The following information was received from the State via e-mail:
On June 25, 2009, at 1630 the Agency was notified by the licensee that while looking for information requested by the State for a previous gauge failure ( EN 44981), the licensee discovered an additional event. This event occurred in early September, 2008, and had not been reported as required. The event involved an Ohmart/VEGA Model SH-F2 level gauge containing a 300 millicurie ( 228 mCi calculated current activity) Cesium (Cs) - 137 source. The gauge shutter would not close due to a sheared shutter rotor. The manufacturer was notified, and a technician arrived at the facility [located in Odessa, TX] on September 30, 2009. The technician replaced the rotor, cleaned and lubricated the gauge, and the gauge operated normally. The technician found that a great deal of material had collected in the source holder causing the failure. The licensee has removed all of their nuclear gauges and is in the process of terminating their license.
Texas Incident Report # I-8642