The following was received from the Commonwealth of
Kentucky via fax:
On 6/28/18 while conducting a 6-month physical inventory and shutter check (on-off mechanism) the licensee, ARKEMA, Inc., discovered the shutters malfunctioned on two separate fixed gauging devices, namely:
1) TN Technologies fixed gauging device model 5208 Serial No. MB 4251 containing 8 Ci Cs-137. The licensee has attempted to contact Thermo Scientific to repair the shutter. The root cause of the failure of the on/off mechanism was due to corrosion and dust due to the gauging device's environment. This licensee produces industrial chemicals and the gauging device is operated in a corrosive environment. This event was reported to the [Kentucky Radiation Health Branch] KY RHB as required by regulation on 6/29/18.
2) Ronan Engineering fixed gauging device model SA-1 Serial Number M8107, containing 5 Ci Cs-137. The licensee has contacted Ronan Engineering and repairs will be conducted during the week starting on 7/16/18. This licensee produces industrial chemicals and the gauging device is operated in a corrosive environment. The event was reported to KY RHB as required by regulation on 6/29/18.
No accidental overexposures have been reported due to the malfunction.
Kentucky Event Report ID No.: KY180002