ML020390002

From kanterella
Jump to navigation Jump to search
Summary of Telephone Conversation on 02/07/2002
ML020390002
Person / Time
Issue date: 02/07/2002
From: Samson Lee
NRC/NMSS/IMNS/MSIB
To: Kelly B
eV Products
References
SSD 01-30
Download: ML020390002 (1)


Text

CONVERSATION RECORD (K:\\Document Processing Center_Ascent\\ADAMSP8-2039\\downloads\\001 -

100\\065\\ML020390002.wpd)

TYPE:

Outgoing Telephone Incoming Telephone x

Meeting NAME OF PERSON CONTACTED:

Kevin Parham, et al., Brian Kelly ORGANIZATION: eV Products TIME: 03:00 PM DATE:

02/07/02

SUBJECT:

CPC-48 Users Manual

SUMMARY

eV Products called us (SJL, JPJ2, Larry Campbell) and discussed the user manual instruction for cleaning of device. I pointed out that region 1 has potential high exposure rate to the cleaning personnel so that personnel should have cleaning training, which is not complied with 10 CFR 32.51(a) requirements (i.e., any training not required). We suggested eV Products to send an official letter to interpret that requirements.

ACTION REQUIRED:

PLACE THIS RECORD IN:

Registration File NR-QA File Incident File General File PERSON DOCUMENTING THE CONVERSATION:

Seung J. Lee SIGNATURE:

/RA/

DATE:

02/07/02