During a 2/16/22 inspection, inspectors discovered an unreported medical event at The Carle Foundation (IL-01156-01) that occurred on December 30, 2021. No adverse patient impact reported. This was reported to the Headquarters Operations Officers on 2/17/2022.
DETAILS: Agency inspectors performed a routine inspection on 2/16/2022, of The Carle Foundation d/b/a The Carle Foundation Hospital. During a review of Y-90 Theraspheres procedures on the afternoon of 2/16/2022, inspectors noted an unreported medical event which occurred on December 30, 2021. Inspectors noted that on 12/30/2021, a written directive to deliver 0.30 GBq Y-90 Theraspheres to the left hepatic artery was prepared; however, only 0.11 GBq (37 percent) was delivered. The licensee reported using a smaller microcatheter than usual.
No further treatment to the patient is planned. The authorized user advised that the dose delivered was medically satisfactory for this case.
The event remains under investigation.
Item Number: IL220005
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.