L-90-199, Submits Rept Detailing Investigative Analysis of Unsatisfactory Blind Specimen Results,Identification of Causes & Corrective Actions Taken by Lab to Prevent Recurrence,Per 10CFR26,App A.2.8(e)(4)

From kanterella
(Redirected from L-90-199)
Jump to navigation Jump to search
Submits Rept Detailing Investigative Analysis of Unsatisfactory Blind Specimen Results,Identification of Causes & Corrective Actions Taken by Lab to Prevent Recurrence,Per 10CFR26,App A.2.8(e)(4)
ML17348A265
Person / Time
Site: Turkey Point, Saint Lucie  NextEra Energy icon.png
Issue date: 05/29/1990
From: Acosta R
FLORIDA POWER & LIGHT CO.
To:
NRC OFFICE OF INFORMATION RESOURCES MANAGEMENT (IRM)
References
L-90-199, NUDOCS 9006040253
Download: ML17348A265 (6)


Text

ACCELERATED DI UTION DEMONS ATION SYSTEM

~

~

REGULATORY XNFORMATION DISTRIBUTION SYSTEM (RIDS)

ACCESSION NBR:9006040253 DOC.DATE: 90/05/29 NOTARIZED:

NO FACIL:50-250 Turkey Point Plant, Unit 3, Florida Power and Light C 50-251 Turkey Point. Plant, Unit 4, Florida Power and Light C 50-335 St. Lucie Plant, Unit 1, Florida Power 6 Light Co.

50-389 St. Lucie Plant, Unit 2, Florida Power 6 Light Co.

AUTH.NAME AUTHOR AFFILIATION ACOSTA,R.J.

Florida Power

& Light Co.

RECXP.NAME RECXPIENT AFFILXATION Document Control Branch (Document Control Desk)

DOCKET g

05000250 05000251 05000335 05000389

SUBJECT:

Submits rept of unsatisfactory performance testing incident, per 10CFR26,App A,Section 2.8(e).

D1STRIBUTION CODE:

AOOID COPIES RECEIVED:LTR g ENCL /

SIZE:

TITLE: OR Submittal:

General Distribution NOTES:

RZ FILE EXTERNAL: LPDR NSIC 01 RECIPIENT ID CODE/NAME PD2-2 LA EDISON,G INTERNAL'CRS NRR/DOEA/OTSB11 NRR/DST/SELB 8D NRR/DST/SRXB 8E COPIES LTTR ENCL 1

1 5

5 6

6 1

1 1

1 1

1 1

0 1

1 2

2 1

1 RECIPIENT ID CODE/NAME PD2-2 PD NORRIS,J NRR/DET/ECMB 9H NRR/DST 8E2 NRR/DST/SICB 7E NUDOCS-ABSTRACT OGC/HDS2 RES/DSIR/EIB NRC PDR COPIES LTTR ENCL 1

1 5

5 1

1 1

1 1

1 1

1 1

0 1

1 1

1

~ki~ Ks~> i NOTE TO ALL"RIDS" RECIPIENTS:

PLEASE HELP US TO REDUCE WASTE! CONTACT THE DOCUMENT CONTROL DESK, ROOM Pl-37 (EXT. 20079) TO ELIMINATEYOUR NAME FROM DISTRIBUTION LISIS FOR DOCUMENTS YOU DON'T NEED!

Qfi gP TOTAL NUMBER OF COPIES REQUIRED:

LTTR M3 ENCL

P.O. Box14000, Juno Beach, FL 33408-0420 wx 29 199'-90-199 10 CFR 26 U. S. Nuclear Regulatory Commission Attn:

Document Control Desk Washington, D.C.

20555 Gentlemen:

Re:

Turkey Point Units 3 and 4

St Lucie Units 1 and 2

Docket Nos.

50-250, 50-251, 50-335, and 50-389 10 CFR 26 Unsatisfactor Performance Testin Incident Re ort Pursuant to 10 CFR 26 Appendix A.2.8(e)(4), Florida Power 6 Light Company (FPL) is submitting the enclosed report of an unsatisfactory performance testing incident to the NRC.

In accordance with the requirements of 10 CFR 26, Appendix A,

Section 2.8(e),

FPL submitted blind performance test specimens to FPL's contract Department of Health and Human Services (DHHS) certified laboratory, Roche Biomedical Laboratories.

The enclosed report details the investigative analysis of unsatisfactory blind specimen results, the identification of causes, and the corrective actions taken by the laboratory to prevent recurrence.

Please contact us if additional information is required.

Very truly yours, R. Z. Acosta Acting Vice President Nuclear Energy RTA/GRM/slh Enclosure cc:

Stewart D. Ebneter, Regional Administrator, Region II, USNRC Senior Resident Inspector, USNRC, Turkey Point Plant Senior Resident Inspector, USNRC, St. Lucie Plant 9006000253 900529 PDR ADOCK 05000250 PDC

~~ p~~ pcs//s~e aot an FPL Group company

.Roche Biomedical Laboratories a subsidiantr of Koffmann-La Roche Inc.

8

-" 24 Roche Biomedcal Laboratones, Inc.

Post Offce Box 2230 Burtington, North Carohna 27215.2230 Tetephone: 919 584 5171 May 22, 1990 INVESTIGATIVE FINDINGS AND CORRECTIVE ACTIONS FOR UNSATISFACTORY BIIND PERFORMANCE TEST RESULTS PROBLEM STATEMENT:

In fulfillment of the requirements specified in 10 CFR part 26 Fitness for Duty Programs, Appendix A, Section 2.8(e),

Florida Power 5 Light Company (FPL) submitted blind performance test specimens to Roche Biomedical Laboratories.

Of the blind specimens submitted, a fraction of the samples were fortified with the drugs.

After receipt of the results, FPL notified Roche Laboratories of unsatisfactory results for four (4) of the drug fortified urine specimens.

The following report details the investigative analysis of the problem.

the identification of the causes and the corrective actions taken.

CODEINE/BENZODIAZEPINE FORTIFIED SAMPLES Specimen Numbers:

003-000-5042-0 037-000-5090-0 Analysis:

Both of these samples were fortified with benzodiazepine and codeine.

A review of the data revealed that both substances were initially identified by the EMIT screening method.

Gas Chromatography Mass Spectrometry (GC/MS) confirmation testing resulted in reporting a positive for benzodiazepines for both specimens but negative for codeine.

A review of the GC/MS data indicates, that codeine was likely present.

However, in each case the samples were subjected to the GC/MS analysis two times, and in both cases, for each

sample, the relative abundance data for one of the ion masses monitored did not meet our specifications.

Our S.O.P.

requires that the relative abundance for the monitored ion masses must be within 20 4

of that obtained for the standards'n each batch.

Unless all criterion are met for a positive confirmation, a

negative result is reported, as was the case for these specimens.

If the qualitative criterion had been

met, the quantitative results obtained for the specimens for codeine were as follows:

003-000-5042-0 037-000-5090-0 1026 ng/ml, 800 ng/ml, 921 ng/ml 646 ng/ml

Identification of Cause:

Our personnel have audited a number of our GC/MS packets for opiate confirmations.

and have encountered a few other samples which gave similar data to the above two specimens.

The problem could not be identified but is most likely due to an interference in our assay in certain samples.

Based upon our analysis, following action has been taken.

Corrective Actions:

We have initiated development of a method using a different extraction column and a different derivatization.

Previously we have used an XAD resin; the new method uses a bonded phase.

The old procedure used butyl derivatives whereas the new procedure uses TMS derivatives.

It is anticipated that this procedure will provide cleaner extracts and an improved chromatographic performance.

The development and validation of new assay procedure was completed in early May.

The procedure was evaluated on samples which have given problems with the old method, and was found acceptable.

The new procedure was implemented on May 10, 1990.

AMPHETAMINES/OPIATES FORTIFIED SAMPLES Specimen Number:

061-000-5107-0 Analysis:

This sample was fortified with amphetamines and opiates.

A review of the analyses revealed that the initial screening test was positive for amphetamine and opiates.

A technician inadvertently transcribed the opiate result to the worksheet as a negative.

Our procedure requires that the screening results be transcribed into our data system.

A second individual is required to review the result entry.

This procedure was documented as p'erformed.

However, the review had compared the manually transcribed result to the data in the computer and did not identify the error.

A review of approximately 5000 records preceding and following this incident showed this to be an isolated administrative error.

Identification of Cause:

A review of all of the reports indicated that a technician inadvertently transcribed the result as negative and a negative result was issued.

Additionally, the individual responsible for the review of the data entry did not identify the problem because the second review compared the data entry against the manually transcribed worksheet instead of against the instrument printout.

Corrective Actions:

I P p I have reconfirmed with our staff that the confirmation review must compare the data entry to the instrument printout of results.

Additionally, we are in the process of implementing a

bar code identification system with an on-line transmission of data after supervisory and certifying official review.

Implementation of this system is anticipated to be completed by August 1990.

MORPHINE/CODEINE FORTIFIED SAMPLES Specimen Number:

086-000-5069-0 Analysis:

This sample was fortified with morphine and codeine.

The test results reported were a positive for codeine and negative for morphine.

A review of the analyses revealed that the sample screened positive for opiates.

Further GC/MS testing resulted in reporting a positive for codeine.

A review of the GC/MS data indicate that morphine was probably present at approximately 800 ng/ml.

However, after a

repeat analyses.

we were unable to obtain a definitive confirmation for morphine.

The ion ratios were not acceptable for a positive confirmation.

Identification of Cause:

This is similar to the validation problems identified for the first two specimen problems in this report.

The cause is believed to be the same as for samples (003-000-5042-0 and 037-000-5090-0).

Corrective Action:

As stated for the corrective action for the first two samples we have implemented a

new validation procedure and have found it to perform to our satisfaction.

The new procedure was implemented on May 10, 1990.

The procedure uses a solid phase extraction (bonded phase) and derivatization with BSTFA and TMCS.

Several samples which appeared to contain morphine and /or codeine by the old procedure, but failed to meet specifications were tested by the new procedure with satisfactory results.

Summary:

This report is being submitted for FPL to forward to the Nuclear Regulatory Commission in accordance with 10 CFR 26, Appendix A, Section 2.8 (e)

(4) and is signed by the individual responsible for the day to day management and operation of our HHS-certified laboratory.

Respectfully Submitted by (insert name, title)

-70 Si ned date

1[

~ t