The following information was provided by the
Ohio Bureau of Environmental Health and Radiation Protection via email:
Ohio State University reported a medical event on March 8, 2023. A written directive for 200 millicuries of lutetium-177 (PLUVICTO) to be administered intravenously was signed by an authorized user. The dosage was assayed at 195.57 millicures and was administered to the patient on March 7, 2023. During the administration, the Nuclear Medicine Technologist noted some drips from the tubing. An investigation was initiated and the results on March 8, 2023, indicate the patient received 157.57 millicuries of lutetium-177 (PLUVICTO), 21.5 percent less than the written directive. The investigation will continue to determine the root cause of the medical event and determine corrective actions if applicable.
Reference Number OH 2023-006
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.