The following information was received via facsimile:
A patient was prescribed 50 Gray to be delivered to a brain tumor bed via a GliaSite balloon inflated with 200 mCi of I-125 saline solution implanted for 66 hours7.638889e-4 days <br />0.0183 hours <br />1.09127e-4 weeks <br />2.5113e-5 months <br />. Upon removal of the balloon on March 7, 2014, it was discovered that the balloon had not inflated and no dose to the tumor bed had been delivered.
A three position stopcock was incorrectly positioned at the time of inflation and the I-125 saline solution was diverted to another syringe rather than to the balloon. All material is accounted for and is being held for DIS. The three position stopcock is not part of the vendor's kit.
Corrective Actions: In-service by the vendor scheduled, revision of policies and procedures to include verification of balloon via medical imaging post-infusion, and in-service to staff on new policies and procedures.
NYS Incident: #1054
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.