The following information was received via e-mail:
L3 Technologies reported a leaking foil sealed source to the Agency (Massachusetts Radiation Control Program). A wipe of the source was taken on Monday, 10/16/17 and results were received on Wednesday, 10/18/17. The leak test was measured at 0.0089 microCuries.
Source Information: Ni-63, 15 mCi, Model # 19233, Serial # L7947.
Device Information: Hewlett Packard Model # 5890, Serial # C-128/83.
The foil source is housed in a gas chromatograph unit that has not been used in many years. The unit is covered and locked away in a storage room. Personnel do not routinely access the unit. Leak tests have been performed every 6 months. There is no immediate risk of contamination at the facility.
The licensee expressed that they will appropriately dispose of the source and unit.
The licensee is required to submit a written report to the Agency [Massachusetts Radiation Control Program] within 30 days.