ENS 46101
ENS Event | |
|---|---|
04:00 Jul 7, 2010 | |
| Title | Agreement State Report - Delivered Dose Differed from the Prescribed Dose by Greater than 50% |
| Event Description | The following was received via fax from the Pennsylvania Department of Environmental Protection:
On July 7, 2010 a patient was beginning the first of three vaginal treatment fractions with an Ir-192 HDR. It was discovered on July 14, 2010 that the end of the treatment tube was placed 3.5 cm short of its intended location. When the patient returned for the second treatment, she was imaged again, and staff noticed the treatment tube was in a different location from the previous treatment. Licensee estimates that, for the first fraction, the intended treatment volume received only about 10% of the intended dose for that fraction. Per 10CFR35.3045(a)(1)(iii), it is required for the licensee to report any event in which the fractionated dose delivered differs from the prescribed dose, for a single fraction, by 50% or more. It is believed the medical staff mis-identified the treatment location and the end of the treatment tube was placed 3.5cm short of its intended location. The licensee is considering making up this dose by adding a fourth treatment fraction. There is no anticipated adverse effect to the patient. There is a PaDEP/BRP reactive inspection scheduled to investigate this ME at U Penn. The patient and the referring physician were notified. A follow-up written report from the licensee is expected. PA Report # PA100015
The following was received via fax from the Pennsylvania Department of Environmental Protection: Due to further information received from the licensee, an amendment is being made to the original sent July 16, 2010, the CAUSE OF THE EVENT has been modified to reflect updated information. Event Description: On July 7, 2010 a patient was beginning the first of three vaginal treatment fractions with an Ir-192 HDR. It was discovered on July 14, 2010 that the end of the treatment tube was placed 3.5 cm short of its intended location. When the patient returned for the second treatment, she was imaged again, and staff noticed the treatment tube was in a different location from the previous treatment. Licensee estimates that, for the first fraction, the intended treatment volume received only about 10% of the intended dose for that fraction. Per 10CFR35.3045(a)(1)(iii), it is required for the licensee to report any event in which the fractionated dose delivered differs from the prescribed dose, for a single fraction, by 50% or more. CAUSE OF THE EVENT: The applicator was placed correctly by the medical staff as confirmed by MRI but moved 3.5 cm short of its intended location prior to treatment. The licensee is considering making up this dose by adding a fourth treatment fraction. There is no anticipated adverse effect to the patient. ACTION: There is a PaDEP/BRP reactive inspection scheduled to investigate this ME at U Penn. The patient and the referring physician were notified. A follow-up written report from the licensee is expected. A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient. |
| Where | |
|---|---|
| University of Pennsylvania Philadelphia, Pennsylvania (NRC Region 1) | |
| License number: | PA-0131 |
| Organization: | Pa Bureau Of Radiation Protection |
| Reporting | |
| Agreement State | |
| Time - Person (Reporting Time:+227.73 h9.489 days <br />1.356 weeks <br />0.312 months <br />) | |
| Opened: | David Allard 15:44 Jul 16, 2010 |
| NRC Officer: | John Knoke |
| Last Updated: | Jul 23, 2010 |
| 46101 - NRC Website
Loading map... | |
University of Pennsylvania with Agreement State | |
WEEKMONTHYEARENS 573752024-10-09T04:00:0009 October 2024 04:00:00
[Table view]Agreement State Patient Underdose ENS 570812024-04-17T04:00:00017 April 2024 04:00:00 Agreement State Medical Underdose ENS 566612023-08-05T15:07:0005 August 2023 15:07:00 Agreement State Medical Underdose ENS 570862023-01-31T04:00:00031 January 2023 04:00:00 Agreement State Medical Event ENS 570852022-12-29T04:00:00029 December 2022 04:00:00 Agreement State Medical Event ENS 570842022-12-28T04:00:00028 December 2022 04:00:00 Agreement State Medical Event ENS 558892022-05-04T04:00:0004 May 2022 04:00:00 Agreement State Underdose ENS 558512022-04-18T04:00:00018 April 2022 04:00:00 Agreement State Patient Underdose ENS 553372021-07-02T04:00:0002 July 2021 04:00:00 Agreement State Patient Under Dose ENS 549642020-10-20T04:00:00020 October 2020 04:00:00 Agreement State Missing Licensed Material ENS 534302018-05-25T04:00:00025 May 2018 04:00:00 Agreement State En Revision Imported Date 10/15/2018 ENS 525042017-01-19T05:00:00019 January 2017 05:00:00 Agreement State Agreement State Report - Received Dose Less than Prescribed Dose ENS 524572016-12-20T05:00:00020 December 2016 05:00:00 Agreement State Agreement State Report - Medical Underdose ENS 483032012-09-11T04:00:00011 September 2012 04:00:00 Agreement State Agreement State Report - Patient Underexposure Using Yttrium-90 Sir-Spheres Treatment ENS 482292012-08-23T04:00:00023 August 2012 04:00:00 Agreement State Agreement State Report - Patient Underexposure Using Yttrium-90 Sir-Spheres Treatment ENS 481372012-07-26T04:00:00026 July 2012 04:00:00 Agreement State Agreement State Report - Medical Event Involving an Underdose with Yttrium-90 Spheres ENS 477632012-03-20T04:00:00020 March 2012 04:00:00 Agreement State Agreement State Report - Underdose During Liver Disease Treatment ENS 474352011-11-10T12:00:00010 November 2011 12:00:00 Agreement State Agreement State Report - Underdosage of Theraspheres ENS 461012010-07-07T04:00:0007 July 2010 04:00:00 Agreement State Agreement State Report - Delivered Dose Differed from the Prescribed Dose by Greater than 50% ENS 457272010-01-21T05:00:00021 January 2010 05:00:00 Agreement State Agreement State Report - Brachytherapy Implants Outside of the Intended Treatment Area 2024-04-17T04:00:00 | |