IR 05000824/2009014

From kanterella
Jump to navigation Jump to search
Notice of Violation from Insp on 851107.Violation Noted: Linearity Checks Not Performed Between 830712-841001 & Weekly Wipe Surveys Not Performed for Wks of 850824-0914
ML20137R281
Person / Time
Site: 03013661, 05000824
Issue date: 11/21/1985
From:
NRC OFFICE OF INSPECTION & ENFORCEMENT (IE REGION II)
To:
Shared Package
ML20137R271 List:
References
30-13661-85-01, 30-13661-85-1, NUDOCS 8512050466
Download: ML20137R281 (2)


Text

-

I

.

.-

i ENCLOSURE NOTICE OF VIOLATION Fairmont General Hospital License No. 47-17929-01 The following violations were identified during an inspection conducted on November 7,1985. The Severity Levels were assigned in accordance with the NRC Enforcement Policy (10 CFR Part'2, Appendix C).

License Condition 19 requires that except as specifically provided otherwise by this license, the licensee shall possess and use licensee material described in Items 6, 7, and 8 of this license in accordance with state-ments, representations, and procedures contained in application dated December 23, 1977; letters dated July 5, 1978 and August 7, 1983; and L

application dated August 17, 1983.

The Nuclear Regulatory Commission's regulations shall govern the licensee's statements in applications or letters, unless the statements are more restrictive than the regulations.

1.

Item 10 of the application dated August 17, 1983, requires that linearity tests of the dose calibrator be performed on a quarterly i

basis.

,j Contrary to the above, linearity checks were not performed between July 12, 1983 and November 14, 1983, November 14, 1983 and May 17, 1984, and May 7, 1984 and October 1, 1984. These dates reflect three intervals that exceed the requirement.

This is a Severity Level IV violation (Supplement VI).

2.

Item 17 of the application dated August 17, 1983, requires that all l

elution, preparation, and injection areas be surveyed daily with an appropriate low-range survey meter and that a wipe survey be performed weekly.

Contrary to the above, daily surveys have not been performed since

!

January 1983, and weekly wipe surveys were not performed for the weeks

!

of August 24, August 31,. September 7, and September 14, 1985.

This is a Severity Level IV violation (Supplement VI).

,

l

'

Pursuant to 10 CFR 2.201, you are required to submit to this office within

!

30 days of the date of this Notice, a written statement or explanation in reply, l

including:

(1) admission or denial of the alleged violations; (2) the reasons

'

for the violations if admitted; (3) the corrective steps. which have been taken and the results achieved:- (4) corrective steps which will be taken to avoid further violations; and (5) the date when full compliance will be achieved.

,

.

!

'

C512050466 851121 REG 2 LIC30 47-17929-01 PDR l

.-

_

. _..

.

..

,-,

.

O'

Notice of Violation

Security or safeguards.information should be submitted as an enclosure to facilitate withholding' it from public disclosure as required by 10 CFR 2.790(d)

or 10 CFR 73.21.

Date:

NOV 21 MS i

.

l l '

l l.

l-l

..

y g

3-g-

-,-s+

-9-+- - - -

- rf e-.

wyy

,e g

9 m.a--p9 y

,e-,

y a

ywe

-

,

p--.a av-,e-"W-t*9r