The following is a summary of an email received the State of Washington:
The licensee reported a medical event involving an underdose administration of Y-90 Theraspheres on April 16, 2021. The event resulted in an underdose to the liver. Two vials of Y-90 were to be delivered to two different locations in the liver, however surveys after injection of both vials revealed not all activity in vials made it into the patient. The planned activity for each dose was 0.79 GBq. Post treatment measurement of percent dose delivered for dose `A' and dose `B' was 60 percent and 76 percent, respectively. Gamma camera imagining confirmed no Y-90 Theraspheres in the dose-vial. However, imaging visualized activity retained in the delivery tubing. Routine post-imaging demonstrated microsphere distribution in liver segments 4A and 4B. The activity delivered was 0.465 GBq and 0.594 GBq, respectively.
Washington State Incident Number WA-21-008.
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.