The following is a summary of information received by email from the
Illinois Office of Nuclear Safety:
On January 31, 2024, a medical administration of Y-90 microspheres at Rush University Medical Center failed to deliver the intended dose. A patient had been prescribed 34.99 mCi (500 Gy) of yttrium-90 (Y-90) Theraspheres intended for a liver treatment. The post-administration survey indicated a delivered dose of only 23.05 mCi (330 Gy), which was a 34 percent underdose. The patient and physician were notified the day of the treatment. There was no adverse impact on the patient and the patient will not need to be retreated. The administering physician reported significant resistance in the initial and subsequent flushes.
IL NMED Item Number: IL240003
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.