The following was received from the State of
Wisconsin via email:
On May 4, 2015, the Wisconsin Radiation Protection Section received a notice from an Authorized Medical Physicist [AMP] at Gundersen Lutheran Medical Center, Inc., of a medical event that occurred during a Yttrium-90 Therasphere procedure on May 1, 2015. The licensee did not have enough information to determine if the procedure constituted a medical event on May 1, 2015. Upon further evaluation, the AMP determined it was a medical event as of 3 pm CST May 4, 2015. The patient was scheduled to receive two separate Y-90 Therasphere doses to two separate segments of the liver. During the first injection, the overpressure valve opened and filled the overflow vial with the Authorized User applying very little pressure to the saline filled syringe. The patient was to get 147 Gy to Segment 6 of the liver. From residue measurements, the patient received 35.5 percent or 52.2 Gy to Segment 6. The Radiation Protection Section will perform an investigation and update through NMED [Nuclear Material Events Database].
Event Report ID No.: WI150006
The correct Wisconsin event ID is WI150007.
Notified the R3DO (Peterson) and NMSS (via e-mail).
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.