m EH E-4 W O A;C) 0<

' m 1.00-0 .9 W ~Z oz s 10 20 30 AVERAGE FLOWI RATE - q (cfs)

TEST 11MBER 5 0- First Series of Seven Flows CERTIFIED BY: ;i E]- Second Series of Seven Flows INDIAN POINT 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 MAY, JUNE 1979 1§Ž7ZIL

F-- , I--, I

[-- -- I >F.- - - FF- r - f---- F-- [ . r-- , F 7- F-, I-  ?(

r page 1 of 3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 INDIAN POINT DATE: 16 JUNE 1979 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #3 ARL DATA ------- WESTINGHOUSE DATA -------

Line Pressure Weight Time Flow (q) Average Mean Test Run Temp Downstream W t lbs/hr LEFM Meter No. No. OF psi lbs. sec. x 106 Flow Factor 3 Sig %

First Series of Seven Flows 6 47 108 42 97033 44.994 7.764 7.7866 0.9974 0.631 6 48 108 42 97206 45.048 7.768,. 7.7968 0.9967 0.447 6 49 108 42 96849 44.919 7.762 7.7908 0.9966 0.674 6 50 108 42 97033 45.012 7.761 7.7851 0.9972 0.655 6 51 108 42 96941 44.954 7.763 7.7871 0.9972 0.573 6 52 108 42 97113 45.060 7.759 7.7839 0.9971 0.524 6 53 108 48 96346 52.314 6.630 6.6479 0.9976 0.700 6 54 108 48 96484 52.392 .6.630 6.6413 0.9987 0.495 6 55 108 48 96494 52.394 6.630 6.6440 0.9983 0.728 6 56 108 48 96524 52.456 6.624 6.6340 0.9988 0.719 6 57 108 48 96565 52.453 6.627 6.6376 0.9987 0.535 6 58 108 58 96077 61.992 5.579 5.5695 1.0020 0.768 6 59 108 58 96057 62.034 5.574' 5.5684 1.0013 0.868 6 60 108 58 96117 62.067 5.575 5.5639 1.0024 0.727 6 61 108 58 .96097 62.081 5.572 5.5648 1.0017 0.660 6 62 108 58 95843 77.019 4.480' 4.4813 1.0000 0.849

-1 r- I--- -- F.- -- I r-- r- I r--: --- F--- rm rI--- r-- rr-F- C f7 w ra page 2 of 3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 INDIAN POINT DATE: 16 JUNE 1979 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #3 ARL DATA - -- WESTINGHOUSE DATA Line Pressure Weight Time Flow (q) Average Mean Test Run Temp Downstream W . t lbs/hr LEFM Meter No. No. OF psi lbs. sec. x lob Flow Factor 3 Sig %

6 63 109 58 95533 76.881 4.473 4.4761 0.9997 0.705 6 64 109 58 95620 76.913 4.476 4.4760 1.0004 0.821 6 65 109 60 95183 104.214 3.288 3.2791 1.0031 1.195 6 66 109 60 95163 104.898 3.266 3.2585 1.0027 0.903 6 67 109 60 95158 155. 157 2.208 2.2034 1.0024 1.032 6 68 109 60 94979 - 154.917 2.207 2.2077 1.0001 1.018 6 69 109 60 94836 313.642 1;089 1.1094 0.9820 1.837 Second Series of Seven Flows 6 70 109 60 95856 314.357 1.098 1.1057 0.9933 1.949 6 71 109 60 95143 155. 383 2.204 2.2053 0.9997 1.097 6 72 110 60 95000 155.052 2.206 2.2064 1.0002 0.981 6 73 110 60 95345 103.177 3.327 3.3217 1.0020 0.810 6 74 110 60 95295 103. 148 3.326 3.3200 1.0022 0.913 6 75 110 58 95609 77.770 4.426 4.4291 0.9997 0.685 6 76 110 58 95659 77.788 4.427 4.4283 1.0001 0.831 6 77 110 58 95609 77.790 4.425 4.4290 0.9995 0.768 6 78 110 58 95925 62.787 5.500- 5.4879 1.0026 0.769 6 79 110 58 95955 62.870 5.494 5.4808 1.0028 0.505

!&r7JL

-! I rI r-- - r-- re-- r r----. r -4 ,

page 3 of 3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 INDIAN POINT DATE: 16 JUNE 1979 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #3

, ----------------------- ARL DATA ------ WESTINGHOUSE DATA Line Pressure Weight Time Flow (q) Average Mean Test Run Temp Downstream W t lbs/hr LEFM Meter No. No. OF psi lbs. sec. Ox 106, Flow Factor 3 Sig %

6 '80 110 58 96006 62.911 5.494 5.4781 1.0033 0.725 6 81 110 58 95924 62.845 5.495 , 5.4835 1.0025- 0.606 6 82 110 48 96524 52.618 6.603 6.5974 1.0012 0.852 6 9 83 110 48 96743 52.763 6.601 6.6030 1.0001 0.522 6 84 110 48 96524 52.607 6.605 6.6030 1.0007 0.448

.6 85 110 48 96336, 52.535 6.601 6.6050 0.9998 0.719 6 86 110 48 96433 52.612 6.599 6.6036 0.9996 0.711 6 87 110 42 96951 45. 135 7.733 7.7307 1.0007 0.385 6 88 110 t42 96707 45.048 7.728 7.7249 1.0007 0.590 6 89 110 42 97022 45.210 7.726 7.7275 1.0002 0.394 6 90 110 42 96799 45. 114 7.724 7.7263 1.0001 0.473 6 91 110 42 96951 45.193 7.721 7.7241 1.0000 0.551 6 92 110 42 97033 45.237 7.722 7.7290 0.9994 0.843 WITNESSED AND ASSISTED BY CLIENT REPRESENTATIVES ARL DATA CERTIFIED BY:

WITNESSED BY CONSOLIDATED EDISON REPRESENTATIVES 1,

I ____

0,7 f- F-i rF7 F--- F- - rem m} r.--m ru- row F-- r- - -- F~--- r- -- ' re m

1.01 0~

H 1.00 C-,

W 0.99 0.98 I I I 0 2 4 6 8 AVERAGE FLOW RATE - lbs/hr x 106 TEST NUMBER 6 0 - First Series of Seven Flows CERTIFIED BY:

[ - Second Series of Seven Flows INDIAN POINT 18" ULTRASONIC FLOWMETER SERIAL NUMBER WMP #3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 MAY, JUNE 1979 , zarn

APPENDIX PHYSICAL CHARACTERISTIC OF WATER AND AIR AT ARL Sp Wt Water Sp.Wt.Wdter Sp Wt.Water Air at Sp.Wt.Water Air at Temp in vacu Sp.Wt Air Same Temp Kin Vis Temp. in vacu Sp.Wt.Air SameTemp. Kin Vis lb/ft3 lF lb/ft3 lb/ft 3 Sq Ft/Sec aF lb/ft3 lb/ft3 lb/ft3 Sq. Ft/Sec.

32 62 3980 0 0807 62.3173 1.9292 x 10 5 80 62.1963 00735 62.1228 0.9297 x 10'5 33 .3998 0.0805 .3193 1.8933 81 .1864 Q0734 .1131 0.9189 34 .4017 0.0804 .3213 1.8573 82 .1766 Q0732 .1034 0.9082

-- 35 62.4035 0.0802 62.3233 1.8214 83 .1667 0.0731 .0936 0.8974

! 36 .4040 0.0800 .3240 1.7899 84 .1569 Q0729 .0839 0.8867 L37 .4046 0 0799 .3247 1.7585 85 62.1470 a0728 62.0742 Q8759 38 .4051 0.0797 .3254 1.7270 86 .1363 00727 .0636 0 8661 39 .4057 0.0796 .3261 1 6956 87 .1255 00726 .0530 0 8563 K 40 62.4062 0 0794 62.3268 1.6641 88 .1148 10724 .04-23 0 8466 41 .4052 0.0792 .3259 1.6367 89 .1040 010723 .0317 0.8368

- 42 .4041 0 0791 .3250 1.6093 90 62.0933 0.0722 62.0211 0.8270 43 .4031 0.0789 .3242 1 5819 91 .0818 00721 62.0097 0.8181 L 44 .4020 P.0788 .3233 1.5545 92 .0702 0 0719 61.9983 0.8091 45 62.4010 0.0786 62 3224 1.5271 93 .0587 0 0718 .9869 0.8002 46 .3987 00784 .3202 1.5033 94 .0471 00716 .9755 0 7912 47 .3963 00783 .3 180 1 4795 95 62.0356 0 0715 61.9641 0.7823

' 48 .3940 00781 .3159 1.4556 96 .0233 0 0714 .9519 0.7742 49 .3916 0.0780 .3 137 1.4318 97 62.0110 0.0713 .9398 0.7661 50 62.3893 00778 62 3115 1.4080 98 61.9988 0 0711 .9276 0.7579 S1 .3856 Q0777 .3080 1.3871 99 .9865 0 0710 .9155 0.7498 52 .3819 G0775 .3044 1.3662 100 61.9742 0.0709 61.9033 0.7417

  • 53 .3783 00774 1.3452 L 54 55

.3746 62.3709 00772 00771

.3009

.2973 1.3243 101 102

.9610

.9478 0 0708 0 0707

.8902

.8772' 0.7342 0.7268 62.2938 1.3034 103 .9347 0 0705 .8641 0.7193 56 .3660 Q0769 .2891 1.2848 104 .9215 0 0704 .8511 0.7119 57 .3611 00768 .2843 1.2662 105 61.9083 0.0703 61.8380 0.7044 1 58 .3562 00766 .2796 1.2477 106 .8947 0 0702 .8246 0.6976 59 .3516 a00765 .2748 1 2291 107 .8812 0.0700 .8112 0 6908 60 62.3464 00763 62.2701 1.2105 108 .8676 0.0699 .7977 0 6839 61 3404 0.0762 .2643 1.1942 109 .8541 0 0697 .7843 0.6771 62- .3345 00760 .2585 1.1779 110 61.8405 00696 61.7709 0 6703 63 .3285 - 00759 .25 26 1 1615 Ill .8260 0 0695 .7565 0.6640 64 .3226 G0757 2468 1.1452 112 .8115 0.0694 .7422 0.6576 1 65 62.3166 0.0756 62 2410 1.1289 113 .7971 0.0692 .7278 0.6513 L 66 .3096 0.0755 .2341 1 1142 114 .7826 0 0691 .7135 0 6449 L 67 68 69

.3025

.2955

.2884 0.0753 0 0752 0.0750 2272

.2203

.2134 1.0995 1.0847 1.0700 115 116 117 61.7681

.7531

.7381 0 0690 0.0689 0 0688 61.6991 6842

.6693 0 6386 0 6328 0 6270 70 62 2814 0.0749 62.2065 1.0553 118 .7230 0 0686 .6544 0.6211 71 2734 0.0748 .1986 1.0421 119 .7080 0 0685 .6395 0.6153 72 2653 0.0746 .1907 1.0289 120 61.6930 0.0684 61.6246 0.6095 L .2573 0.0745 .18 28 1.0156 121 .6773 0.0683 .6090 0 .6041 74 .2492 0.0743 .1749 1.0024 122 .6615 0 0682 .5933 0.5987 75 62 2412 0 0742 62.1670 0.9892 123 .6458 0 0681 5777 0.5934 76 .2322 0 0741 .1582 0.9773 124 .6300 0.0680 .5620 0.5880 77 .223' 0 0739 1493 0.9654 125 61.6143 0 0679 61.5464 0.5826 78 .2143 0 0738 .1405 0.9535 79 .2053 0.0736 .1316 0 9416 6-1-76

CALIBRATION INDIAN POINT TWO 18" ULTRASONIC FLOWMETERS SERIAL NUMBERS WMP #2, and WMP #3 WESTINGHOUSE ELECTRIC CORPORATION PURCHASE ORDER NUMBER 87GC-WJ-47042 MAY, JUNE 1979 - ARL NO. 106-79/C91 Certified By Albert G. Ferron George E. Hecker, Director ALDEN RESEARCH LABORATORY WORCESTER POLYTECHNIC INSTITUTE HOLDEN, MASSACHUSETTS 01520

SUMMARY

Two 18" Ultrasonic Flowmeters, Serial Numbers WMP #2 and WMP #3, were calibrated at the ALDEN RESEARCH LABORATORY of Worcester Polytechnic Institute for WESTINGHOUSE ELECTRIC CORPORATION under their Purchase Order Number 87GC-WJ-47042.

Both 18" ultrasonic flowmeters were calibrated using straight pipe configurations. The details of the piping immediately upstream and downstream of the ultrasonic meters are shown on the piping diagram.

The results are presented in the form of a tabulated data sheet as well as a plot of the average temperature and pressure corrected mean meter factor (T,P) versus the average flow rate (q) in cubic feet per second or pounds per hour.

The flow meters were calibrated at seven flow rates ranging from 5 to 35 cubic feet per second, or 1 to 7.7 million pounds per hour.

Tests 1 and 4 in May, 1979 identified a scaling error of approximately 0.45% which was corrected for Tests 5 and 6 conducted in June, 1979.

The error was determined to be due to incorrect placement of the paths in the LEFM meter section.