The following is a summary of information provided by the
Washington State Department of Health (
DOH), Radioactive Materials Section via email:
The licensee reported that a Y-90 TheraSphere liver treatment was completed in accordance with manufacturer requirements and nothing out of the ordinary was observed. However, during a subsequent step to flush the system with saline, the technician, who is not an authorized user or radioactive material user, noticed bubbles coming up above the vial. Wipe samples of the procedure equipment were counted and showed radioactivity. A second flush of the system was initiated but the same bubbles were observed and therefore the flush was stopped, and the catheter was retracted from the patient. Additional flushes were not attempted. An evaluation showed that the patient received only 33 percent of the intended dose. The activity contained in the wipes and cleaning materials accounted for the balance of the intended dose (66 percent) and was disposed of as waste. DOH has requested additional information regarding potential occupational exposures. The licensee will submit a detailed report.
Washington Incident Number: WA-23-003.
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.