A prostate seed implant procedure was terminated after the insertion of 2 needles. Only 5 of 106 intended seeds were implanted (1.55mCi of 32.86mCi). The patient's anatomy (pubic arch) presented interference to the placement of needles/seeds for proper dose distribution. The patient will now be treated with external beam IMRT [intensity modulated radiation therapy] once post implant
CT and dose assessment have been performed in approximately 3 weeks. The patient and referring physician have been notified. The facility notified
NYS DOH [New York State Department of Health] same day, written report with corrective actions has been received. To prevent recurrence the urologist will verify during planning volume study that there are no anatomical obstructions to needle placement.
Report No. NY-13-02
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.