Nine breast cancer patients were treated with a multi-channel high dose rate afterloader with the incorrect dwell positions input into the afterloader controller. The intended prescription was for 5 mm between dwell positions. The actual input was 2.5 mm between dwell positions. The prescribed dose was for 3400 cGy and the actual dose delivered was approximately 25% less. No adverse clinical effects are expected.
The doctors and patients will be notified.
- * * UPDATE FROM DR. VIVIAN ARTERBERY TO JOHN SHOEMAKER AT 1507 EST ON 03/07/11 * * *
There were a total of eleven breast cancer patients involved. Seven of the eleven patients had small areas of unintended target that received unplanned dose.
The doctors and patients have been notified.
Notified R3DO (Riemer) and FSME (McIntosh)