The following report was received by the
California Department of Public Health, Radiation Health Branch (
RHB) via email:
The University of California, Los Angeles (UCLA) reported a medical event that had occurred on Monday, August 5, 2024, during a Y-90 Sirtex SIR-Sphere liver cancer treatment.
The target organ was the patient's left liver lobe. The authorized user prescribed an activity of 0.71 GBq (19.14 mCi).
Unfortunately, the target organ only received 7.72 mCi or 40.33 percent of the prescribed dose.
The interventional radiologist reported that the cause of the under-delivery was a leak in the SIR-Sphere tubing system (delivery box and covered table). All contaminated equipment is being held for decay by UCLA.
UCLA will submit a 15-day written report to the RHB.
California Event Number: 080524