At approximately 2200 EST, on December 09, 2014, an area operator attempted to open the shutter on source holder 10071961 (Hemlock identifier). The shutter operation failed with the shutter stuck in the closed position. The operator then contacted an area Equipment Care Operator who verified the shutter would not operate. The area operator informed a day tech of the failed shutter operation on December 10, 2014. Operating the shutter was part of a routine annual calibration
PM [preventive maintenance]. The process was previously shut down, therefore, this incident caused no loss of production. The source holder is located at S-457 on
HP tank 35600.
Fixed gauge: Vega America Source Holder Model SHF 2-B with Cs-137 800 millicurie source.
- * * UPDATE FROM CHRISTOPHER VAN HAMPLER TO DANIEL MILLS AT 0904 EST ON 12/15/2014 * * *
The licensee performed a radiation survey of the fixed gauge which indicated higher than expected radiation readings (40 mR/hr at <30 cm). Based on these readings, the licensee has determined that the shutter is stuck in a partially open position. The licensee has contacted the manufacturer to repair the device.
Notified R3DO (Skokowski) and
NMSS Events Notification (via email)