The following information was obtained from the Commonwealth of
Pennsylvania via facsimile:
Notifications: The Department [Pennsylvania Department of Environmental Protection] was notified of this event on Wednesday, February 11, 2015. It is immediately reportable as per 10 CFR 35.3045(a)(1)(i).
Event Description: A patient was receiving Y-90 Sirtex Sirspheres when the administering device failed and a portion of the dose was lost in the apparatus. The patient received 10.4 mCi, or 58% of the prescribed 17.82 mCi dose.
Cause of the event: The device came apart during the procedure and the remainder of the dose was contained in the packaging.
Actions: The Department plans a reactive inspection.
Pennsylvania Event Report ID No: PA150004
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.